首页 > 最新文献

Zeitschrift fur Medizinische Physik最新文献

英文 中文
Determination of the intraocular irradiance and potential retinal hazards at various positions in the eye during transscleral equatorial illumination for different applied pressures. 在不同的加压条件下,测定经巩膜赤道照明时眼球内不同位置的辐照度和潜在的视网膜危害。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2022-12-10 DOI: 10.1016/j.zemedi.2022.11.005
Nicole Fehler, Christian Lingenfelder, Sebastian Kupferschmid, Martin Hessling

Purpose: With diaphanoscopic illumination of the eye, the intensity of light entering its interior depends on the transmission properties of the eyewall. Light that passes through the eyewall can cause damage to the retina. Therefore, in this study, the intraocular irradiances are determined at different positions on the retina, directly behind the illuminated eyewall, the opposite eyewall and near the macula of ex-vivo porcine eyes. These irradiances are examined for their dependence on the pressure applied on the eyewall with the illuminating fiber and for the influence of the pigmentation of the eye.

Methods: In total 221 ex-vivo porcine eyes were investigated. For transscleral illumination an illumination fiber with a diffusing adapter cap is pressed against the equatorial eyewall. The illumination fiber is pressed onto the eye and the pressure is measured in the anterior chamber. Three different pressures are applied, 23, 78 and 132 mmHg. A detection fiber with diffusing fiber tip is inserted into the eye at the desired position. The eyes were divided in groups with high and less pigmentation to investigate the influence of the pigmentation on the intraocular irradiance.

Results: The intraocular irradiances Eintra increases for various increasing applied pressures with the illumination fiber on the eyewall and for various positions inside the eye. With this the irradiances weighted with the photochemical and thermal hazard weighting function, EA-R and EVIR-R, also increases. Differences in Eintra, EA-R and EVIR-R could be found for different pigmented eyes as these values are higher for less pigmented eyes than for strong pigmented ones.

Conclusion: The hazard to the retina during diaphanoscopic illumination of the eye depends on how strong the surgeon presses the illumination fiber on the eyewall. Depending on the applied pressure and the measuring position in the eye, the specified limit for the photochemical hazard to the retina is partly exceeded. The pigmentation of the eye also plays a role. The irradiance in less pigmented eyes appears to be higher than in strongly pigmented eyes. Because of this, the surgeon should be able to adjust the intensity of the light source to the color of the patient's eye.

目的:在对眼睛进行透射照明时,进入眼睛内部的光线强度取决于眼球壁的透射特性。穿过眼球壁的光线会对视网膜造成损伤。因此,本研究在视网膜的不同位置测定了眼内辐照度,这些位置分别位于被照射眼球的正后方、对侧眼球和活体猪眼的黄斑附近。这些辐照度与照明光纤对眼球施加的压力有关,也受眼球色素的影响:方法:总共研究了 221 只活体猪眼睛。经巩膜照明时,将带有扩散适配帽的照明光纤压在赤道眼球壁上。将照明光纤压在眼球上,测量前房的压力。采用三种不同的压力,分别为 23、78 和 132 毫米汞柱。在所需位置将带有扩散光纤尖端的检测光纤插入眼球。将眼球分为色素沉着程度高和色素沉着程度低两组,以研究色素沉着对眼内辐照度的影响:结果:随着眼球壁上的照明纤维和眼球内的不同位置,眼内辐照度 Eintra 随施加的压力不同而增加。与此同时,使用光化学和热危害加权函数加权的辐照度 EA-R 和 EVIR-R 也会增加。不同色素眼的 Eintra、EA-R 和 EVIR-R 值存在差异,因为色素较少的眼比色素较多的眼的 Eintra、EA-R 和 EVIR-R 值要高:结论:眼球透视照明对视网膜的危害取决于外科医生将照明纤维压在眼球壁上的力度。根据所施加的压力和在眼球中的测量位置,视网膜所受的光化学危害会部分超出规定限值。眼睛的色素沉着也有影响。色素较少的眼睛的辐照度似乎高于色素较多的眼睛。因此,外科医生应该能够根据患者眼睛的颜色调整光源的强度。
{"title":"Determination of the intraocular irradiance and potential retinal hazards at various positions in the eye during transscleral equatorial illumination for different applied pressures.","authors":"Nicole Fehler, Christian Lingenfelder, Sebastian Kupferschmid, Martin Hessling","doi":"10.1016/j.zemedi.2022.11.005","DOIUrl":"10.1016/j.zemedi.2022.11.005","url":null,"abstract":"<p><strong>Purpose: </strong>With diaphanoscopic illumination of the eye, the intensity of light entering its interior depends on the transmission properties of the eyewall. Light that passes through the eyewall can cause damage to the retina. Therefore, in this study, the intraocular irradiances are determined at different positions on the retina, directly behind the illuminated eyewall, the opposite eyewall and near the macula of ex-vivo porcine eyes. These irradiances are examined for their dependence on the pressure applied on the eyewall with the illuminating fiber and for the influence of the pigmentation of the eye.</p><p><strong>Methods: </strong>In total 221 ex-vivo porcine eyes were investigated. For transscleral illumination an illumination fiber with a diffusing adapter cap is pressed against the equatorial eyewall. The illumination fiber is pressed onto the eye and the pressure is measured in the anterior chamber. Three different pressures are applied, 23, 78 and 132 mmHg. A detection fiber with diffusing fiber tip is inserted into the eye at the desired position. The eyes were divided in groups with high and less pigmentation to investigate the influence of the pigmentation on the intraocular irradiance.</p><p><strong>Results: </strong>The intraocular irradiances E<sub>intra</sub> increases for various increasing applied pressures with the illumination fiber on the eyewall and for various positions inside the eye. With this the irradiances weighted with the photochemical and thermal hazard weighting function, E<sub>A-R</sub> and E<sub>VIR-R</sub>, also increases. Differences in E<sub>intra</sub>, E<sub>A-R</sub> and E<sub>VIR-R</sub> could be found for different pigmented eyes as these values are higher for less pigmented eyes than for strong pigmented ones.</p><p><strong>Conclusion: </strong>The hazard to the retina during diaphanoscopic illumination of the eye depends on how strong the surgeon presses the illumination fiber on the eyewall. Depending on the applied pressure and the measuring position in the eye, the specified limit for the photochemical hazard to the retina is partly exceeded. The pigmentation of the eye also plays a role. The irradiance in less pigmented eyes appears to be higher than in strongly pigmented eyes. Because of this, the surgeon should be able to adjust the intensity of the light source to the color of the patient's eye.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"610-619"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"10338500","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Considerations of a thick lens formula for intraocular lens power calculation. 厚晶状体眼内透镜功率计算公式的考虑因素。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2022-12-19 DOI: 10.1016/j.zemedi.2022.11.007
Achim Langenbucher, Peter Hoffmann, Alan Cayless, Damien Gatinel, Guillaume Debellemanière, Jascha Wendelstein, Nóra Szentmáry

Background: In recent years, some lens manufacturers have committed to providing lens shape data for some of their lens models. The purpose of this study is to present a strategy for prediction of intraocular lens power and residual refraction based on a pseudophakic model eye containing 5 refractive surfaces and to show its applicability using worked examples.

Methods: A pseudophakic model eye with a thin spectacle correction, a thick cornea (radius of curvatures for both surfaces and central thickness) and a thick IOL (either radius of curvatures RLa and RLp for front and back surface or equivalent power PL and Coddington factor CL; and either central thickness LT or edge thickness and optic diameter) was set up. Calculations were performed based on linear Gaussian optics (vergence formulae). Formulae were provided to derive the lens power/shape and the residual equivalent spectacle refraction SEQ. From the lens shape the location of the haptic plane HP, the image sided principal plane of the lens HL, and the ocular magnification OM were extracted.

Results: The calculation of a thick intraocular lens and the prediction of residual refraction is presented with reference to 3 working examples: A) lens varied in PL and shifted with its haptic plane keeping the CL constant, B) lens varied in CL and shifted with its haptic plane keeping PL constant, and C) CL and PL of the lens varied keeping its haptic plane position in the eye constant. For each combination of parameters (PL, CL, or haptic plane shift) the parameters influencing SEQ, OM and HL-HP were analysed.

Conclusion: Some modern optical biometers currently on the market provide the radii of curvature of both corneal surface and all relevant distances in the eye. With additional data on the lens shape, it would be possible to improve lens power calculations by switching from thin to thick lens models for the cornea and for the lens. This would overcome one of the major drawbacks of current lens power calculation methods.

背景:近年来,一些晶状体制造商承诺为其部分晶状体模型提供晶状体形状数据。本研究的目的是根据包含 5 个屈光面的假性视网膜模型眼,提出一种预测眼内透镜功率和残余屈光度的策略,并通过工作实例说明其适用性:方法: 建立了一个假性角膜模型眼,该模型眼具有薄眼镜矫正、厚角膜(两个表面的曲率半径和中心厚度)和厚人工晶体(前后表面的曲率半径 RLa 和 RLp 或等效功率 PL 和科丁顿因子 CL;以及中心厚度 LT 或边缘厚度和光学直径)。计算基于线性高斯光学(会聚公式)。计算公式可得出镜片功率/形状和残余等效眼镜折射率 SEQ。从晶状体形状中提取出触觉平面 HP、晶状体像侧主平面 HL 和眼球放大率 OM 的位置:结果:通过 3 个工作实例介绍了厚眼内透镜的计算和残余屈光度的预测:A) 在保持 CL 不变的情况下,改变晶状体的 PL 并随其触觉平面移动;B) 在保持 PL 不变的情况下,改变晶状体的 CL 并随其触觉平面移动;C) 在保持其触觉平面在眼球中的位置不变的情况下,改变晶状体的 CL 和 PL。对每种参数组合(PL、CL 或触觉平面移动)影响 SEQ、OM 和 HL-HP 的参数进行了分析:结论:目前市场上的一些现代光学生物测量仪可提供角膜表面的曲率半径和眼内所有相关距离。有了关于晶状体形状的额外数据,就有可能通过将角膜和晶状体的薄晶状体模型转换为厚晶状体模型来改进晶状体功率计算。这将克服目前镜片功率计算方法的一个主要缺点。
{"title":"Considerations of a thick lens formula for intraocular lens power calculation.","authors":"Achim Langenbucher, Peter Hoffmann, Alan Cayless, Damien Gatinel, Guillaume Debellemanière, Jascha Wendelstein, Nóra Szentmáry","doi":"10.1016/j.zemedi.2022.11.007","DOIUrl":"10.1016/j.zemedi.2022.11.007","url":null,"abstract":"<p><strong>Background: </strong>In recent years, some lens manufacturers have committed to providing lens shape data for some of their lens models. The purpose of this study is to present a strategy for prediction of intraocular lens power and residual refraction based on a pseudophakic model eye containing 5 refractive surfaces and to show its applicability using worked examples.</p><p><strong>Methods: </strong>A pseudophakic model eye with a thin spectacle correction, a thick cornea (radius of curvatures for both surfaces and central thickness) and a thick IOL (either radius of curvatures RLa and RLp for front and back surface or equivalent power PL and Coddington factor CL; and either central thickness LT or edge thickness and optic diameter) was set up. Calculations were performed based on linear Gaussian optics (vergence formulae). Formulae were provided to derive the lens power/shape and the residual equivalent spectacle refraction SEQ. From the lens shape the location of the haptic plane HP, the image sided principal plane of the lens HL, and the ocular magnification OM were extracted.</p><p><strong>Results: </strong>The calculation of a thick intraocular lens and the prediction of residual refraction is presented with reference to 3 working examples: A) lens varied in PL and shifted with its haptic plane keeping the CL constant, B) lens varied in CL and shifted with its haptic plane keeping PL constant, and C) CL and PL of the lens varied keeping its haptic plane position in the eye constant. For each combination of parameters (PL, CL, or haptic plane shift) the parameters influencing SEQ, OM and HL-HP were analysed.</p><p><strong>Conclusion: </strong>Some modern optical biometers currently on the market provide the radii of curvature of both corneal surface and all relevant distances in the eye. With additional data on the lens shape, it would be possible to improve lens power calculations by switching from thin to thick lens models for the cornea and for the lens. This would overcome one of the major drawbacks of current lens power calculation methods.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"620-631"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"10764181","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Heterogeneity of absorbed dose distribution in kidney tissues and dose-response modelling of nephrotoxicity in radiopharmaceutical therapy with beta-particle emitters: A review. 肾脏组织吸收剂量分布的异质性和β粒子发射体放射性药物治疗中肾毒性的剂量反应模型:综述。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-04-06 DOI: 10.1016/j.zemedi.2023.02.006
Clarita Saldarriaga Vargas, Michelle Andersson, Céline Bouvier-Capely, Wei Bo Li, Balázs Madas, Peter Covens, Lara Struelens, Lidia Strigari

Absorbed dose heterogeneity in kidney tissues is an important issue in radiopharmaceutical therapy. The effect of absorbed dose heterogeneity in nephrotoxicity is, however, not fully understood yet, which hampers the implementation of treatment optimization by obscuring the interpretation of clinical response data and the selection of optimal treatment options. Although some dosimetry methods have been developed for kidney dosimetry to the level of microscopic renal substructures, the clinical assessment of the microscopic distribution of radiopharmaceuticals in kidney tissues currently remains a challenge. This restricts the anatomical resolution of clinical dosimetry, which hinders a thorough clinical investigation of the impact of absorbed dose heterogeneity. The potential of absorbed dose-response modelling to support individual treatment optimization in radiopharmaceutical therapy is recognized and gaining attraction. However, biophysical modelling is currently underexplored for the kidney, where particular modelling challenges arise from the convolution of a complex functional organization of renal tissues with the function-mediated dose distribution of radiopharmaceuticals. This article reviews and discusses the heterogeneity of absorbed dose distribution in kidney tissues and the absorbed dose-response modelling of nephrotoxicity in radiopharmaceutical therapy. The review focuses mainly on the peptide receptor radionuclide therapy with beta-particle emitting somatostatin analogues, for which the scientific literature reflects over two decades of clinical experience. Additionally, detailed research perspectives are proposed to address various identified challenges to progress in this field.

肾脏组织的吸收剂量异质性是放射性药物治疗中的一个重要问题。然而,人们对吸收剂量异质性对肾毒性的影响尚未完全了解,这就妨碍了临床反应数据的解读和最佳治疗方案的选择,从而阻碍了治疗优化的实施。虽然目前已开发出一些剂量测定方法,可将肾脏剂量测定到微观肾脏亚结构的水平,但目前临床评估放射性药物在肾脏组织中的微观分布仍是一项挑战。这限制了临床剂量测定的解剖分辨率,妨碍了对吸收剂量异质性影响的全面临床研究。在放射性药物治疗中,吸收剂量-反应建模在支持个体治疗优化方面的潜力已得到认可,并越来越具有吸引力。然而,目前对肾脏的生物物理建模尚未充分开发,肾脏组织的复杂功能组织与放射性药物的功能介导剂量分布之间的卷积关系给建模带来了特殊的挑战。本文回顾并讨论了肾脏组织吸收剂量分布的异质性以及放射性药物治疗中肾毒性的吸收剂量-反应建模。该综述主要侧重于使用发射β粒子的体生长激素类似物进行肽受体放射性核素治疗,这方面的科学文献反映了二十多年的临床经验。此外,还提出了详细的研究观点,以应对该领域取得进展所面临的各种挑战。
{"title":"Heterogeneity of absorbed dose distribution in kidney tissues and dose-response modelling of nephrotoxicity in radiopharmaceutical therapy with beta-particle emitters: A review.","authors":"Clarita Saldarriaga Vargas, Michelle Andersson, Céline Bouvier-Capely, Wei Bo Li, Balázs Madas, Peter Covens, Lara Struelens, Lidia Strigari","doi":"10.1016/j.zemedi.2023.02.006","DOIUrl":"10.1016/j.zemedi.2023.02.006","url":null,"abstract":"<p><p>Absorbed dose heterogeneity in kidney tissues is an important issue in radiopharmaceutical therapy. The effect of absorbed dose heterogeneity in nephrotoxicity is, however, not fully understood yet, which hampers the implementation of treatment optimization by obscuring the interpretation of clinical response data and the selection of optimal treatment options. Although some dosimetry methods have been developed for kidney dosimetry to the level of microscopic renal substructures, the clinical assessment of the microscopic distribution of radiopharmaceuticals in kidney tissues currently remains a challenge. This restricts the anatomical resolution of clinical dosimetry, which hinders a thorough clinical investigation of the impact of absorbed dose heterogeneity. The potential of absorbed dose-response modelling to support individual treatment optimization in radiopharmaceutical therapy is recognized and gaining attraction. However, biophysical modelling is currently underexplored for the kidney, where particular modelling challenges arise from the convolution of a complex functional organization of renal tissues with the function-mediated dose distribution of radiopharmaceuticals. This article reviews and discusses the heterogeneity of absorbed dose distribution in kidney tissues and the absorbed dose-response modelling of nephrotoxicity in radiopharmaceutical therapy. The review focuses mainly on the peptide receptor radionuclide therapy with beta-particle emitting somatostatin analogues, for which the scientific literature reflects over two decades of clinical experience. Additionally, detailed research perspectives are proposed to address various identified challenges to progress in this field.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"491-509"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9629564","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Optimization of the regularization parameter in the Dual Annealing method used for the reconstruction of energy spectrum of electron beam generated by the AQURE mobile accelerator. 优化用于重建 AQURE 移动加速器产生的电子束能谱的双重退火法中的正则化参数。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-04-21 DOI: 10.1016/j.zemedi.2023.03.003
Adam Ryczkowski, Tomasz Piotrowski, Marcin Staszczak, Marcin Wiktorowicz, Przemysław Adrich

Introduction: The shape of the energy spectrum is an essential component of any electron beam Monte Carlo model. Due to specialized equipment and the long measurement time for the direct methods for determining the energy spectrum, attractive alternatives are backward spectrum reconstructions from the measured data. One such approach is solving the first-degree Fredholm integral equation with appropriate regularization. It makes it possible to calculate the depth distribution as the sum of the distributions from monoenergetic beams. This study aims to determine the optimal value of the regularization parameter for the problem of determining the spectrum of the electron beam produced by a mobile accelerator used during intraoperative radiotherapy.

Material and methods: The Geant4 package was used to generate the distributions of deep doses for monoenergetic beams for two models with different degrees of complexity, i.e. simple (theoretical) and full (for the mobile accelerator). The dose distributions for four different shapes of energy spectrum (for each model) were obtained similarly. They were established as the reference data for further calculations. The Dual Annealing optimization method was used to obtain the reconstructed spectrum. The multiple optimizations that differ by the regularization parameter (ranging from 0 to 1) were performed. For each reconstruction, similarity indicators of the energy spectrum and the dose distribution to the referenced data were calculated to determine the optimal regularization parameters.

Results: Optimal regularization parameters determined by similarity indicators for the spectrum and the dose distribution differ for geometry models considered in the study. The regularization parameter for the simple geometry ranged from 0.03 to 0.05, while for full geometry, they were from 0.05 to 0.06. The results for conventional linear accelerators found in the literature range from 0.5 to 1.1.

Conclusion: The Dual Annealing optimization method can be effectively used to solve the Fredholm equation with Tikhonov regularization to reconstruct an electron beam's energy spectrum. The regularization parameter value depends on the beam-forming system. Its value for the mobile accelerator considered in the study ranges from 0.05 to 0.06, depending on the nominal beam energy value.

简介能谱形状是任何电子束蒙特卡罗模型的重要组成部分。由于直接确定能谱的方法需要专门的设备和较长的测量时间,因此从测量数据中重建后向能谱是很有吸引力的替代方法。其中一种方法是通过适当的正则化求解一阶弗雷德霍姆积分方程。这样就可以将深度分布作为单能量光束分布之和进行计算。本研究旨在确定正则化参数的最佳值,以确定术中放疗时使用的移动加速器产生的电子束频谱:使用Geant4软件包为两种复杂程度不同的模型,即简单模型(理论模型)和完整模型(移动加速器模型)生成单能量束的深部剂量分布。四种不同能谱形状(每种模型)的剂量分布也是以类似方式获得的。它们被确定为进一步计算的参考数据。双退火优化法用于获得重建能谱。根据正则化参数(从 0 到 1)的不同,进行了多次优化。对于每次重建,计算能谱和剂量分布与参考数据的相似性指标,以确定最佳正则化参数:结果:根据能谱和剂量分布的相似性指标确定的最佳正则化参数因研究中考虑的几何模型而异。简单几何模型的正则化参数在 0.03 至 0.05 之间,而全几何模型的正则化参数在 0.05 至 0.06 之间。文献中传统线性加速器的正则化参数从 0.5 到 1.1 不等:双重退火优化法可以有效地利用提霍诺夫正则化来求解弗雷德霍姆方程,从而重建电子束能谱。正则化参数值取决于电子束形成系统。研究中考虑的移动加速器的正则化参数值在 0.05 到 0.06 之间,具体取决于标称电子束能量值。
{"title":"Optimization of the regularization parameter in the Dual Annealing method used for the reconstruction of energy spectrum of electron beam generated by the AQURE mobile accelerator.","authors":"Adam Ryczkowski, Tomasz Piotrowski, Marcin Staszczak, Marcin Wiktorowicz, Przemysław Adrich","doi":"10.1016/j.zemedi.2023.03.003","DOIUrl":"10.1016/j.zemedi.2023.03.003","url":null,"abstract":"<p><strong>Introduction: </strong>The shape of the energy spectrum is an essential component of any electron beam Monte Carlo model. Due to specialized equipment and the long measurement time for the direct methods for determining the energy spectrum, attractive alternatives are backward spectrum reconstructions from the measured data. One such approach is solving the first-degree Fredholm integral equation with appropriate regularization. It makes it possible to calculate the depth distribution as the sum of the distributions from monoenergetic beams. This study aims to determine the optimal value of the regularization parameter for the problem of determining the spectrum of the electron beam produced by a mobile accelerator used during intraoperative radiotherapy.</p><p><strong>Material and methods: </strong>The Geant4 package was used to generate the distributions of deep doses for monoenergetic beams for two models with different degrees of complexity, i.e. simple (theoretical) and full (for the mobile accelerator). The dose distributions for four different shapes of energy spectrum (for each model) were obtained similarly. They were established as the reference data for further calculations. The Dual Annealing optimization method was used to obtain the reconstructed spectrum. The multiple optimizations that differ by the regularization parameter (ranging from 0 to 1) were performed. For each reconstruction, similarity indicators of the energy spectrum and the dose distribution to the referenced data were calculated to determine the optimal regularization parameters.</p><p><strong>Results: </strong>Optimal regularization parameters determined by similarity indicators for the spectrum and the dose distribution differ for geometry models considered in the study. The regularization parameter for the simple geometry ranged from 0.03 to 0.05, while for full geometry, they were from 0.05 to 0.06. The results for conventional linear accelerators found in the literature range from 0.5 to 1.1.</p><p><strong>Conclusion: </strong>The Dual Annealing optimization method can be effectively used to solve the Fredholm equation with Tikhonov regularization to reconstruct an electron beam's energy spectrum. The regularization parameter value depends on the beam-forming system. Its value for the mobile accelerator considered in the study ranges from 0.05 to 0.06, depending on the nominal beam energy value.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"510-520"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9775836","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Preconditioning of clinical data for intraocular lens formula constant optimisation using Random Forest Quantile Regression Trees. 使用随机森林定量回归树为优化眼内透镜配方常数的临床数据进行预处理。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-02-20 DOI: 10.1016/j.zemedi.2022.11.009
Achim Langenbucher, Nóra Szentmáry, Alan Cayless, Jascha Wendelstein, Peter Hoffmann

Purpose: To implement a fully data driven strategy for identifying outliers in clinical datasets used for formula constant optimisation, in order to achieve proper formula predicted refraction after cataract surgery, and to assess the capabilities of this outlier detection method.

Methods: 2 clinical datasets (DS1/DS2: N = 888/403) of eyes treated with a monofocal aspherical intraocular lens (Hoya XY1/Johnson&Johnson Vision Z9003) containing preoperative biometric data, power of the lens implant and postoperative spherical equivalent (SEQ) were transferred to us for formula constant optimisation. Original datasets were used to generate baseline formula constants. A random forest quantile regression algorithm was set up using bootstrap resampling with replacement. Quantile regression trees were grown and the 25% and 75% quantile, and the interquartile range were extracted from SEQ and formula predicted refraction REF for the SRKT, Haigis and Castrop formulae. Fences were defined from the quantiles and data points outside the fences were marked and removed as outliers before recalculating the formula constants.

Results: NB = 1000 bootstrap samples were derived from both datasets, and random forest quantile regression trees were grown to model SEQ versus REF and to estimate the median and 25% and 75% quantiles. The fence boundaries were defined as being from 25% quantile - 1.5·IQR to 75% quantile + 1.5·IQR, with data points outside the fence being marked as outliers. In total, for DS1 and DS2, 25/27/32 and 4/5/4 data points were identified as outliers for the SRKT/Haigis/Castrop formulae respectively. The respective root mean squared formula prediction errors for the three formulae were slightly reduced from: 0.4370 dpt;0.4449 dpt/0.3625 dpt;0.4056 dpt/and 0.3376 dpt;0.3532 dpt to: 0.4271 dpt;0.4348 dpt/0.3528 dpt;0.3952 dpt/0.3277 dpt;0.3432 dpt for DS1;DS2.

Conclusion: We were able to prove that with random forest quantile regression trees a fully data driven outlier identification strategy acting in the response space is achievable. In a real life scenario this strategy has to be complemented by an outlier identification method acting in the parameter space for a proper qualification of datasets prior to formula constant optimisation.

目的:在用于公式常数优化的临床数据集中实施一种完全数据驱动的异常值识别策略,以便在白内障手术后获得适当的公式预测屈光度,并评估这种异常值检测方法的能力。方法:我们获得了两个临床数据集(DS1/DS2:N = 888/403),这些数据集包含术前生物测量数据、晶体植入功率和术后球面等值(SEQ),这些数据集使用单焦点非球面眼内透镜(Hoya XY1/Johnson&Johnson Vision Z9003)进行治疗。原始数据集用于生成基线公式常数。使用带替换的引导重采样建立了随机森林量子回归算法。生成量值回归树,并从 SEQ 和 SRKT、Haigis 和 Castrop 公式的预测折射率 REF 中提取 25% 和 75% 的量值以及四分位数间范围。在重新计算公式常量之前,根据量值定义栅栏,并将栅栏外的数据点标记为异常值并清除:从两个数据集中获得了 NB = 1000 个引导样本,并建立了随机森林量化回归树,以模拟 SEQ 与 REF 的关系,并估算出中位数、25% 和 75% 的量化值。栅栏边界被定义为从 25% 量值-1.5-IQR 到 75% 量值+1.5-IQR,栅栏外的数据点被标记为异常值。对于 DS1 和 DS2,SRKT/Haigis/Castrop 公式分别将 25/27/32 和 4/5/4 个数据点确定为离群值。三个公式的均方根公式预测误差分别从 0.4370 dpt;0.4449 dpt/0.3625 dpt;0.4056 dpt/ 和 0.3376 dpt;0.3532 dpt 略微降低到 0.4271 dpt;0.4271 dpt;0.4271 dpt:DS1 和 DS2 的 0.4271 dpt;0.4348 dpt/0.3528 dpt;0.3952 dpt/0.3277 dpt;0.3432 dpt:我们能够证明,利用随机森林量化回归树,可以在响应空间中实现完全由数据驱动的离群值识别策略。在现实生活中,这一策略必须辅以参数空间中的离群值识别方法,以便在公式常数优化之前对数据集进行适当的鉴定。
{"title":"Preconditioning of clinical data for intraocular lens formula constant optimisation using Random Forest Quantile Regression Trees.","authors":"Achim Langenbucher, Nóra Szentmáry, Alan Cayless, Jascha Wendelstein, Peter Hoffmann","doi":"10.1016/j.zemedi.2022.11.009","DOIUrl":"10.1016/j.zemedi.2022.11.009","url":null,"abstract":"<p><strong>Purpose: </strong>To implement a fully data driven strategy for identifying outliers in clinical datasets used for formula constant optimisation, in order to achieve proper formula predicted refraction after cataract surgery, and to assess the capabilities of this outlier detection method.</p><p><strong>Methods: </strong>2 clinical datasets (DS1/DS2: N = 888/403) of eyes treated with a monofocal aspherical intraocular lens (Hoya XY1/Johnson&Johnson Vision Z9003) containing preoperative biometric data, power of the lens implant and postoperative spherical equivalent (SEQ) were transferred to us for formula constant optimisation. Original datasets were used to generate baseline formula constants. A random forest quantile regression algorithm was set up using bootstrap resampling with replacement. Quantile regression trees were grown and the 25% and 75% quantile, and the interquartile range were extracted from SEQ and formula predicted refraction REF for the SRKT, Haigis and Castrop formulae. Fences were defined from the quantiles and data points outside the fences were marked and removed as outliers before recalculating the formula constants.</p><p><strong>Results: </strong>N<sub>B</sub> = 1000 bootstrap samples were derived from both datasets, and random forest quantile regression trees were grown to model SEQ versus REF and to estimate the median and 25% and 75% quantiles. The fence boundaries were defined as being from 25% quantile - 1.5·IQR to 75% quantile + 1.5·IQR, with data points outside the fence being marked as outliers. In total, for DS1 and DS2, 25/27/32 and 4/5/4 data points were identified as outliers for the SRKT/Haigis/Castrop formulae respectively. The respective root mean squared formula prediction errors for the three formulae were slightly reduced from: 0.4370 dpt;0.4449 dpt/0.3625 dpt;0.4056 dpt/and 0.3376 dpt;0.3532 dpt to: 0.4271 dpt;0.4348 dpt/0.3528 dpt;0.3952 dpt/0.3277 dpt;0.3432 dpt for DS1;DS2.</p><p><strong>Conclusion: </strong>We were able to prove that with random forest quantile regression trees a fully data driven outlier identification strategy acting in the response space is achievable. In a real life scenario this strategy has to be complemented by an outlier identification method acting in the parameter space for a proper qualification of datasets prior to formula constant optimisation.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"632-640"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"10816824","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Imaging doses for different CBCT protocols on the Halcyon 3.0 linear accelerator - TLD measurements in an anthropomorphic phantom. Halcyon 3.0 直线加速器上不同 CBCT 方案的成像剂量 - 在人体模型中进行的 TLD 测量。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-04-22 DOI: 10.1016/j.zemedi.2023.03.002
Angelika Altergot, Michaela Schürmann, Tanja Jungert, Hendrik Auerbach, Frank Nüsken, Jan Palm, Christian Rübe, Claudia E Rübe, Yvonne Dzierma

Introduction: Image guided radiotherapy allows for particularly conformal tumour irradiation through precise patient positioning. Becoming the standard for radiotherapy, this increases imaging doses to the patient. The Halcyon 3.0 linear accelerator (Varian Medical Systems, Palo Alto, CA) requires daily imaging due to its geometry. For this reason, the accelerator is equipped with on-line kV and MV imaging. However, daily CBCT images required for irradiation apply additional radiation, which increases the dose to normal tissue and therefore can affect the patient's secondary cancer risk. In this study, actual organ doses were measured for the kV system, and a comparison of normal tissue doses for all available kV CBCT protocols was presented to demonstrate differences in imaging doses across entities and protocols. In addition, effective dose and secondary cancer risk from imaging are evaluated.

Material and methods: Measurements were performed with thermoluminescent dosimeters in an anthropomorphic phantom positioned according to each entity (brain, head and neck, breast, lung, pelvis). CBCT images were obtained, using all available pre-set protocols without further adjustment of the parameters. Measured doses for each position and each protocol were then compared and secondary cancer risk of relevant and specifically radiosensitive organs was calculated.

Results: It was found that imaging doses for protocols such as Pelvis and Head could be reduced by up to half using the corresponding Fast and Low Dose modes, respectively. On the other hand, larger field sizes or the Large mode yielded higher doses than their initial protocols. Image Gently was found to spare normal tissue best, however it is not suitable for certain entities due to low image quality or insufficient projection data.

Discussion: By using appropriate kV-CBCT protocols, it is possible to reduce imaging doses to a significant extent and therefore spare healthy tissue. Combined with studies of image quality, the results of this study could lead to adjustments in workflow regarding the choice of protocols used in daily routine. This could prevent unnecessary radiation exposure and reduce secondary cancer risk.

简介影像引导放射治疗通过对患者进行精确定位,可对肿瘤进行特别适形的照射。影像引导放射治疗已成为放射治疗的标准,可增加患者的成像剂量。Halcyon 3.0 直线加速器(瓦里安医疗系统公司,加利福尼亚州帕洛阿尔托)因其几何形状需要每天进行成像。因此,加速器配备了在线 kV 和 MV 成像。然而,每日照射所需的 CBCT 图像会产生额外的辐射,从而增加正常组织的剂量,因此会影响患者的继发性癌症风险。本研究测量了 kV 系统的实际器官剂量,并比较了所有可用 kV CBCT 方案的正常组织剂量,以显示不同实体和方案的成像剂量差异。此外,还评估了成像的有效剂量和二次癌症风险:使用热释光剂量计在根据各实体(脑、头颈、乳腺、肺、骨盆)定位的拟人化模型中进行测量。使用所有可用的预设方案获取 CBCT 图像,无需进一步调整参数。然后对每个位置和每个方案的测量剂量进行比较,并计算相关和特定辐射敏感器官的二次癌症风险:结果发现,使用相应的快速和低剂量模式,骨盆和头部等方案的成像剂量最多可分别减少一半。另一方面,较大的视野尺寸或大剂量模式产生的剂量高于其初始方案。研究发现,"图像平缓 "模式对正常组织的损伤最小,但由于图像质量低或投影数据不足,该模式不适用于某些实体:讨论:通过使用适当的 kV-CBCT 方案,可以在很大程度上减少成像剂量,从而保护健康组织。结合对图像质量的研究,这项研究的结果可能会调整工作流程中对日常使用方案的选择。这可以避免不必要的辐射照射,降低继发性癌症风险。
{"title":"Imaging doses for different CBCT protocols on the Halcyon 3.0 linear accelerator - TLD measurements in an anthropomorphic phantom.","authors":"Angelika Altergot, Michaela Schürmann, Tanja Jungert, Hendrik Auerbach, Frank Nüsken, Jan Palm, Christian Rübe, Claudia E Rübe, Yvonne Dzierma","doi":"10.1016/j.zemedi.2023.03.002","DOIUrl":"10.1016/j.zemedi.2023.03.002","url":null,"abstract":"<p><strong>Introduction: </strong>Image guided radiotherapy allows for particularly conformal tumour irradiation through precise patient positioning. Becoming the standard for radiotherapy, this increases imaging doses to the patient. The Halcyon 3.0 linear accelerator (Varian Medical Systems, Palo Alto, CA) requires daily imaging due to its geometry. For this reason, the accelerator is equipped with on-line kV and MV imaging. However, daily CBCT images required for irradiation apply additional radiation, which increases the dose to normal tissue and therefore can affect the patient's secondary cancer risk. In this study, actual organ doses were measured for the kV system, and a comparison of normal tissue doses for all available kV CBCT protocols was presented to demonstrate differences in imaging doses across entities and protocols. In addition, effective dose and secondary cancer risk from imaging are evaluated.</p><p><strong>Material and methods: </strong>Measurements were performed with thermoluminescent dosimeters in an anthropomorphic phantom positioned according to each entity (brain, head and neck, breast, lung, pelvis). CBCT images were obtained, using all available pre-set protocols without further adjustment of the parameters. Measured doses for each position and each protocol were then compared and secondary cancer risk of relevant and specifically radiosensitive organs was calculated.</p><p><strong>Results: </strong>It was found that imaging doses for protocols such as Pelvis and Head could be reduced by up to half using the corresponding Fast and Low Dose modes, respectively. On the other hand, larger field sizes or the Large mode yielded higher doses than their initial protocols. Image Gently was found to spare normal tissue best, however it is not suitable for certain entities due to low image quality or insufficient projection data.</p><p><strong>Discussion: </strong>By using appropriate kV-CBCT protocols, it is possible to reduce imaging doses to a significant extent and therefore spare healthy tissue. Combined with studies of image quality, the results of this study could lead to adjustments in workflow regarding the choice of protocols used in daily routine. This could prevent unnecessary radiation exposure and reduce secondary cancer risk.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"580-595"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9775868","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Source-detector trajectory optimization for CBCT metal artifact reduction based on PICCS reconstruction. 基于 PICCS 重建的 CBCT 金属伪影减少源-探测器轨迹优化。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-03-25 DOI: 10.1016/j.zemedi.2023.02.001
Sepideh Hatamikia, Ander Biguri, Gernot Kronreif, Tom Russ, Joachim Kettenbach, Wolfgang Birkfellner

Precise instrument placement plays a critical role in all interventional procedures, especially percutaneous procedures such as needle biopsies, to achieve successful tumor targeting and increased diagnostic accuracy. C-arm cone beam computed tomography (CBCT) has the potential to precisely visualize the anatomy in direct vicinity of the needle and evaluate the adequacy of needle placement during the intervention, allowing for instantaneous adjustment in case of misplacement. However, even with the most advanced C-arm CBCT devices, it can be difficult to identify the exact needle position on CBCT images due to the strong metal artifacts around the needle. In this study, we proposed a framework for customized trajectory design in CBCT imaging based on Prior Image Constrained Compressed Sensing (PICCS) reconstruction with the goal of reducing metal artifacts in needle-based procedures. We proposed to optimize out-of-plane rotations in three-dimensional (3D) space and minimize projection views while reducing metal artifacts at specific volume of interests (VOIs). An anthropomorphic thorax phantom with a needle inserted inside and two tumor models as the imaging targets were used to validate the proposed approach. The performance of the proposed approach was also evaluated for CBCT imaging under kinematic constraints by simulating some collision areas on the geometry of the C-arm. We compared the result of optimized 3D trajectories using the PICCS algorithm and 20 projections with the result of a circular trajectory with sparse view using PICCS and Feldkamp, Davis, and Kress (FDK), both using 20 projections, and the circular FDK method with 313 projections. For imaging targets 1 and 2, the highest values of structural similarity index measure (SSIM) and universal quality index (UQI) between the reconstructed image from the optimized trajectories and the initial CBCT image at the VOI was calculated 0.7521, 0.7308 and 0.7308, 0.7248 respectively. These results significantly outperformed the FDK method (with 20 and 313 projections) and the PICCS method (20 projections) both using the circular trajectory. Our results showed that the proposed optimized trajectories not only significantly reduce metal artifacts but also suggest a dose reduction for needle-based CBCT interventions, considering the small number of projections used. Furthermore, our results showed that the optimized trajectories are compatible with spatially constrained situations and enable CBCT imaging under kinematic constraints when the standard circular trajectory is not feasible.

精确的器械放置在所有介入手术中都起着至关重要的作用,尤其是经皮手术,如针穿活检,这样才能成功瞄准肿瘤并提高诊断准确性。C 型臂锥形束计算机断层扫描(CBCT)可以精确地观察到穿刺针直接附近的解剖结构,并在介入过程中评估穿刺针放置的适当性,以便在放置错误时立即进行调整。然而,即使是最先进的 C 型臂 CBCT 设备,也很难在 CBCT 图像上确定针的准确位置,因为针周围有很强的金属伪影。在这项研究中,我们提出了一个基于先验图像约束压缩传感(PICCS)重建的 CBCT 成像定制轨迹设计框架,目的是减少针式手术中的金属伪影。我们建议优化三维(3D)空间中的平面外旋转,尽量减少投影视图,同时减少特定感兴趣容积(VOI)上的金属伪影。为了验证所提出的方法,我们使用了一个拟人胸腔模型,模型内部插入了一根针,并使用两个肿瘤模型作为成像目标。我们还通过模拟 C 臂几何形状上的一些碰撞区域,评估了所提方法在运动学约束条件下的 CBCT 成像性能。我们比较了使用 PICCS 算法和 20 个投影的优化三维轨迹结果,以及使用 PICCS 和 Feldkamp、Davis 和 Kress(FDK)(均使用 20 个投影)的稀疏视图圆形轨迹结果,以及使用 313 个投影的圆形 FDK 方法。对于目标 1 和目标 2 的成像,优化轨迹重建图像与 VOI 处初始 CBCT 图像之间的结构相似性指数(SSIM)和通用质量指数(UQI)的最高值分别为 0.7521、0.7308 和 0.7308、0.7248。这些结果明显优于使用圆形轨迹的 FDK 方法(20 个和 313 个投影)和 PICCS 方法(20 个投影)。我们的结果表明,考虑到所使用的投影数量较少,建议的优化轨迹不仅能显著减少金属伪影,还能减少针式 CBCT 介入的剂量。此外,我们的结果表明,优化轨迹与空间受限情况兼容,当标准圆形轨迹不可行时,可在运动学受限的情况下进行 CBCT 成像。
{"title":"Source-detector trajectory optimization for CBCT metal artifact reduction based on PICCS reconstruction.","authors":"Sepideh Hatamikia, Ander Biguri, Gernot Kronreif, Tom Russ, Joachim Kettenbach, Wolfgang Birkfellner","doi":"10.1016/j.zemedi.2023.02.001","DOIUrl":"10.1016/j.zemedi.2023.02.001","url":null,"abstract":"<p><p>Precise instrument placement plays a critical role in all interventional procedures, especially percutaneous procedures such as needle biopsies, to achieve successful tumor targeting and increased diagnostic accuracy. C-arm cone beam computed tomography (CBCT) has the potential to precisely visualize the anatomy in direct vicinity of the needle and evaluate the adequacy of needle placement during the intervention, allowing for instantaneous adjustment in case of misplacement. However, even with the most advanced C-arm CBCT devices, it can be difficult to identify the exact needle position on CBCT images due to the strong metal artifacts around the needle. In this study, we proposed a framework for customized trajectory design in CBCT imaging based on Prior Image Constrained Compressed Sensing (PICCS) reconstruction with the goal of reducing metal artifacts in needle-based procedures. We proposed to optimize out-of-plane rotations in three-dimensional (3D) space and minimize projection views while reducing metal artifacts at specific volume of interests (VOIs). An anthropomorphic thorax phantom with a needle inserted inside and two tumor models as the imaging targets were used to validate the proposed approach. The performance of the proposed approach was also evaluated for CBCT imaging under kinematic constraints by simulating some collision areas on the geometry of the C-arm. We compared the result of optimized 3D trajectories using the PICCS algorithm and 20 projections with the result of a circular trajectory with sparse view using PICCS and Feldkamp, Davis, and Kress (FDK), both using 20 projections, and the circular FDK method with 313 projections. For imaging targets 1 and 2, the highest values of structural similarity index measure (SSIM) and universal quality index (UQI) between the reconstructed image from the optimized trajectories and the initial CBCT image at the VOI was calculated 0.7521, 0.7308 and 0.7308, 0.7248 respectively. These results significantly outperformed the FDK method (with 20 and 313 projections) and the PICCS method (20 projections) both using the circular trajectory. Our results showed that the proposed optimized trajectories not only significantly reduce metal artifacts but also suggest a dose reduction for needle-based CBCT interventions, considering the small number of projections used. Furthermore, our results showed that the optimized trajectories are compatible with spatially constrained situations and enable CBCT imaging under kinematic constraints when the standard circular trajectory is not feasible.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"565-579"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9199379","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Quantifying myocardial perfusion during MR-guided interventions without exogenous contrast agents: intra-arterial spin labeling. 在无外源性造影剂的磁共振引导下进行介入治疗时量化心肌灌注:动脉内自旋标记。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-01-28 DOI: 10.1016/j.zemedi.2023.01.002
Simon Reiss, Kevin Wäscher, Ali Caglar Özen, Thomas Lottner, Timo Heidt, Constantin von Zur Mühlen, Michael Bock

Purpose: To test intra-arterial spin labeling (iASL) using active guiding catheters for myocardial perfusion measurements during magnetic resonance (MR)-guided interventions in a pig study.

Methods: In this work, a single-loop radiofrequency (RF) coil at the tip of a 6F active coronary catheter was used as a transmit coil for local spin labeling. The transmit magnetic RF field (B1) of the coil and the labeling efficiency were determined, and iASL was tested in two pigs after the catheter was engaged in the aortic root, the ostium of the left coronary artery (LCA) under MR-guidance. The iASL effect was assessed by the signal difference between spin-labeling On and control (spin-labeling OFF) images, and in a cross-correlation between ON/Off states of spin-labeling a binary labeling paradigm. In addition, quantitative myocardial perfusion was calculated from the iASL experiments.

Results: The maximum B1 in the vicinity of the catheter coil was 2.1 µT. A strong local labeling effect with a labeling efficiency of 0.45 was achieved with iASL both in vitro and in vivo. In both pigs, the proximal myocardial segments supplied by the LCA showed significant labelling effect up to distances of 60 mm from the aortic root with a relative signal difference of (3.14 ± 2.89)% in the first and (3.50 ± 1.25)% in the second animal. The mean correlation coefficients were R = 0.63 ± 0.22 and 0.42 ± 0.16, respectively. The corresponding computed myocardial perfusion values in this region of the myocardium were similar to those obtained with contrast perfusion methods ((1.2 ± 1.1) mL/min/g and (0.8 ± 0.6) mL/min/g).

Conclusion: The proposed iASL method demonstrates the feasibility of selective myocardial perfusion measurements during MR-guided coronary interventions, which with further technical improvements may provide an alternative to exogenous contrast-based perfusion. Due to the invasive nature of the iASL method, it can potentially be used in concert with MRI-guided coronary angioplasty.

目的:在一项猪研究中测试使用主动引导导管进行动脉内自旋标记(iASL),以便在磁共振(MR)引导下进行介入治疗时测量心肌灌注:在这项研究中,6F 有源冠状动脉导管顶端的单环射频线圈被用作局部自旋标记的发射线圈。测定了线圈的射频发射磁场(B1)和标记效率,并在磁共振引导下将导管插入左冠状动脉(LCA)的主动脉根部后,在两头猪身上测试了 iASL。iASL 的效果通过自旋标记打开和对照(自旋标记关闭)图像之间的信号差异,以及二元标记范式下自旋标记打开/关闭状态之间的交叉相关性进行评估。此外,iASL 实验还计算了定量心肌灌注:导管线圈附近的最大 B1 为 2.1 µT。iASL 在体外和体内都实现了很强的局部标记效果,标记效率为 0.45。在两头猪身上,由 LCA 供血的近端心肌节段在距离主动脉根部 60 mm 的范围内都显示出显著的标记效果,第一头猪和第二头猪的相对信号差分别为 (3.14 ± 2.89)% 和 (3.50 ± 1.25)%。平均相关系数分别为 R = 0.63 ± 0.22 和 0.42 ± 0.16。该心肌区域相应的计算心肌灌注值与对比灌注法获得的值相似((1.2 ± 1.1) mL/min/g 和 (0.8 ± 0.6) mL/min/g):拟议的 iASL 方法证明了在磁共振引导下冠状动脉介入治疗期间进行选择性心肌灌注测量的可行性,随着技术的进一步改进,该方法可能成为外源性造影剂灌注的替代方法。由于 iASL 方法的侵入性,它有可能与 MRI 引导下的冠状动脉血管成形术配合使用。
{"title":"Quantifying myocardial perfusion during MR-guided interventions without exogenous contrast agents: intra-arterial spin labeling.","authors":"Simon Reiss, Kevin Wäscher, Ali Caglar Özen, Thomas Lottner, Timo Heidt, Constantin von Zur Mühlen, Michael Bock","doi":"10.1016/j.zemedi.2023.01.002","DOIUrl":"10.1016/j.zemedi.2023.01.002","url":null,"abstract":"<p><strong>Purpose: </strong>To test intra-arterial spin labeling (iASL) using active guiding catheters for myocardial perfusion measurements during magnetic resonance (MR)-guided interventions in a pig study.</p><p><strong>Methods: </strong>In this work, a single-loop radiofrequency (RF) coil at the tip of a 6F active coronary catheter was used as a transmit coil for local spin labeling. The transmit magnetic RF field (B<sub>1</sub>) of the coil and the labeling efficiency were determined, and iASL was tested in two pigs after the catheter was engaged in the aortic root, the ostium of the left coronary artery (LCA) under MR-guidance. The iASL effect was assessed by the signal difference between spin-labeling On and control (spin-labeling OFF) images, and in a cross-correlation between ON/Off states of spin-labeling a binary labeling paradigm. In addition, quantitative myocardial perfusion was calculated from the iASL experiments.</p><p><strong>Results: </strong>The maximum B<sub>1</sub> in the vicinity of the catheter coil was 2.1 µT. A strong local labeling effect with a labeling efficiency of 0.45 was achieved with iASL both in vitro and in vivo. In both pigs, the proximal myocardial segments supplied by the LCA showed significant labelling effect up to distances of 60 mm from the aortic root with a relative signal difference of (3.14 ± 2.89)% in the first and (3.50 ± 1.25)% in the second animal. The mean correlation coefficients were R = 0.63 ± 0.22 and 0.42 ± 0.16, respectively. The corresponding computed myocardial perfusion values in this region of the myocardium were similar to those obtained with contrast perfusion methods ((1.2 ± 1.1) mL/min/g and (0.8 ± 0.6) mL/min/g).</p><p><strong>Conclusion: </strong>The proposed iASL method demonstrates the feasibility of selective myocardial perfusion measurements during MR-guided coronary interventions, which with further technical improvements may provide an alternative to exogenous contrast-based perfusion. Due to the invasive nature of the iASL method, it can potentially be used in concert with MRI-guided coronary angioplasty.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"596-609"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"10587236","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Reduction of patient specific quality assurance through plan complexity metrics for VMAT plans with an open-source TPS script. 通过使用开源 TPS 脚本的 VMAT 计划复杂性指标,减少针对患者的质量保证。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-03-29 DOI: 10.1016/j.zemedi.2023.02.003
Klara Uher, Stefanie Ehrbar, Stephanie Tanadini-Lang, Riccardo Dal Bello

Purpose: Volumetric modulated arc therapy (VMAT) is a widespread technique for the delivery of normo-fractionated radiation therapy (NFRT) and stereotactic body radiation therapy (SBRT). It is associated with a significant hardware burden requiring dose rate modulation, collimator movement and gantry rotation synchronisation. Patient specific quality assurance (PSQA) guarantees that the linacs can precisely and accurately deliver the planned dose. However, PSQA requires a significant time allocation and class solutions to reduce this while guaranteeing the deliverability of the plans should be investigated.

Methods: In this study, an in-house developed Eclipse Scripting API (ESAPI) script was used to extract five independent plan complexity metrics from N = 667 VMAT treatment fields. The correlation between metrics and portal dosimetry measurements was investigated with Pearson correlation, box plot analysis and receiver operating characteristic curves, which were used to defined the best performing metric and its threshold.

Results: The incidence of fields failing the clinical PSQA criteria of 3%/2mm (NFRT) and 3%/1.5mm (SBRT) was low (N = 1). The mean MLC opening was the metric with the highest correlation with the portal dosimetry data and among the best in discriminating the requirement of PSQA. The thresholds of 16.12 mm (NFRT) and 7.96 mm (SBRT) corresponded to true positive rates higher than 90%.

Conclusions: This work presents a quantitative approach to reduce the time allocation for PSQA by identifying the most complex plans demanding a dedicated measurement. The proposed method requires PSQA for approximately 10% of the plans. The ESAPI script is distributed open-source to ease the investigation and implementation at other institutions.

目的:容积调制弧形疗法(VMAT)是一种广泛应用于普通分次放射治疗(NFRT)和立体定向体放射治疗(SBRT)的技术。该技术需要进行剂量率调制、准直器移动和龙门旋转同步,硬件负担很重。患者特定质量保证(PSQA)可确保直列加速器能够精确地提供计划剂量。然而,PSQA需要花费大量时间,因此应研究既能减少时间又能保证计划可实施性的一类解决方案:本研究使用内部开发的 Eclipse Scripting API (ESAPI) 脚本从 N = 667 个 VMAT 治疗区域中提取五个独立的计划复杂性指标。利用皮尔逊相关性、箱形图分析和接收器操作特征曲线研究了指标与门户剂量测定测量之间的相关性,并以此确定了性能最佳的指标及其阈值:未达到临床 PSQA 标准 3%/2mm(NFRT)和 3%/1.5mm(SBRT)的视野发生率较低(N = 1)。平均 MLC 开口是与门静脉剂量测定数据相关性最高的指标,也是区分 PSQA 要求的最佳指标之一。16.12毫米(NFRT)和7.96毫米(SBRT)的阈值对应的真实阳性率高于90%:这项工作提出了一种定量方法,通过识别需要专门测量的最复杂计划来减少 PSQA 的时间分配。所提出的方法需要对大约 10% 的计划进行 PSQA。ESAPI 脚本是开源的,便于其他机构进行研究和实施。
{"title":"Reduction of patient specific quality assurance through plan complexity metrics for VMAT plans with an open-source TPS script.","authors":"Klara Uher, Stefanie Ehrbar, Stephanie Tanadini-Lang, Riccardo Dal Bello","doi":"10.1016/j.zemedi.2023.02.003","DOIUrl":"10.1016/j.zemedi.2023.02.003","url":null,"abstract":"<p><strong>Purpose: </strong>Volumetric modulated arc therapy (VMAT) is a widespread technique for the delivery of normo-fractionated radiation therapy (NFRT) and stereotactic body radiation therapy (SBRT). It is associated with a significant hardware burden requiring dose rate modulation, collimator movement and gantry rotation synchronisation. Patient specific quality assurance (PSQA) guarantees that the linacs can precisely and accurately deliver the planned dose. However, PSQA requires a significant time allocation and class solutions to reduce this while guaranteeing the deliverability of the plans should be investigated.</p><p><strong>Methods: </strong>In this study, an in-house developed Eclipse Scripting API (ESAPI) script was used to extract five independent plan complexity metrics from N = 667 VMAT treatment fields. The correlation between metrics and portal dosimetry measurements was investigated with Pearson correlation, box plot analysis and receiver operating characteristic curves, which were used to defined the best performing metric and its threshold.</p><p><strong>Results: </strong>The incidence of fields failing the clinical PSQA criteria of 3%/2mm (NFRT) and 3%/1.5mm (SBRT) was low (N = 1). The mean MLC opening was the metric with the highest correlation with the portal dosimetry data and among the best in discriminating the requirement of PSQA. The thresholds of 16.12 mm (NFRT) and 7.96 mm (SBRT) corresponded to true positive rates higher than 90%.</p><p><strong>Conclusions: </strong>This work presents a quantitative approach to reduce the time allocation for PSQA by identifying the most complex plans demanding a dedicated measurement. The proposed method requires PSQA for approximately 10% of the plans. The ESAPI script is distributed open-source to ease the investigation and implementation at other institutions.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"555-564"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9279104","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Development of a PTV margin for preclinical irradiation of orthotopic pancreatic tumors derived from a well-known recipe for humans. 为临床前照射正位胰腺肿瘤开发 PTV 边界,该边界源自著名的人体配方。
IF 2.4 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-11-01 Epub Date: 2023-05-22 DOI: 10.1016/j.zemedi.2023.03.005
Severin Kampfer, Sophie Dobiasch, Stephanie E Combs, Jan J Wilkens

In human radiotherapy a safety margin (PTV margin) is essential for successful irradiation and is usually part of clinical treatment planning. In preclinical radiotherapy research with small animals, most uncertainties and inaccuracies are present as well, but according to the literature a margin is used only scarcely. In addition, there is only little experience about the appropriate size of the margin, which should carefully be investigated and considered, since sparing of organs at risk or normal tissue is affected. Here we estimate the needed margin for preclinical irradiation by adapting a well-known human margin recipe from van Herck et al. to the dimensions and requirements of the specimen on a small animal radiation research platform (SARRP). We adjusted the factors of the described formula to the specific challenges in an orthotopic pancreatic tumor mouse model to establish an appropriate margin concept. The SARRP was used with its image-guidance irradiation possibility for arc irradiation with a field size of 10 × 10 mm2 for 5 fractions. Our goal was to irradiate the clinical target volume (CTV) of at least 90% of our mice with at least 95% of the prescribed dose. By carefully analyzing all relevant factors we gain a CTV to planning target volume (PTV) margin of 1.5 mm for our preclinical setup. The stated safety margin is strongly dependent on the exact setting of the experiment and has to be adjusted for other experimental settings. The few stated values in literature correspond well to our result. Even if using margins in the preclinical setting might be an additional challenge, we think it is crucial to use them to produce reliable results and improve the efficacy of radiotherapy.

在人类放射治疗中,安全裕度(PTV裕度)是成功照射的必要条件,通常也是临床治疗计划的一部分。在使用小动物进行临床前放疗研究时,也存在着大多数不确定性和不准确性,但根据文献资料,很少使用安全边际。此外,关于边缘的适当大小也鲜有经验可循,这一点应仔细研究和考虑,因为这将影响对危险器官或正常组织的保护。在此,我们根据小动物辐射研究平台(SARRP)上标本的尺寸和要求,对 van Herck 等人的著名人体边缘配方进行了调整,从而估算出临床前辐照所需的边缘。我们根据正位胰腺肿瘤小鼠模型所面临的具体挑战调整了所述公式中的系数,以建立适当的边缘概念。SARRP 可在图像引导下进行弧形照射,照射野大小为 10 × 10 平方毫米,共照射 5 次。我们的目标是用至少 95% 的规定剂量照射至少 90% 的小鼠的临床靶体积(CTV)。通过仔细分析所有相关因素,我们将临床前设置的 CTV 与规划靶体积 (PTV) 的裕度定为 1.5 毫米。所述安全裕度在很大程度上取决于实验的具体设置,必须根据其他实验设置进行调整。文献中给出的几个数值与我们的结果非常吻合。尽管在临床前设置中使用安全裕度可能是一个额外的挑战,但我们认为使用安全裕度对产生可靠的结果和提高放疗疗效至关重要。
{"title":"Development of a PTV margin for preclinical irradiation of orthotopic pancreatic tumors derived from a well-known recipe for humans.","authors":"Severin Kampfer, Sophie Dobiasch, Stephanie E Combs, Jan J Wilkens","doi":"10.1016/j.zemedi.2023.03.005","DOIUrl":"10.1016/j.zemedi.2023.03.005","url":null,"abstract":"<p><p>In human radiotherapy a safety margin (PTV margin) is essential for successful irradiation and is usually part of clinical treatment planning. In preclinical radiotherapy research with small animals, most uncertainties and inaccuracies are present as well, but according to the literature a margin is used only scarcely. In addition, there is only little experience about the appropriate size of the margin, which should carefully be investigated and considered, since sparing of organs at risk or normal tissue is affected. Here we estimate the needed margin for preclinical irradiation by adapting a well-known human margin recipe from van Herck et al. to the dimensions and requirements of the specimen on a small animal radiation research platform (SARRP). We adjusted the factors of the described formula to the specific challenges in an orthotopic pancreatic tumor mouse model to establish an appropriate margin concept. The SARRP was used with its image-guidance irradiation possibility for arc irradiation with a field size of 10 × 10 mm<sup>2</sup> for 5 fractions. Our goal was to irradiate the clinical target volume (CTV) of at least 90% of our mice with at least 95% of the prescribed dose. By carefully analyzing all relevant factors we gain a CTV to planning target volume (PTV) margin of 1.5 mm for our preclinical setup. The stated safety margin is strongly dependent on the exact setting of the experiment and has to be adjusted for other experimental settings. The few stated values in literature correspond well to our result. Even if using margins in the preclinical setting might be an additional challenge, we think it is crucial to use them to produce reliable results and improve the efficacy of radiotherapy.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":" ","pages":"533-541"},"PeriodicalIF":2.4,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9516757","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Zeitschrift fur Medizinische Physik
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1