{"title":"An acoustic reflex technique of establishing hearing aid settings.","authors":"R W Keith","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>An acoustic reflex method of establishing hearing aid gain settings for most comfortable loudness was tested. Forty hearing-aid users were examined to determine relationships among MCL, loudness discomfort level, and the contralateral acoustic reflex to speech, under unaided and aided conditions. Results indicate that a simple 2-part rule results in amplification to most comfortable loudness for 76% of subjects. Implications for establishing hearing aid saturation sound pressure level settings are discussed.</p>","PeriodicalId":76027,"journal":{"name":"Journal of the American Auditory Society","volume":"5 2","pages":"71-5"},"PeriodicalIF":0.0000,"publicationDate":"1979-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the American Auditory Society","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
An acoustic reflex method of establishing hearing aid gain settings for most comfortable loudness was tested. Forty hearing-aid users were examined to determine relationships among MCL, loudness discomfort level, and the contralateral acoustic reflex to speech, under unaided and aided conditions. Results indicate that a simple 2-part rule results in amplification to most comfortable loudness for 76% of subjects. Implications for establishing hearing aid saturation sound pressure level settings are discussed.