Michael Günther , Leonie Schuster , Christian Boßelmann , Holger Lerche , Ulf Ziemann , Katharina Feil , Justus Marquetand
{"title":"海绵脑电图在信号质量上与常规脑电图相当,但速度比常规脑电图快","authors":"Michael Günther , Leonie Schuster , Christian Boßelmann , Holger Lerche , Ulf Ziemann , Katharina Feil , Justus Marquetand","doi":"10.1016/j.cnp.2023.03.002","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>Emergency diagnostics, such as acquisition of an electroencephalogram (EEG), are of great diagnostic importance, but there is often a lack of experienced personnel. Wet active electrode sponge-based electroencephalogram (sp-EEG) systems can be applied rapidly and by inexperienced personnel. This makes them an attractive alternative to routine EEG (r-EEG) systems in these settings. Here, we examined the feasibility and signal quality of sp-EEG compared to r-EEG.</p></div><div><h3>Methods</h3><p>In this case-control, single-blind, non-randomized study, EEG recordings using a sp- and a r-EEG system were performed in 18 individuals with a variety of epileptiform discharges and 11 healthy control subjects. The time was stopped until all electrodes in both systems displayed adequate skin-electrode impedances. The resulting 58 EEGs were visually inspected by 7 experienced, blinded neurologists. Raters were asked to score physiological and pathological graphoelements, and to distinguish between the different systems by visual inspection of the EEGs.</p></div><div><h3>Results</h3><p>Time to signal acquisition for sp-EEG was significantly faster (4.8 min (SD 2.01) vs. r-EEG 13.3 min (SD 2.72), p < 0.001). All physiological and pathological graphoelements of all 58 EEGs could be identified. Raters were unable to distinguish between sp-EEG or r-EEG based on visual inspection of the EEGs alone.</p></div><div><h3>Conclusions</h3><p>Sp-EEG represents a feasible alternative to r-EEG in emergency diagnostics or resource-limited settings.</p></div><div><h3>Significance</h3><p>Given shortage of trained personnel or resources, the easy implementation and comparable quality of a novel sp-EEG system may increase general availability of EEG and thus improve patient care.</p></div>","PeriodicalId":45697,"journal":{"name":"Clinical Neurophysiology Practice","volume":"8 ","pages":"Pages 58-64"},"PeriodicalIF":2.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10074306/pdf/main.pdf","citationCount":"0","resultStr":"{\"title\":\"Sponge EEG is equivalent regarding signal quality, but faster than routine EEG\",\"authors\":\"Michael Günther , Leonie Schuster , Christian Boßelmann , Holger Lerche , Ulf Ziemann , Katharina Feil , Justus Marquetand\",\"doi\":\"10.1016/j.cnp.2023.03.002\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><p>Emergency diagnostics, such as acquisition of an electroencephalogram (EEG), are of great diagnostic importance, but there is often a lack of experienced personnel. Wet active electrode sponge-based electroencephalogram (sp-EEG) systems can be applied rapidly and by inexperienced personnel. This makes them an attractive alternative to routine EEG (r-EEG) systems in these settings. Here, we examined the feasibility and signal quality of sp-EEG compared to r-EEG.</p></div><div><h3>Methods</h3><p>In this case-control, single-blind, non-randomized study, EEG recordings using a sp- and a r-EEG system were performed in 18 individuals with a variety of epileptiform discharges and 11 healthy control subjects. The time was stopped until all electrodes in both systems displayed adequate skin-electrode impedances. The resulting 58 EEGs were visually inspected by 7 experienced, blinded neurologists. Raters were asked to score physiological and pathological graphoelements, and to distinguish between the different systems by visual inspection of the EEGs.</p></div><div><h3>Results</h3><p>Time to signal acquisition for sp-EEG was significantly faster (4.8 min (SD 2.01) vs. r-EEG 13.3 min (SD 2.72), p < 0.001). All physiological and pathological graphoelements of all 58 EEGs could be identified. Raters were unable to distinguish between sp-EEG or r-EEG based on visual inspection of the EEGs alone.</p></div><div><h3>Conclusions</h3><p>Sp-EEG represents a feasible alternative to r-EEG in emergency diagnostics or resource-limited settings.</p></div><div><h3>Significance</h3><p>Given shortage of trained personnel or resources, the easy implementation and comparable quality of a novel sp-EEG system may increase general availability of EEG and thus improve patient care.</p></div>\",\"PeriodicalId\":45697,\"journal\":{\"name\":\"Clinical Neurophysiology Practice\",\"volume\":\"8 \",\"pages\":\"Pages 58-64\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2023-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10074306/pdf/main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical Neurophysiology Practice\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2467981X23000045\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"NEUROSCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Neurophysiology Practice","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2467981X23000045","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"NEUROSCIENCES","Score":null,"Total":0}
Sponge EEG is equivalent regarding signal quality, but faster than routine EEG
Objective
Emergency diagnostics, such as acquisition of an electroencephalogram (EEG), are of great diagnostic importance, but there is often a lack of experienced personnel. Wet active electrode sponge-based electroencephalogram (sp-EEG) systems can be applied rapidly and by inexperienced personnel. This makes them an attractive alternative to routine EEG (r-EEG) systems in these settings. Here, we examined the feasibility and signal quality of sp-EEG compared to r-EEG.
Methods
In this case-control, single-blind, non-randomized study, EEG recordings using a sp- and a r-EEG system were performed in 18 individuals with a variety of epileptiform discharges and 11 healthy control subjects. The time was stopped until all electrodes in both systems displayed adequate skin-electrode impedances. The resulting 58 EEGs were visually inspected by 7 experienced, blinded neurologists. Raters were asked to score physiological and pathological graphoelements, and to distinguish between the different systems by visual inspection of the EEGs.
Results
Time to signal acquisition for sp-EEG was significantly faster (4.8 min (SD 2.01) vs. r-EEG 13.3 min (SD 2.72), p < 0.001). All physiological and pathological graphoelements of all 58 EEGs could be identified. Raters were unable to distinguish between sp-EEG or r-EEG based on visual inspection of the EEGs alone.
Conclusions
Sp-EEG represents a feasible alternative to r-EEG in emergency diagnostics or resource-limited settings.
Significance
Given shortage of trained personnel or resources, the easy implementation and comparable quality of a novel sp-EEG system may increase general availability of EEG and thus improve patient care.
期刊介绍:
Clinical Neurophysiology Practice (CNP) is a new Open Access journal that focuses on clinical practice issues in clinical neurophysiology including relevant new research, case reports or clinical series, normal values and didactic reviews. It is an official journal of the International Federation of Clinical Neurophysiology and complements Clinical Neurophysiology which focuses on innovative research in the specialty. It has a role in supporting established clinical practice, and an educational role for trainees, technicians and practitioners.