Changhan Jun, Sung-hwan Lim, J. Wolinsky, T. Garzon-Muvdi, D. Petrisor, K. Cleary, D. Stoianovici
{"title":"MR Safe Robot Assisted Needle Access of the Brain: Preclinical Study","authors":"Changhan Jun, Sung-hwan Lim, J. Wolinsky, T. Garzon-Muvdi, D. Petrisor, K. Cleary, D. Stoianovici","doi":"10.1142/S2424905X18500034","DOIUrl":null,"url":null,"abstract":"We report the results of preclinical experiments for direct MRI-guided needle interventions in the brain. An MR Safe robot was incorporated into an intraoperative MRI system. Deep regions of the brain simulated in a cranial mockup were targeted with a needle under robotic assistance. The 3D accuracy of in-scanner targeting at an average depth of 95mm was 1.55mm, with no manual corrections.","PeriodicalId":447761,"journal":{"name":"J. Medical Robotics Res.","volume":"25 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2018-02-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"17","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"J. Medical Robotics Res.","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1142/S2424905X18500034","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 17
Abstract
We report the results of preclinical experiments for direct MRI-guided needle interventions in the brain. An MR Safe robot was incorporated into an intraoperative MRI system. Deep regions of the brain simulated in a cranial mockup were targeted with a needle under robotic assistance. The 3D accuracy of in-scanner targeting at an average depth of 95mm was 1.55mm, with no manual corrections.