{"title":"可穿戴式多普勒超声测量颈动脉校正血流时间准确检测动态志愿者被动抬腿时脑卒中容量的变化。","authors":"Jon-Émile S Kenny,&nbsp;Christine Horner,&nbsp;Mai Elfarnawany,&nbsp;Andrew M Eibl,&nbsp;Joseph K Eibl","doi":"10.31083/j.fbe1502012","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The change in the corrected flow time of the common carotid artery (ccFTΔ) has been used as a surrogate of changing stroke volume (SVΔ) in the critically-ill. Thus, this relatively easy-to-obtain Doppler measure may help clinicians better define the intended effect of intravenous fluids. Yet the temporal evolution of SVΔ and ccFTΔ has not been reported in volunteers undergoing a passive leg raise (PLR).</p><p><strong>Methods: </strong>We recruited clinically-euvolemic, non-fasted, adult, volunteers in a local physiology lab to perform 2 PLR maneuvers, each separated by a 5 minute 'wash-out'. During each PLR, SV was measured by a non-invasive pulse contour analysis device. SV was temporally-synchronized with a wireless, wearable Doppler ultrasound worn over the common carotid artery that continuously measured ccFT.</p><p><strong>Results: </strong>36 PLR maneuvers were obtained across 19 ambulatory volunteers. 8856 carotid Doppler cardiac cycles were analyzed. The ccFT increased nearly ubiquitously during the PLR and within 40-60 seconds of PLR onset; the rise in SV from the pulse contour device was more gradual. SVΔ by +5% and +10% were both detected by a +7% ccFTΔ with sensitivities, specificities and areas under the receiver operator curve of 59%, 95% and 0.77 (<i>p</i> < 0.001) and 66%, 76% and 0.73 (<i>p</i> < 0.001), respectively.</p><p><strong>Conclusions: </strong>The ccFTΔ during the PLR in ambulatory volunteers was rapid and sustained. Within the limits of precision for detecting a clinically-significant rise in SV by a non-invasive pulse contour analysis device, simultaneously-acquired ccFT from a wireless, wearable ultrasound system was accurate at detecting 'preload responsiveness'.</p>","PeriodicalId":73068,"journal":{"name":"Frontiers in bioscience (Elite edition)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-05-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Accurately Detects Changing Stroke Volume During the Passive Leg Raise in Ambulatory Volunteers.\",\"authors\":\"Jon-Émile S Kenny,&nbsp;Christine Horner,&nbsp;Mai Elfarnawany,&nbsp;Andrew M Eibl,&nbsp;Joseph K Eibl\",\"doi\":\"10.31083/j.fbe1502012\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The change in the corrected flow time of the common carotid artery (ccFTΔ) has been used as a surrogate of changing stroke volume (SVΔ) in the critically-ill. Thus, this relatively easy-to-obtain Doppler measure may help clinicians better define the intended effect of intravenous fluids. Yet the temporal evolution of SVΔ and ccFTΔ has not been reported in volunteers undergoing a passive leg raise (PLR).</p><p><strong>Methods: </strong>We recruited clinically-euvolemic, non-fasted, adult, volunteers in a local physiology lab to perform 2 PLR maneuvers, each separated by a 5 minute 'wash-out'. During each PLR, SV was measured by a non-invasive pulse contour analysis device. SV was temporally-synchronized with a wireless, wearable Doppler ultrasound worn over the common carotid artery that continuously measured ccFT.</p><p><strong>Results: </strong>36 PLR maneuvers were obtained across 19 ambulatory volunteers. 8856 carotid Doppler cardiac cycles were analyzed. The ccFT increased nearly ubiquitously during the PLR and within 40-60 seconds of PLR onset; the rise in SV from the pulse contour device was more gradual. SVΔ by +5% and +10% were both detected by a +7% ccFTΔ with sensitivities, specificities and areas under the receiver operator curve of 59%, 95% and 0.77 (<i>p</i> < 0.001) and 66%, 76% and 0.73 (<i>p</i> < 0.001), respectively.</p><p><strong>Conclusions: </strong>The ccFTΔ during the PLR in ambulatory volunteers was rapid and sustained. Within the limits of precision for detecting a clinically-significant rise in SV by a non-invasive pulse contour analysis device, simultaneously-acquired ccFT from a wireless, wearable ultrasound system was accurate at detecting 'preload responsiveness'.</p>\",\"PeriodicalId\":73068,\"journal\":{\"name\":\"Frontiers in bioscience (Elite edition)\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-05-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Frontiers in bioscience (Elite edition)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.31083/j.fbe1502012\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in bioscience (Elite edition)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31083/j.fbe1502012","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

背景:在危重患者中,颈总动脉校正血流时间(ccFTΔ)的变化已被用作改变卒中容量(SVΔ)的替代指标。因此,这种相对容易获得的多普勒测量可以帮助临床医生更好地确定静脉输液的预期效果。然而,在接受被动抬腿(PLR)的志愿者中,SVΔ和ccFTΔ的时间进化尚未见报道。方法:我们在当地的生理实验室招募了临床大容量、非禁食的成年志愿者,进行2次PLR动作,每次间隔5分钟的“洗脱期”。在每次PLR期间,通过无创脉冲轮廓分析装置测量SV。SV与戴在颈总动脉上的无线可穿戴多普勒超声同步,连续测量ccFT。结果:19名非卧床志愿者共完成36个PLR动作。分析8856例颈动脉多普勒心动周期。在PLR期间和PLR发生后40-60秒内,ccFT几乎无处不在地增加;脉冲轮廓装置的SV上升更为平缓。SVΔ +5%和+10%的检出率均为+7% ccFTΔ,灵敏度、特异度和受体操作者曲线下面积分别为59%、95%和0.77 (p < 0.001)和66%、76%和0.73 (p < 0.001)。结论:在非卧床志愿者的PLR过程中ccFTΔ是快速且持续的。在通过无创脉冲轮廓分析设备检测临床显著的SV升高的精度范围内,从无线可穿戴超声系统同时获取的ccFT在检测“预加载响应性”方面是准确的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Accurately Detects Changing Stroke Volume During the Passive Leg Raise in Ambulatory Volunteers.

Background: The change in the corrected flow time of the common carotid artery (ccFTΔ) has been used as a surrogate of changing stroke volume (SVΔ) in the critically-ill. Thus, this relatively easy-to-obtain Doppler measure may help clinicians better define the intended effect of intravenous fluids. Yet the temporal evolution of SVΔ and ccFTΔ has not been reported in volunteers undergoing a passive leg raise (PLR).

Methods: We recruited clinically-euvolemic, non-fasted, adult, volunteers in a local physiology lab to perform 2 PLR maneuvers, each separated by a 5 minute 'wash-out'. During each PLR, SV was measured by a non-invasive pulse contour analysis device. SV was temporally-synchronized with a wireless, wearable Doppler ultrasound worn over the common carotid artery that continuously measured ccFT.

Results: 36 PLR maneuvers were obtained across 19 ambulatory volunteers. 8856 carotid Doppler cardiac cycles were analyzed. The ccFT increased nearly ubiquitously during the PLR and within 40-60 seconds of PLR onset; the rise in SV from the pulse contour device was more gradual. SVΔ by +5% and +10% were both detected by a +7% ccFTΔ with sensitivities, specificities and areas under the receiver operator curve of 59%, 95% and 0.77 (p < 0.001) and 66%, 76% and 0.73 (p < 0.001), respectively.

Conclusions: The ccFTΔ during the PLR in ambulatory volunteers was rapid and sustained. Within the limits of precision for detecting a clinically-significant rise in SV by a non-invasive pulse contour analysis device, simultaneously-acquired ccFT from a wireless, wearable ultrasound system was accurate at detecting 'preload responsiveness'.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Quercetin, the Potential Powerful Flavonoid for Human and Food: A Review. Streptomyces as a Novel Biotool for Azo Pigments Remediation in Contaminated Scenarios. Biotechnological Advances Utilizing Aptamers and Peptides Refining PD-L1 Targeting. Obtaining Melanin-Synthesizing Strains of Bacillus thuringiensis and their Use for Biological Preparations. Implementation of a Macroporous Polyhydroxyethylmethacrylate Cryogel-Based Mini-Bioreactor System to Improve Monoclonal Antibody Production.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1