M. Kyro, J. Simola, K. Haneda, K. Takizawa, H. Hagiwara, P. Vainikainen
{"title":"Development of a channel model for 60 GHz radio systems in an angiography room","authors":"M. Kyro, J. Simola, K. Haneda, K. Takizawa, H. Hagiwara, P. Vainikainen","doi":"10.1109/BIOWIRELESS.2011.5724338","DOIUrl":null,"url":null,"abstract":"This paper presents the development of a radio channel model for the feasibility study of 60 GHz high-speed radio systems for a medical operation in a hospital environment. Our particular usage scenario is an angiography room, where the 60 GHz radio system sends video data from an x-ray machine to a movable screen. Our channel model is based on an extensive radio channel measurement in an angiography room in a hospital. Pathloss and multipath characteristics were statistically modeled, leading to a generic channel model in an angiography room. A channel model structure, complete set of model parameters, and implementation recipe of the developed model is provided.","PeriodicalId":430449,"journal":{"name":"2011 IEEE Topical Conference on Biomedical Wireless Technologies, Networks, and Sensing Systems","volume":"33 1-2 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2011-03-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2011 IEEE Topical Conference on Biomedical Wireless Technologies, Networks, and Sensing Systems","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BIOWIRELESS.2011.5724338","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5
Abstract
This paper presents the development of a radio channel model for the feasibility study of 60 GHz high-speed radio systems for a medical operation in a hospital environment. Our particular usage scenario is an angiography room, where the 60 GHz radio system sends video data from an x-ray machine to a movable screen. Our channel model is based on an extensive radio channel measurement in an angiography room in a hospital. Pathloss and multipath characteristics were statistically modeled, leading to a generic channel model in an angiography room. A channel model structure, complete set of model parameters, and implementation recipe of the developed model is provided.