{"title":"基于MEMS制造超声换能器的P0-1高分辨率基频和谐波成像","authors":"C. Chandrana, N.A. Kharin, A. Nair, K. Waters, D. Vince, B. Kuban, G. Lockwood, S. Roy, A. Fleischman","doi":"10.1109/ULTSYM.2007.298","DOIUrl":null,"url":null,"abstract":"Intravascular ultrasound (IVUS) imaging is increasingly employed to assist in selecting and evaluating therapeutic intervention. Recent work in IVUS backscatter analysis demonstrates the capability of IVUS to characterize specific lesions and identify plaques that lead to various clinical syndromes. Correct identification of plaque types depends on their structure, composition and sufficient image resolution (< 30 mum axially). Tissue harmonic imaging (THI) has shown promise to increase quality of IVUS. However, Current IVUS catheters, with unfocused transducers and narrow bandwidths, might not benefit from advantages due to harmonic imaging. Hence, we are developing miniature focused large bandwidth ultrasonic transducers with high resolutions and harmonic imaging capabilities.We made spherically focused broad bandwidth (120%) PVDF- TrFE transducers using MEMS techniques. We characterized the transducers. We also developed a method for multiple ultrasonic modalities using existing hardware and employing pulse-inversion techniques where a single transducer can be used to produce four modes of images. In this study, we obtained images of the coronary ostium in four imaging modes (Fundamental 20 MHz, Fundamental 40 MHz, Harmonic 40 MHz and Harmonic 80 MHz). We also excited our transducer with 40 MHz monocycle pulse for maximal axial resolution to generate standard mode images. Our standard mode images were compared to histology and commercial IVUS Systems. Axial resolutions with a monocycle pulse were typically less than 19 mum with a bandwidth of ~ 120 %. The lateral resolution characterization was done using a 1 mm transducer with an f-number 2.6. Harmonic signals showed better resolution compared to the fundamental signals. Ex vivo circumferential images of the human coronary ostium at F20, F40, H40 and H80 were acquired. The standard mode ultrasonic image of the human aorta using the MEMS transducer clearly displayed better delineation of the media compared to the existing commercial IVUS systems. When comparing the standard mode 40 MHz ultrasonic image to its histology, our ultrasonic image showed near histological resolution and identified various features in the histology. Such promising results suggest that the high resolution transducers with multi modality imaging capabilities can improve the clinical use of IVUS for cardiovascular diseases.","PeriodicalId":6355,"journal":{"name":"2007 IEEE Ultrasonics Symposium Proceedings","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2007-12-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"6","resultStr":"{\"title\":\"P0-1 High Resolution Fundamental and Harmonic Imaging Using a MEMS Fabricated Ultrasonic Transducer\",\"authors\":\"C. Chandrana, N.A. Kharin, A. Nair, K. Waters, D. Vince, B. Kuban, G. Lockwood, S. Roy, A. Fleischman\",\"doi\":\"10.1109/ULTSYM.2007.298\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Intravascular ultrasound (IVUS) imaging is increasingly employed to assist in selecting and evaluating therapeutic intervention. Recent work in IVUS backscatter analysis demonstrates the capability of IVUS to characterize specific lesions and identify plaques that lead to various clinical syndromes. Correct identification of plaque types depends on their structure, composition and sufficient image resolution (< 30 mum axially). Tissue harmonic imaging (THI) has shown promise to increase quality of IVUS. However, Current IVUS catheters, with unfocused transducers and narrow bandwidths, might not benefit from advantages due to harmonic imaging. Hence, we are developing miniature focused large bandwidth ultrasonic transducers with high resolutions and harmonic imaging capabilities.We made spherically focused broad bandwidth (120%) PVDF- TrFE transducers using MEMS techniques. We characterized the transducers. We also developed a method for multiple ultrasonic modalities using existing hardware and employing pulse-inversion techniques where a single transducer can be used to produce four modes of images. In this study, we obtained images of the coronary ostium in four imaging modes (Fundamental 20 MHz, Fundamental 40 MHz, Harmonic 40 MHz and Harmonic 80 MHz). We also excited our transducer with 40 MHz monocycle pulse for maximal axial resolution to generate standard mode images. Our standard mode images were compared to histology and commercial IVUS Systems. Axial resolutions with a monocycle pulse were typically less than 19 mum with a bandwidth of ~ 120 %. The lateral resolution characterization was done using a 1 mm transducer with an f-number 2.6. Harmonic signals showed better resolution compared to the fundamental signals. Ex vivo circumferential images of the human coronary ostium at F20, F40, H40 and H80 were acquired. The standard mode ultrasonic image of the human aorta using the MEMS transducer clearly displayed better delineation of the media compared to the existing commercial IVUS systems. When comparing the standard mode 40 MHz ultrasonic image to its histology, our ultrasonic image showed near histological resolution and identified various features in the histology. Such promising results suggest that the high resolution transducers with multi modality imaging capabilities can improve the clinical use of IVUS for cardiovascular diseases.\",\"PeriodicalId\":6355,\"journal\":{\"name\":\"2007 IEEE Ultrasonics Symposium Proceedings\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2007-12-26\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"6\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2007 IEEE Ultrasonics Symposium Proceedings\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/ULTSYM.2007.298\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2007 IEEE Ultrasonics Symposium Proceedings","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ULTSYM.2007.298","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6

摘要

血管内超声(IVUS)成像越来越多地用于辅助选择和评估治疗干预措施。最近在IVUS反向散射分析方面的工作表明,IVUS能够表征特定病变并识别导致各种临床综合征的斑块。正确识别斑块类型取决于它们的结构、组成和足够的图像分辨率(轴向< 30 μ m)。组织谐波成像(THI)已显示出提高IVUS质量的希望。然而,目前的IVUS导管,具有未聚焦换能器和窄带宽,可能无法从谐波成像的优势中获益。因此,我们正在开发具有高分辨率和谐波成像能力的微型聚焦大带宽超声换能器。我们利用MEMS技术制作了球形聚焦宽带宽(120%)PVDF- TrFE换能器。我们对换能器进行了表征。我们还开发了一种使用现有硬件和脉冲反演技术的多种超声模式的方法,其中单个换能器可用于产生四种模式的图像。在本研究中,我们获得了四种成像模式(基本20 MHz,基本40 MHz,谐波40 MHz和谐波80 MHz)的冠状动脉口图像。我们还用40 MHz单周脉冲激励换能器以获得最大轴向分辨率以生成标准模式图像。我们的标准模式图像与组织学和商业IVUS系统进行了比较。单周期脉冲的轴向分辨率通常小于19 μ m,带宽约为120%。横向分辨率表征使用1毫米换能器,f值为2.6。谐波信号比基波信号具有更好的分辨率。获取人冠状动脉口F20、F40、H40和H80的离体周长图像。与现有的商用IVUS系统相比,使用MEMS换能器的人体主动脉标准模式超声图像清楚地显示出更好的介质描绘。将标准模式40 MHz超声图像与其组织学进行比较,我们的超声图像显示出接近组织学的分辨率,并识别出组织学中的各种特征。这些有希望的结果表明,具有多模态成像能力的高分辨率传感器可以提高IVUS在心血管疾病中的临床应用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
P0-1 High Resolution Fundamental and Harmonic Imaging Using a MEMS Fabricated Ultrasonic Transducer
Intravascular ultrasound (IVUS) imaging is increasingly employed to assist in selecting and evaluating therapeutic intervention. Recent work in IVUS backscatter analysis demonstrates the capability of IVUS to characterize specific lesions and identify plaques that lead to various clinical syndromes. Correct identification of plaque types depends on their structure, composition and sufficient image resolution (< 30 mum axially). Tissue harmonic imaging (THI) has shown promise to increase quality of IVUS. However, Current IVUS catheters, with unfocused transducers and narrow bandwidths, might not benefit from advantages due to harmonic imaging. Hence, we are developing miniature focused large bandwidth ultrasonic transducers with high resolutions and harmonic imaging capabilities.We made spherically focused broad bandwidth (120%) PVDF- TrFE transducers using MEMS techniques. We characterized the transducers. We also developed a method for multiple ultrasonic modalities using existing hardware and employing pulse-inversion techniques where a single transducer can be used to produce four modes of images. In this study, we obtained images of the coronary ostium in four imaging modes (Fundamental 20 MHz, Fundamental 40 MHz, Harmonic 40 MHz and Harmonic 80 MHz). We also excited our transducer with 40 MHz monocycle pulse for maximal axial resolution to generate standard mode images. Our standard mode images were compared to histology and commercial IVUS Systems. Axial resolutions with a monocycle pulse were typically less than 19 mum with a bandwidth of ~ 120 %. The lateral resolution characterization was done using a 1 mm transducer with an f-number 2.6. Harmonic signals showed better resolution compared to the fundamental signals. Ex vivo circumferential images of the human coronary ostium at F20, F40, H40 and H80 were acquired. The standard mode ultrasonic image of the human aorta using the MEMS transducer clearly displayed better delineation of the media compared to the existing commercial IVUS systems. When comparing the standard mode 40 MHz ultrasonic image to its histology, our ultrasonic image showed near histological resolution and identified various features in the histology. Such promising results suggest that the high resolution transducers with multi modality imaging capabilities can improve the clinical use of IVUS for cardiovascular diseases.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
10B-3 Vibrating Interventional Device Detection Using Real-Time 3D Color Doppler P5E-8 The Method of Reverberation-Ray Matrix - A New Matrix Analysis of Waves in Piezoelectric Laminates P1D-4 Characteristics of a Novel Magnetic Field Sensor Using Piezoelectric Vibrations P5C-3 Field Simulation Parameters Design for Realistic Statistical Parameters of Radio - Frequency Ultrasound Images 2F-1 Fabrication and Performance of a High-Frequency Geometrically Focussed Composite Transducer with Triangular Pillar Geometry
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1