Practical considerations for the use of a pulmonary artery thermistor catheter.

IF 2.7 2区 医学 Q1 SURGERY Surgery Pub Date : 1976-04-01
M Woods, R N Scott, A H Harken
{"title":"Practical considerations for the use of a pulmonary artery thermistor catheter.","authors":"M Woods,&nbsp;R N Scott,&nbsp;A H Harken","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The use of a pulmonary artery thermistor catheter for pressure measurement and thermodilution cardiac output determination was evaluated in 11 dogs. Pulmonary artery wedge pressure was a reliable index of left atrial pressure at end-expiratory pressures less than 10 cm. H2O. Fluctuations in pulmonary artery temperature occurred at a frequency equal to the respiratory rate and an amplitude of 0.010 to 0.086 degrees C. Changes in amplitude were associated with changes in ventilatory waveform, respiratory rate, and level of anesthesia. Intermittent and continuous positive-pressure ventilation generally dampened and reversed the pulmonary artery temperature pattern exhibited during spontaneous breathing. This suggested that when end expiration is used to time indicator injection, cardiac output will be underestimated during spontaneous breathing and overestimated during continuous or intermittent positive-pressure ventilation. When indicator was injected at the same point in the ventilatory cycle, successive values of cardiac output deviated from one another by 0.0 to 6.7 percent. Deviations as large as 14 percent resulted if sequential injections were out of phase by half a respiratory cycle. These deviations can be minimized by injecting indicator at the same point in the respiratory cycle, if it is not feasible to measure cardiac output during apnea. The clinical utility of a pulmonary artery thermistor catheter can be optimized through appreciation of its specific strengths and limitations.</p>","PeriodicalId":22152,"journal":{"name":"Surgery","volume":"79 4","pages":"469-75"},"PeriodicalIF":2.7000,"publicationDate":"1976-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Surgery","FirstCategoryId":"3","ListUrlMain":"","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"SURGERY","Score":null,"Total":0}
引用次数: 0

Abstract

The use of a pulmonary artery thermistor catheter for pressure measurement and thermodilution cardiac output determination was evaluated in 11 dogs. Pulmonary artery wedge pressure was a reliable index of left atrial pressure at end-expiratory pressures less than 10 cm. H2O. Fluctuations in pulmonary artery temperature occurred at a frequency equal to the respiratory rate and an amplitude of 0.010 to 0.086 degrees C. Changes in amplitude were associated with changes in ventilatory waveform, respiratory rate, and level of anesthesia. Intermittent and continuous positive-pressure ventilation generally dampened and reversed the pulmonary artery temperature pattern exhibited during spontaneous breathing. This suggested that when end expiration is used to time indicator injection, cardiac output will be underestimated during spontaneous breathing and overestimated during continuous or intermittent positive-pressure ventilation. When indicator was injected at the same point in the ventilatory cycle, successive values of cardiac output deviated from one another by 0.0 to 6.7 percent. Deviations as large as 14 percent resulted if sequential injections were out of phase by half a respiratory cycle. These deviations can be minimized by injecting indicator at the same point in the respiratory cycle, if it is not feasible to measure cardiac output during apnea. The clinical utility of a pulmonary artery thermistor catheter can be optimized through appreciation of its specific strengths and limitations.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
肺动脉热敏电阻导管使用的实际考虑。
我们对11只狗进行了肺动脉热敏电阻导管的压力测量和热稀释心输出量测定。肺动脉楔压是呼气末气压小于10 cm时左房压的可靠指标。H2O。肺动脉温度波动的频率与呼吸频率相等,幅度为0.010 ~ 0.086℃,幅度的变化与通气波形、呼吸频率和麻醉程度的变化有关。间歇和持续正压通气通常会抑制和逆转自发呼吸时表现出的肺动脉温度模式。这表明,当使用末端呼气来计时指标注射时,在自主呼吸时心输出量会被低估,而在连续或间歇正压通气时心输出量会被高估。当在通气周期的同一点注射指示剂时,心输出量的连续值彼此偏离0.0 ~ 6.7%。如果连续注射在半个呼吸周期内不同步,则会导致14%的偏差。如果在呼吸暂停期间无法测量心输出量,可以通过在呼吸周期的同一点注射指示剂来最小化这些偏差。肺动脉热敏电阻导管的临床应用可以通过了解其特定的优势和局限性来优化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Surgery
Surgery 医学-外科
CiteScore
5.40
自引率
5.30%
发文量
687
审稿时长
64 days
期刊介绍: For 66 years, Surgery has published practical, authoritative information about procedures, clinical advances, and major trends shaping general surgery. Each issue features original scientific contributions and clinical reports. Peer-reviewed articles cover topics in oncology, trauma, gastrointestinal, vascular, and transplantation surgery. The journal also publishes papers from the meetings of its sponsoring societies, the Society of University Surgeons, the Central Surgical Association, and the American Association of Endocrine Surgeons.
期刊最新文献
ERGO (ERGonomics in the Operating room) study: A cross-sectional international online survey. Contents Acknowledgments of Peer Reviewers Recorder's notes from the 82nd annual meeting of the Central surgical association held in Indianapolis, in on June 5th to 7th, 2025 Cover 1(with editorial board)
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1