Automated Light Transmission Aggregometry with and without Platelet Poor Plasma Reference: A Method Comparison.

Ulrich J Sachs, Lida Röder, Nina Cooper, Christian Radon, Hans-Jürgen Kolde
{"title":"Automated Light Transmission Aggregometry with and without Platelet Poor Plasma Reference: A Method Comparison.","authors":"Ulrich J Sachs,&nbsp;Lida Röder,&nbsp;Nina Cooper,&nbsp;Christian Radon,&nbsp;Hans-Jürgen Kolde","doi":"10.1055/s-0043-1762588","DOIUrl":null,"url":null,"abstract":"<p><p><b>Background</b>  Light transmission aggregometry (LTA) is considered the gold standard for the evaluation of platelet function but is labor-intensive and involves numerous manual steps. Automation may contribute to standardization. Here, we evaluate the performance characteristics of a new automated instrument, Thrombomate XRA (TXRA), and compare it against a manual instrument (PAP-8). <b>Materials and Methods</b>  Leftover blood samples from blood donors or patients were tested in parallel with identical reagents and in identical concentrations both manually using PAP-8 and automated on the TXRA. In addition to precision and method comparison, an additional evaluation was performed on the TXRA against \"virtual\" platelet-poor plasma (VPPP) based on artificial intelligence. The main focus was on comparing the maximum aggregation (MA%) values. <b>Results</b>  Precision for MA% ranged from 1.4 to 4.6% on TXRA for all reagents. Normal ranges for 100 healthy blood donors on both instruments were in a similar range for all reagents, with a tendency to slightly higher values with TXRA. Most agonists resulted in normally distributed MA%. Comparing 47 patient samples on both devices showed a good correlation for both slope and MA% with some differences in individual samples with epinephrine and TRAP. Correlation between the TXRA measurement against PPP and \"virtual\" PPP demonstrated excellent correlation. Reaction signatures of both devices were very similar. <b>Conclusion</b>  TXRA provides reproducible LTA results that correlate with an established manual method when tested against PPP or VPPP. Its ability to perform LTA only from platelet-rich plasma without requiring autologous PPP simplifies LTA. TXRA is an important step not only for further standardizing LTA but also for a more widespread use of this important method.</p>","PeriodicalId":22238,"journal":{"name":"TH Open: Companion Journal to Thrombosis and Haemostasis","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9946786/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"TH Open: Companion Journal to Thrombosis and Haemostasis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1055/s-0043-1762588","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background  Light transmission aggregometry (LTA) is considered the gold standard for the evaluation of platelet function but is labor-intensive and involves numerous manual steps. Automation may contribute to standardization. Here, we evaluate the performance characteristics of a new automated instrument, Thrombomate XRA (TXRA), and compare it against a manual instrument (PAP-8). Materials and Methods  Leftover blood samples from blood donors or patients were tested in parallel with identical reagents and in identical concentrations both manually using PAP-8 and automated on the TXRA. In addition to precision and method comparison, an additional evaluation was performed on the TXRA against "virtual" platelet-poor plasma (VPPP) based on artificial intelligence. The main focus was on comparing the maximum aggregation (MA%) values. Results  Precision for MA% ranged from 1.4 to 4.6% on TXRA for all reagents. Normal ranges for 100 healthy blood donors on both instruments were in a similar range for all reagents, with a tendency to slightly higher values with TXRA. Most agonists resulted in normally distributed MA%. Comparing 47 patient samples on both devices showed a good correlation for both slope and MA% with some differences in individual samples with epinephrine and TRAP. Correlation between the TXRA measurement against PPP and "virtual" PPP demonstrated excellent correlation. Reaction signatures of both devices were very similar. Conclusion  TXRA provides reproducible LTA results that correlate with an established manual method when tested against PPP or VPPP. Its ability to perform LTA only from platelet-rich plasma without requiring autologous PPP simplifies LTA. TXRA is an important step not only for further standardizing LTA but also for a more widespread use of this important method.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
自动光透射聚集法与无血小板差血浆参比:一种方法比较。
背景光透射聚合法(LTA)被认为是评估血小板功能的金标准,但它是劳动密集型的,涉及许多手动步骤。自动化可能有助于标准化。在这里,我们评估了一种新的自动化仪器,血栓XRA (TXRA)的性能特征,并将其与手动仪器(PAP-8)进行比较。材料和方法采用相同的试剂和相同的浓度对献血者或患者的剩余血液样本进行平行检测,手动使用PAP-8和自动在TXRA上进行检测。除了精度和方法比较外,还对基于人工智能的TXRA对抗“虚拟”血小板缺乏血浆(VPPP)进行了额外的评估。主要焦点是比较最大聚合(MA%)值。结果TXRA对所有试剂的MA%精密度为1.4 ~ 4.6%。100名健康献血者在两种仪器上的正常范围与所有试剂的范围相似,TXRA的值有略高的趋势。大多数激动剂导致正态分布的MA%。比较两种装置上的47例患者样本显示斜率和MA%具有良好的相关性,肾上腺素和TRAP的个体样本存在一些差异。TXRA测量与PPP和“虚拟”PPP之间的相关性显示出极好的相关性。两种装置的反应特征非常相似。结论TXRA提供了可重复的LTA结果与建立的人工方法在PPP或VPPP测试时相关。它能够仅从富含血小板的血浆中进行LTA,而不需要自体PPP,从而简化了LTA。TXRA不仅是进一步规范LTA的重要一步,也是这一重要方法得到更广泛应用的重要一步。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Management of Urgent Bleeding in Patients with Hemophilia A: Focus on the Use of Emicizumab Disease Burden, Clinical Outcomes, and Quality of Life in People with Hemophilia A without Inhibitors in Europe: Analyses from CHESS II/CHESS PAEDs TH Open Continues to Highlight the State-of-the-Art on Thrombosis and Hemostasis with a Renewed Editorial Board The Effect of Rituximab on Antiphospholipid Titers in Patients with Antiphospholipid Syndrome. Potential for a Virtual Care Model in the Perioperative Management of Anticoagulant Therapy: A 5-Year Retrospective Clinic Review.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1