M. Campos-García, Cesar Cossio-Guerrero, O. Huerta-Carranza, V. Moreno-Oliva
{"title":"Advances in corneal topography measurements with conical null-screens","authors":"M. Campos-García, Cesar Cossio-Guerrero, O. Huerta-Carranza, V. Moreno-Oliva","doi":"10.1117/12.2192137","DOIUrl":null,"url":null,"abstract":"In this work we report the design of a null-screen for corneal topography. To avoid the difficulties in the alignment of the test system due to the face contour (eyebrows, nose, or eyelids), we design a conical null-screen with a novel radial points distribution drawn on it in such a way that its image, which is formed by reflection on the test surface, becomes an exact array of circular spots if the surface is perfect. Additionally, an algorithm to compute the sagittal and meridional radii of curvature for the corneal surface is presented. The sagittal radius is obtained from the surface normal, and the meridional radius is calculated from a function fitted to the derivative of the sagittal curvature by using the surfacenormals raw data. Experimental results for the testing a calibration spherical surface are shown. Also, we perform some corneal topography measurements.","PeriodicalId":212434,"journal":{"name":"SPIE Optical Systems Design","volume":"8 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2015-09-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"SPIE Optical Systems Design","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.2192137","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5
Abstract
In this work we report the design of a null-screen for corneal topography. To avoid the difficulties in the alignment of the test system due to the face contour (eyebrows, nose, or eyelids), we design a conical null-screen with a novel radial points distribution drawn on it in such a way that its image, which is formed by reflection on the test surface, becomes an exact array of circular spots if the surface is perfect. Additionally, an algorithm to compute the sagittal and meridional radii of curvature for the corneal surface is presented. The sagittal radius is obtained from the surface normal, and the meridional radius is calculated from a function fitted to the derivative of the sagittal curvature by using the surfacenormals raw data. Experimental results for the testing a calibration spherical surface are shown. Also, we perform some corneal topography measurements.