{"title":"气球膨胀辅助捕获和成功回收部分断裂和捕获的导丝","authors":"S. Sinha, M. Razi, Awadesh K. Sharma, U. Pandey","doi":"10.4103/jicc.jicc_58_20","DOIUrl":null,"url":null,"abstract":"Guidewire fracture (either partial or complete) is an exceedingly rare complication of which is mostly encountered in percutaneous coronary intervention (PCI) of complex lesions. It can be managed mostly with using percutaneous techniques, but at times requires a surgical approach as well. Here, we report a case of a 67-year-old male who underwent PCI of totally occluded proximal left anterior descending artery. Due to excessive manipulation, tip of Sion blue wire (polymer-jacket guidewire) got trapped into the lesion. To unloop the wire, spring tip got partially fractured. The guiding catheter was pushed over the fractured wire near the fracture point and a 2.5 mm × 10 mm Sapphire semi-compliant balloon (Orbus Neisch, Netherland) was inflated near the tip of guiding catheter at 12 atm pressure, and whole assembly was pulled out to successfully retrieve the partially fractured and retained wire.","PeriodicalId":100789,"journal":{"name":"Journal of Indian College of Cardiology","volume":"11 1","pages":"145 - 147"},"PeriodicalIF":0.0000,"publicationDate":"2021-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Balloon inflation-assisted trapping and successful retrieval of partially fractured and entrapped guidewire\",\"authors\":\"S. Sinha, M. Razi, Awadesh K. Sharma, U. Pandey\",\"doi\":\"10.4103/jicc.jicc_58_20\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Guidewire fracture (either partial or complete) is an exceedingly rare complication of which is mostly encountered in percutaneous coronary intervention (PCI) of complex lesions. It can be managed mostly with using percutaneous techniques, but at times requires a surgical approach as well. Here, we report a case of a 67-year-old male who underwent PCI of totally occluded proximal left anterior descending artery. Due to excessive manipulation, tip of Sion blue wire (polymer-jacket guidewire) got trapped into the lesion. To unloop the wire, spring tip got partially fractured. The guiding catheter was pushed over the fractured wire near the fracture point and a 2.5 mm × 10 mm Sapphire semi-compliant balloon (Orbus Neisch, Netherland) was inflated near the tip of guiding catheter at 12 atm pressure, and whole assembly was pulled out to successfully retrieve the partially fractured and retained wire.\",\"PeriodicalId\":100789,\"journal\":{\"name\":\"Journal of Indian College of Cardiology\",\"volume\":\"11 1\",\"pages\":\"145 - 147\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Indian College of Cardiology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/jicc.jicc_58_20\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Indian College of Cardiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jicc.jicc_58_20","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
导丝骨折(部分或完全)是一种极为罕见的并发症,主要见于复杂病变的经皮冠状动脉介入治疗(PCI)。大多数情况下可以通过经皮技术进行治疗,但有时也需要手术治疗。在此,我们报告一位67岁男性接受PCI治疗完全闭塞的左前降支近端。由于操作过度,锡安蓝丝(聚合物护套导丝)尖端卡在病灶内。为了解开铁丝,弹簧头部分断裂。在骨折点附近将导尿管推过骨折丝,在导尿管尖端附近以12atm的压力充气2.5 mm × 10 mm Sapphire半弯曲球囊(Orbus Neisch,荷兰),并将整个组件拔出,成功取出部分断裂和保留的导丝。
Balloon inflation-assisted trapping and successful retrieval of partially fractured and entrapped guidewire
Guidewire fracture (either partial or complete) is an exceedingly rare complication of which is mostly encountered in percutaneous coronary intervention (PCI) of complex lesions. It can be managed mostly with using percutaneous techniques, but at times requires a surgical approach as well. Here, we report a case of a 67-year-old male who underwent PCI of totally occluded proximal left anterior descending artery. Due to excessive manipulation, tip of Sion blue wire (polymer-jacket guidewire) got trapped into the lesion. To unloop the wire, spring tip got partially fractured. The guiding catheter was pushed over the fractured wire near the fracture point and a 2.5 mm × 10 mm Sapphire semi-compliant balloon (Orbus Neisch, Netherland) was inflated near the tip of guiding catheter at 12 atm pressure, and whole assembly was pulled out to successfully retrieve the partially fractured and retained wire.