{"title":"End-to-end testing for stereotactic radiotherapy including the development of a Multi-Modality phantom","authors":"","doi":"10.1016/j.zemedi.2022.11.006","DOIUrl":null,"url":null,"abstract":"<div><h3><strong>Purpose</strong></h3><p>A new insert for a commercially available end-to-end test phantom was designed and in-house manufactured by 3D printing. Subsequently, the insert was tested for different stereotactic radiation therapy workflows (SRS, SBRT, FSRT, and Multimet) also in comparison to the original insert.</p></div><div><h3><strong>Material and methods</strong></h3><p>Workflows contained imaging (MR, CT), treatment planning, positioning, and irradiation. Positioning accuracy was evaluated for non-coplanar x-ray, kV- and MV-CBCT systems, as well as surface guided radiation therapy. Dosimetric accuracy of the irradiation was measured with an ionization chamber at four different linear accelerators including dynamic tumor tracking for SBRT.</p></div><div><h3><strong>Results</strong></h3><p>CT parameters of the insert were within the specification. For MR images, the new insert allowed quantitative analysis of the MR distortion. Positioning accuracy of the phantom with the new insert using the imaging systems of the different linacs was < 1 mm/degree also for MV-CBCT and a non-coplanar imaging system which caused > 3 mm deviation with the original insert. Deviation of point dose values was <<!--> <!-->3% for SRS, FSRT, and SBRT for both inserts. For the Multimet plans deviations exceeded 10% because the ionization chamber was not positioned in each metastasis, but in the center of phantom and treatment plan.</p></div><div><h3><strong>Conclusion</strong></h3><p>The in-house manufactured insert performed well in all steps of four stereotactic treatment end-to-end tests. Advantages over the commercially available alternative were seen for quantitative analysis of deformation correction in MR images, applicability for non-coplanar x-ray imaging, and dynamic tumor tracking.</p></div>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":"34 3","pages":"Pages 477-484"},"PeriodicalIF":2.4000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0939388922001234/pdfft?md5=ddec9b262290ef8f0ca939378a7e3a7f&pid=1-s2.0-S0939388922001234-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zeitschrift fur Medizinische Physik","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0939388922001234","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
A new insert for a commercially available end-to-end test phantom was designed and in-house manufactured by 3D printing. Subsequently, the insert was tested for different stereotactic radiation therapy workflows (SRS, SBRT, FSRT, and Multimet) also in comparison to the original insert.
Material and methods
Workflows contained imaging (MR, CT), treatment planning, positioning, and irradiation. Positioning accuracy was evaluated for non-coplanar x-ray, kV- and MV-CBCT systems, as well as surface guided radiation therapy. Dosimetric accuracy of the irradiation was measured with an ionization chamber at four different linear accelerators including dynamic tumor tracking for SBRT.
Results
CT parameters of the insert were within the specification. For MR images, the new insert allowed quantitative analysis of the MR distortion. Positioning accuracy of the phantom with the new insert using the imaging systems of the different linacs was < 1 mm/degree also for MV-CBCT and a non-coplanar imaging system which caused > 3 mm deviation with the original insert. Deviation of point dose values was < 3% for SRS, FSRT, and SBRT for both inserts. For the Multimet plans deviations exceeded 10% because the ionization chamber was not positioned in each metastasis, but in the center of phantom and treatment plan.
Conclusion
The in-house manufactured insert performed well in all steps of four stereotactic treatment end-to-end tests. Advantages over the commercially available alternative were seen for quantitative analysis of deformation correction in MR images, applicability for non-coplanar x-ray imaging, and dynamic tumor tracking.
期刊介绍:
Zeitschrift fur Medizinische Physik (Journal of Medical Physics) is an official organ of the German and Austrian Society of Medical Physic and the Swiss Society of Radiobiology and Medical Physics.The Journal is a platform for basic research and practical applications of physical procedures in medical diagnostics and therapy. The articles are reviewed following international standards of peer reviewing.
Focuses of the articles are:
-Biophysical methods in radiation therapy and nuclear medicine
-Dosimetry and radiation protection
-Radiological diagnostics and quality assurance
-Modern imaging techniques, such as computed tomography, magnetic resonance imaging, positron emission tomography
-Ultrasonography diagnostics, application of laser and UV rays
-Electronic processing of biosignals
-Artificial intelligence and machine learning in medical physics
In the Journal, the latest scientific insights find their expression in the form of original articles, reviews, technical communications, and information for the clinical practice.