{"title":"A robust six-to-four port reduction technique for the calibration of six-port microwave network analysers","authors":"A. Wright, S. Judah","doi":"10.1109/IMTC.1990.65977","DOIUrl":null,"url":null,"abstract":"The authors introduce a simple method by which the robust calibration technique developed by S.K. Judah (1985) may be unified with the thru-reflect-line technique, yielding initial estimates of the six-port instrumental parameters that are insensitive to noise and random errors. This calibration technique has been employed in a single six-port reflectometer operating at 2 GHz. Experimental experience has shown the technique to be robust; problems encountered using the method recommended by G.F. Engen (1979) have not arisen. Investigation by computer simulation has also shown the technique to be reliable. This methodology requires a substantial increase in software effort owing to the effective operational requirement of two full calibration techniques.<<ETX>>","PeriodicalId":404761,"journal":{"name":"7th IEEE Conference on Instrumentation and Measurement Technology","volume":"147 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1990-02-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"7th IEEE Conference on Instrumentation and Measurement Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IMTC.1990.65977","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2
Abstract
The authors introduce a simple method by which the robust calibration technique developed by S.K. Judah (1985) may be unified with the thru-reflect-line technique, yielding initial estimates of the six-port instrumental parameters that are insensitive to noise and random errors. This calibration technique has been employed in a single six-port reflectometer operating at 2 GHz. Experimental experience has shown the technique to be robust; problems encountered using the method recommended by G.F. Engen (1979) have not arisen. Investigation by computer simulation has also shown the technique to be reliable. This methodology requires a substantial increase in software effort owing to the effective operational requirement of two full calibration techniques.<>