{"title":"Stress shielding of femoral component with and without collar","authors":"H. A. Mansour, J. Ray, D. P. Mukherjee","doi":"10.1109/SBEC.1995.514429","DOIUrl":null,"url":null,"abstract":"Stress shielding in metallic implants used in total hip arthroplasty is often mentioned as the cause of loosening in many clinical cases. In this study the stress shielding was evaluated in Ti6Al4V alloy commercial femoral components with and without collar. The results from the strain gauge measurements showed that the stress shielding was lower for femoral components collar than that seen without collar.","PeriodicalId":332563,"journal":{"name":"Proceedings of the 1995 Fourteenth Southern Biomedical Engineering Conference","volume":"44 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1995-04-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the 1995 Fourteenth Southern Biomedical Engineering Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/SBEC.1995.514429","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
Abstract
Stress shielding in metallic implants used in total hip arthroplasty is often mentioned as the cause of loosening in many clinical cases. In this study the stress shielding was evaluated in Ti6Al4V alloy commercial femoral components with and without collar. The results from the strain gauge measurements showed that the stress shielding was lower for femoral components collar than that seen without collar.