Wei Zhao, Gang Feng, Jun Wan, Xing Wang, Jiping Feng, Wanwan Zhang, Yalan Yu, Rui Pu, Yong Du, Yongtong Cao
{"title":"玻璃管和聚对苯二甲酸乙二醇酯管中的EDTA血浆适用于在迈瑞CL-6000i上测量B型钠尿肽。","authors":"Wei Zhao, Gang Feng, Jun Wan, Xing Wang, Jiping Feng, Wanwan Zhang, Yalan Yu, Rui Pu, Yong Du, Yongtong Cao","doi":"10.1177/00045632231162289","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>To explore differences in B-type natriuretic peptide (BNP) concentration and stability and evaluate BNP accuracy in different collection tubes.</p><p><strong>Methods: </strong>BNP concentrations in heparin/glass, EDTA/glass, and EDTA/polyethylene terephthalate (PET) tubes were measured on the Mindray CL-6000i at 0.5, 1, 2, and 4 h after collection. Differences were evaluated using Wilcoxon's paired tests and Bland-Altman plots. BNP stability and measurement accuracies were estimated using Kruskal-Wallis H tests and recovery tests.</p><p><strong>Results: </strong>BNP concentrations in EDTA/glass tubes were 31.4% higher than those in heparin/glass tubes and 3.04% lower than those in EDTA/PET tubes. BNP stability significantly decreased in the heparin/glass tube. BNP remained stable in EDTA/glass and EDTA/PET tubes at room temperature for 4 h. BNP recovery rates in heparin/glass, EDTA/glass, and EDTA/PET tubes were 77.46, 86.04, and 88.23%, respectively.</p><p><strong>Conclusions: </strong>Plasma in EDTA/glass and EDTA/PET tubes is suitable for BNP measurement on the Mindray CL-6000i.</p>","PeriodicalId":8005,"journal":{"name":"Annals of Clinical Biochemistry","volume":" ","pages":"306-312"},"PeriodicalIF":2.1000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"EDTA plasma in glass and polyethylene terephthalate tubes is suitable for measurement of B-type natriuretic peptide on the Mindray CL-6000i.\",\"authors\":\"Wei Zhao, Gang Feng, Jun Wan, Xing Wang, Jiping Feng, Wanwan Zhang, Yalan Yu, Rui Pu, Yong Du, Yongtong Cao\",\"doi\":\"10.1177/00045632231162289\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Aims: </strong>To explore differences in B-type natriuretic peptide (BNP) concentration and stability and evaluate BNP accuracy in different collection tubes.</p><p><strong>Methods: </strong>BNP concentrations in heparin/glass, EDTA/glass, and EDTA/polyethylene terephthalate (PET) tubes were measured on the Mindray CL-6000i at 0.5, 1, 2, and 4 h after collection. Differences were evaluated using Wilcoxon's paired tests and Bland-Altman plots. BNP stability and measurement accuracies were estimated using Kruskal-Wallis H tests and recovery tests.</p><p><strong>Results: </strong>BNP concentrations in EDTA/glass tubes were 31.4% higher than those in heparin/glass tubes and 3.04% lower than those in EDTA/PET tubes. BNP stability significantly decreased in the heparin/glass tube. BNP remained stable in EDTA/glass and EDTA/PET tubes at room temperature for 4 h. BNP recovery rates in heparin/glass, EDTA/glass, and EDTA/PET tubes were 77.46, 86.04, and 88.23%, respectively.</p><p><strong>Conclusions: </strong>Plasma in EDTA/glass and EDTA/PET tubes is suitable for BNP measurement on the Mindray CL-6000i.</p>\",\"PeriodicalId\":8005,\"journal\":{\"name\":\"Annals of Clinical Biochemistry\",\"volume\":\" \",\"pages\":\"306-312\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2023-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annals of Clinical Biochemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1177/00045632231162289\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/3/14 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Clinical Biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/00045632231162289","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/3/14 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
EDTA plasma in glass and polyethylene terephthalate tubes is suitable for measurement of B-type natriuretic peptide on the Mindray CL-6000i.
Aims: To explore differences in B-type natriuretic peptide (BNP) concentration and stability and evaluate BNP accuracy in different collection tubes.
Methods: BNP concentrations in heparin/glass, EDTA/glass, and EDTA/polyethylene terephthalate (PET) tubes were measured on the Mindray CL-6000i at 0.5, 1, 2, and 4 h after collection. Differences were evaluated using Wilcoxon's paired tests and Bland-Altman plots. BNP stability and measurement accuracies were estimated using Kruskal-Wallis H tests and recovery tests.
Results: BNP concentrations in EDTA/glass tubes were 31.4% higher than those in heparin/glass tubes and 3.04% lower than those in EDTA/PET tubes. BNP stability significantly decreased in the heparin/glass tube. BNP remained stable in EDTA/glass and EDTA/PET tubes at room temperature for 4 h. BNP recovery rates in heparin/glass, EDTA/glass, and EDTA/PET tubes were 77.46, 86.04, and 88.23%, respectively.
Conclusions: Plasma in EDTA/glass and EDTA/PET tubes is suitable for BNP measurement on the Mindray CL-6000i.
期刊介绍:
Annals of Clinical Biochemistry is the fully peer reviewed international journal of the Association for Clinical Biochemistry and Laboratory Medicine.
Annals of Clinical Biochemistry accepts papers that contribute to knowledge in all fields of laboratory medicine, especially those pertaining to the understanding, diagnosis and treatment of human disease. It publishes papers on clinical biochemistry, clinical audit, metabolic medicine, immunology, genetics, biotechnology, haematology, microbiology, computing and management where they have both biochemical and clinical relevance. Papers describing evaluation or implementation of commercial reagent kits or the performance of new analysers require substantial original information. Unless of exceptional interest and novelty, studies dealing with the redox status in various diseases are not generally considered within the journal''s scope. Studies documenting the association of single nucleotide polymorphisms (SNPs) with particular phenotypes will not normally be considered, given the greater strength of genome wide association studies (GWAS). Research undertaken in non-human animals will not be considered for publication in the Annals.