{"title":"Microwave Transmission Method for The Detection of Water Holdup in Oil-water Mixtures Based on a Yagi Antenna","authors":"Hongwei Qin, Pei Yang, Ruirong Dang","doi":"10.1784/insi.2023.65.8.443","DOIUrl":null,"url":null,"abstract":"In this paper, Yagi antennas are introduced into microwave transmission measuring instruments as the transmit and receive antennas for detecting moisture content in oil-water mixtures. A Yagi antenna is designed, where the simulation results show a peak gain of 9 dBi and the reflection\n coefficient S11 is lower than ???10 dB in a frequency band of 2.25-4 GHz. Meanwhile, the measurement result using a vector network analyser shows that the Yagi antenna works in the frequency band of 2.9-4.15 GHz, indicating that measurement results shifted to high frequencies. Based on the\n Yagi antennas, a moisture content measuring system using the microwave transmission method is designed and constructed, where the measurement results show the Yagi directional high-gain microwave antenna used in this paper can achieve water holdup measurement in the range of 0-100%, where\n the relative error is less than ??15% and the absolute error is within ??2.341%.","PeriodicalId":344397,"journal":{"name":"Insight - Non-Destructive Testing and Condition Monitoring","volume":"145 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Insight - Non-Destructive Testing and Condition Monitoring","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1784/insi.2023.65.8.443","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
In this paper, Yagi antennas are introduced into microwave transmission measuring instruments as the transmit and receive antennas for detecting moisture content in oil-water mixtures. A Yagi antenna is designed, where the simulation results show a peak gain of 9 dBi and the reflection
coefficient S11 is lower than ???10 dB in a frequency band of 2.25-4 GHz. Meanwhile, the measurement result using a vector network analyser shows that the Yagi antenna works in the frequency band of 2.9-4.15 GHz, indicating that measurement results shifted to high frequencies. Based on the
Yagi antennas, a moisture content measuring system using the microwave transmission method is designed and constructed, where the measurement results show the Yagi directional high-gain microwave antenna used in this paper can achieve water holdup measurement in the range of 0-100%, where
the relative error is less than ??15% and the absolute error is within ??2.341%.