基于TSL2561传感器校准热科学genys30可见分光光度计的溶解氨水平测量原型

Titin Putri Aripta, S. W. Suciyati, A. Supriyanto, J. Junaidi
{"title":"基于TSL2561传感器校准热科学genys30可见分光光度计的溶解氨水平测量原型","authors":"Titin Putri Aripta, S. W. Suciyati, A. Supriyanto, J. Junaidi","doi":"10.23960/jemit.v3i3.110","DOIUrl":null,"url":null,"abstract":"In this study, a prototype measuring instrument for dissolved ammonia levels based on the TSL2561 sensor and calibrated thermo scientific Genesys 30 visible spectrophotometers has been realized, which aims to create a system for reading dissolved ammonia levels. This measuring instrument uses a violet LED as a light source, Arduino UNO as the central processor, and an I2C LCD to display measured values. This research was carried out by reading sensor tests on artificial instruments and spectrophotometer with dissolved ammonia samples with levels varying from 0-0.3 mg/l to obtain an equation for converting the absorbance value of the artificial measuring instrument into the dissolved ammonia level value, which was implemented in the Arduino program. Furthermore, an artificial measuring instrument is applied by measuring the dissolved ammonia level in the wastewater sample, namely the wastewater from the shrimp seeds tank, artemia tank, and tilapia tank. This measuring instrument has a measurement range from 0-0.3 mg/l. The sensor test results show that the greater the dissolved ammonia level, the greater the absorbance value. The results of the application of artificial measuring instruments obtained the value of dissolved ammonia levels in the wastewater of the shrimp seeds tank of 0.2811 mg/l, the wastewater of the artemia tank of 0.0672 mg/l, and the wastewater of the tilapia pond at 0.0156 mg/l. Based on the calculation results, it was obtained that the average accuracy and precision for the shrimp seeds tank wastewater was 98.63% and 98.47%, the Artemia tank wastewater was 97.72% and 98.08%, while for the pond wastewater tilapia by 95.71% and 99.74 %.","PeriodicalId":15738,"journal":{"name":"Journal of Energy, Material, and Instrumentation Technology","volume":"21 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Prototype Measuring Levels of Dissolved Ammonia Based on TSL2561 Sensor Calibrated Thermo Scientific Genesys 30 Visible Spectrophotometer\",\"authors\":\"Titin Putri Aripta, S. W. Suciyati, A. Supriyanto, J. Junaidi\",\"doi\":\"10.23960/jemit.v3i3.110\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In this study, a prototype measuring instrument for dissolved ammonia levels based on the TSL2561 sensor and calibrated thermo scientific Genesys 30 visible spectrophotometers has been realized, which aims to create a system for reading dissolved ammonia levels. This measuring instrument uses a violet LED as a light source, Arduino UNO as the central processor, and an I2C LCD to display measured values. This research was carried out by reading sensor tests on artificial instruments and spectrophotometer with dissolved ammonia samples with levels varying from 0-0.3 mg/l to obtain an equation for converting the absorbance value of the artificial measuring instrument into the dissolved ammonia level value, which was implemented in the Arduino program. Furthermore, an artificial measuring instrument is applied by measuring the dissolved ammonia level in the wastewater sample, namely the wastewater from the shrimp seeds tank, artemia tank, and tilapia tank. This measuring instrument has a measurement range from 0-0.3 mg/l. The sensor test results show that the greater the dissolved ammonia level, the greater the absorbance value. The results of the application of artificial measuring instruments obtained the value of dissolved ammonia levels in the wastewater of the shrimp seeds tank of 0.2811 mg/l, the wastewater of the artemia tank of 0.0672 mg/l, and the wastewater of the tilapia pond at 0.0156 mg/l. Based on the calculation results, it was obtained that the average accuracy and precision for the shrimp seeds tank wastewater was 98.63% and 98.47%, the Artemia tank wastewater was 97.72% and 98.08%, while for the pond wastewater tilapia by 95.71% and 99.74 %.\",\"PeriodicalId\":15738,\"journal\":{\"name\":\"Journal of Energy, Material, and Instrumentation Technology\",\"volume\":\"21 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-08-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Energy, Material, and Instrumentation Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.23960/jemit.v3i3.110\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Energy, Material, and Instrumentation Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.23960/jemit.v3i3.110","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在本研究中,基于TSL2561传感器和经过校准的热科学Genesys 30可见分光光度计,实现了溶解氨水平的原型测量仪器,旨在创建一个溶解氨水平的读取系统。该测量仪采用紫光LED作为光源,Arduino UNO作为中央处理器,I2C LCD显示测量值。本研究采用0-0.3 mg/l溶解氨样品在人工仪器和分光光度计上进行读数传感器测试,得到人工测量仪器的吸光度值转换为溶解氨水平值的方程,并在Arduino程序中实现。此外,采用人工测量仪器,测量废水样品中溶解氨的水平,即虾种池、蒿池和罗非鱼池的废水。本仪器测量范围为0-0.3 mg/l。传感器测试结果表明,溶解氨浓度越大,吸光度值越大。人工测量仪器的应用结果表明,虾种池废水中溶解氨含量为0.2811 mg/l,蒿池废水为0.0672 mg/l,罗非鱼池废水为0.0156 mg/l。计算结果表明,对虾种子池废水的平均准确度和精密度分别为98.63%和98.47%,蒿池废水的平均准确度和精密度分别为97.72%和98.08%,罗非鱼池废水的平均准确度和精密度分别为95.71%和99.74%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Prototype Measuring Levels of Dissolved Ammonia Based on TSL2561 Sensor Calibrated Thermo Scientific Genesys 30 Visible Spectrophotometer
In this study, a prototype measuring instrument for dissolved ammonia levels based on the TSL2561 sensor and calibrated thermo scientific Genesys 30 visible spectrophotometers has been realized, which aims to create a system for reading dissolved ammonia levels. This measuring instrument uses a violet LED as a light source, Arduino UNO as the central processor, and an I2C LCD to display measured values. This research was carried out by reading sensor tests on artificial instruments and spectrophotometer with dissolved ammonia samples with levels varying from 0-0.3 mg/l to obtain an equation for converting the absorbance value of the artificial measuring instrument into the dissolved ammonia level value, which was implemented in the Arduino program. Furthermore, an artificial measuring instrument is applied by measuring the dissolved ammonia level in the wastewater sample, namely the wastewater from the shrimp seeds tank, artemia tank, and tilapia tank. This measuring instrument has a measurement range from 0-0.3 mg/l. The sensor test results show that the greater the dissolved ammonia level, the greater the absorbance value. The results of the application of artificial measuring instruments obtained the value of dissolved ammonia levels in the wastewater of the shrimp seeds tank of 0.2811 mg/l, the wastewater of the artemia tank of 0.0672 mg/l, and the wastewater of the tilapia pond at 0.0156 mg/l. Based on the calculation results, it was obtained that the average accuracy and precision for the shrimp seeds tank wastewater was 98.63% and 98.47%, the Artemia tank wastewater was 97.72% and 98.08%, while for the pond wastewater tilapia by 95.71% and 99.74 %.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The Microbial Fuel Cell Characteristics of the PVA/Chitosan Membrane with Variations of Phosphate Acid and Sodium Alginate Derived from Vegetable Waste The Influence of Heatmap Correlation-based Feature Selection on Predictive Modeling of Low Alloy Steel Mechanical Properties Using Artificial Neural Network (ANN) Algorithm Extraction of Nanocellulose from Softwood Pine (Pinus merkusii) using Acid Hydrolysis Method Analysis of the Effects of CO2, H2S Composition, and Temperature on Steam Towards Corrosion Rate in Geothermal Power Plants Design of a Classroom Noise Monitoring Tool Using a KY-037 Sound Sensor Based on Wemos D1R1
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1