血清雌二醇水平对了解两性生理/临床环境至关重要:低雌二醇的测量极限和敏感免疫测定的评估

Q3 Medicine Open Biomarkers Journal Pub Date : 2019-01-01 DOI:10.36648/2472-1646.5.2.62
C. Nencetti, M. Sessa, P. Vitti, M. C. Anelli, M. Tonacchera, P. Agretti
{"title":"血清雌二醇水平对了解两性生理/临床环境至关重要:低雌二醇的测量极限和敏感免疫测定的评估","authors":"C. Nencetti, M. Sessa, P. Vitti, M. C. Anelli, M. Tonacchera, P. Agretti","doi":"10.36648/2472-1646.5.2.62","DOIUrl":null,"url":null,"abstract":"Background: Measure of serum 17β-Estradiol is essential to understand physiology, development and health of reproductive processes in both sexes. Commercially immunoassays are not enough sensitive and accurate to assess low concentrations of E2. Purpose of this study was to compare a new sensitive immunoassay for estradiol measurement with respect to method current in our laboratory and to evaluate the performance of the new method. Methods: Four pools of patient sera with E2 concentrations close to clinical decision values were prepared. To test the repeatability of new method the 5x5 protocol was used and CVs were calculated. To evaluate the performance of new method, 50 samples with E2 concentrations covering the whole measurable interval were selected and assayed. Linearity dilution, LoB (Limit of Blank) and LoD (Limit of Detection) were determined. Results: The new assay showed good total repeatability demonstrated by low CV values, and good linear relationship with respect to current method (R=0.9926) as demonstrated by linear regression. Non-parametric regression showed for the new method a slight constant and proportional systematic error that, however, resulted not significant from difference plot analysis, with a general tendency to overestimate results for the current method. Performances of the new method resulted acceptable within the maximum admissible error derived from the literature, and a good linearity over a wide range of dilutions was showed. LoD value confirmed an amelioration in measurement of low estradiol. Conclusion: In conclusion, sensitive assay is suitable to replace the method used in our laboratory with significant improvement in the measurement of low serum estradiol levels.","PeriodicalId":39398,"journal":{"name":"Open Biomarkers Journal","volume":"66 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Estradiol Serum Levels are Crucial to Understand Physiological/Clinical Setting in Both Sexes: Limits of Measurement of Low Estradiol and Evaluation of a Sensitive Immunoassay\",\"authors\":\"C. Nencetti, M. Sessa, P. Vitti, M. C. Anelli, M. Tonacchera, P. Agretti\",\"doi\":\"10.36648/2472-1646.5.2.62\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background: Measure of serum 17β-Estradiol is essential to understand physiology, development and health of reproductive processes in both sexes. Commercially immunoassays are not enough sensitive and accurate to assess low concentrations of E2. Purpose of this study was to compare a new sensitive immunoassay for estradiol measurement with respect to method current in our laboratory and to evaluate the performance of the new method. Methods: Four pools of patient sera with E2 concentrations close to clinical decision values were prepared. To test the repeatability of new method the 5x5 protocol was used and CVs were calculated. To evaluate the performance of new method, 50 samples with E2 concentrations covering the whole measurable interval were selected and assayed. Linearity dilution, LoB (Limit of Blank) and LoD (Limit of Detection) were determined. Results: The new assay showed good total repeatability demonstrated by low CV values, and good linear relationship with respect to current method (R=0.9926) as demonstrated by linear regression. Non-parametric regression showed for the new method a slight constant and proportional systematic error that, however, resulted not significant from difference plot analysis, with a general tendency to overestimate results for the current method. Performances of the new method resulted acceptable within the maximum admissible error derived from the literature, and a good linearity over a wide range of dilutions was showed. LoD value confirmed an amelioration in measurement of low estradiol. Conclusion: In conclusion, sensitive assay is suitable to replace the method used in our laboratory with significant improvement in the measurement of low serum estradiol levels.\",\"PeriodicalId\":39398,\"journal\":{\"name\":\"Open Biomarkers Journal\",\"volume\":\"66 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Open Biomarkers Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.36648/2472-1646.5.2.62\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Open Biomarkers Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.36648/2472-1646.5.2.62","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

背景:血清17β-雌二醇的测定对了解两性生殖过程的生理、发育和健康至关重要。商业免疫测定法在评估低浓度E2时不够灵敏和准确。本研究的目的是比较一种新的用于测定雌二醇的灵敏免疫分析法与我们实验室现有的方法,并评价新方法的性能。方法:制备4个E2浓度接近临床判定值的患者血清。为验证新方法的重复性,采用5x5方案并计算cv。为了评价新方法的性能,选择了50个E2浓度覆盖整个测量区间的样品进行了分析。测定线性稀释度、空白限(LoB)和检出限(LoD)。结果:新方法重复性好,CV值低,与现有方法线性关系良好(R=0.9926)。非参数回归表明,新方法有轻微的常数和比例系统误差,但差异图分析结果不显著,普遍倾向于高估现有方法的结果。新方法的性能在文献中得出的最大允许误差范围内是可接受的,并且在广泛的稀释度范围内显示出良好的线性。LoD值证实了低雌二醇测量的改善。结论:本实验室对低血清雌二醇水平的测定有明显改善,敏感性测定法适合替代本实验室使用的方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Estradiol Serum Levels are Crucial to Understand Physiological/Clinical Setting in Both Sexes: Limits of Measurement of Low Estradiol and Evaluation of a Sensitive Immunoassay
Background: Measure of serum 17β-Estradiol is essential to understand physiology, development and health of reproductive processes in both sexes. Commercially immunoassays are not enough sensitive and accurate to assess low concentrations of E2. Purpose of this study was to compare a new sensitive immunoassay for estradiol measurement with respect to method current in our laboratory and to evaluate the performance of the new method. Methods: Four pools of patient sera with E2 concentrations close to clinical decision values were prepared. To test the repeatability of new method the 5x5 protocol was used and CVs were calculated. To evaluate the performance of new method, 50 samples with E2 concentrations covering the whole measurable interval were selected and assayed. Linearity dilution, LoB (Limit of Blank) and LoD (Limit of Detection) were determined. Results: The new assay showed good total repeatability demonstrated by low CV values, and good linear relationship with respect to current method (R=0.9926) as demonstrated by linear regression. Non-parametric regression showed for the new method a slight constant and proportional systematic error that, however, resulted not significant from difference plot analysis, with a general tendency to overestimate results for the current method. Performances of the new method resulted acceptable within the maximum admissible error derived from the literature, and a good linearity over a wide range of dilutions was showed. LoD value confirmed an amelioration in measurement of low estradiol. Conclusion: In conclusion, sensitive assay is suitable to replace the method used in our laboratory with significant improvement in the measurement of low serum estradiol levels.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Open Biomarkers Journal
Open Biomarkers Journal Medicine-Medicine (miscellaneous)
CiteScore
0.80
自引率
0.00%
发文量
9
期刊介绍: The Open Biomarkers Journal is an Open Access online journal, which publishes original full-length, short research articles and reviews on biomarkers in clinical, medical and pharmaceutical research. The coverage includes biomarkers of disease, new biomarkers, exposure to drugs, genetic effects, and applications of biomarkers. The Open Biomarkers Journal, a peer reviewed journal, aims to provide the most complete and reliable source of information on current developments in the field. The emphasis will be on publishing quality articles rapidly and freely available to researchers worldwide.
期刊最新文献
The Cytoglobin Expression Under Hypoxic Conditions in Covid-19 Cases microRNA-146a: A biomarker for Epileptogenesis, Epilepsy Prognosis, and Treatment Resistance Study of Gender-based Changes in P53 in Preeclampsia A Mini-review on the Effects of (Carbon) Nanoparticles and Oxidative Stress in Animals Diagnostic Accuracy of Red Cell Distribution Width to Platelet Ratio for Detection of Liver Fibrosis Compared with Fibroscan in Chronic Hepatitis B Egyptian patients
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1