The Impact of Cross-Talk in a Flat Panel Detector on CT Image Quality

Seemeen Karimi;James Hall;Jerel A. Smith;Joseph Tringe
{"title":"The Impact of Cross-Talk in a Flat Panel Detector on CT Image Quality","authors":"Seemeen Karimi;James Hall;Jerel A. Smith;Joseph Tringe","doi":"10.1109/OJIM.2023.3332342","DOIUrl":null,"url":null,"abstract":"Spatial resolution and image noise are two aspects of image quality of an X-ray computed tomography (CT) system and are determined by the X-ray source, the detector, and mathematical operations for image reconstruction. In CT scanners with flat panel detectors (FPDs), there is cross-talk (signal leakage) between detector pixels. The contribution of the cross-talk to spatial resolution and noise in reconstructed images has not been adequately modeled. Previously, we estimated cross-talk from autocovariance measurements in air, and modeled the impact of cross-talk on spatial resolution. We have extended that work to calculate the impact of cross-talk on signal-to-noise ratio in radiographs and to reconstructed image noise. We modeled the spatial resolution and noise of a CT scanner that uses a flat-panel detector with 0.2-mm pixels and a gadolinium oxysulfide scintillator, and a 450 kVp, dual-focus X-ray tube. Our noise model agrees with measurements from experimental data and simulations to within 10%. We show that cross-talk in FPDs can reduce resolution by over 30%, reduce noise by approximately a factor of two, and introduce correlation in the noise, and therefore, cannot be disregarded when assessing CT image quality.","PeriodicalId":100630,"journal":{"name":"IEEE Open Journal of Instrumentation and Measurement","volume":"2 ","pages":"1-9"},"PeriodicalIF":0.0000,"publicationDate":"2023-11-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10317879","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Open Journal of Instrumentation and Measurement","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10317879/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Spatial resolution and image noise are two aspects of image quality of an X-ray computed tomography (CT) system and are determined by the X-ray source, the detector, and mathematical operations for image reconstruction. In CT scanners with flat panel detectors (FPDs), there is cross-talk (signal leakage) between detector pixels. The contribution of the cross-talk to spatial resolution and noise in reconstructed images has not been adequately modeled. Previously, we estimated cross-talk from autocovariance measurements in air, and modeled the impact of cross-talk on spatial resolution. We have extended that work to calculate the impact of cross-talk on signal-to-noise ratio in radiographs and to reconstructed image noise. We modeled the spatial resolution and noise of a CT scanner that uses a flat-panel detector with 0.2-mm pixels and a gadolinium oxysulfide scintillator, and a 450 kVp, dual-focus X-ray tube. Our noise model agrees with measurements from experimental data and simulations to within 10%. We show that cross-talk in FPDs can reduce resolution by over 30%, reduce noise by approximately a factor of two, and introduce correlation in the noise, and therefore, cannot be disregarded when assessing CT image quality.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
平板探测器中的串扰对 CT 图像质量的影响
空间分辨率和图像噪声是x射线计算机断层扫描(CT)系统图像质量的两个方面,由x射线源、探测器和图像重建的数学运算决定。在使用平板检测器(FPDs)的CT扫描仪中,检测器像素之间存在串扰(信号泄漏)。在重建图像中,串扰对空间分辨率和噪声的贡献尚未得到充分的模拟。以前,我们从空气中的自协方差测量中估计串扰,并模拟串扰对空间分辨率的影响。我们已经扩展了这项工作,以计算串扰对x光片信噪比的影响,并重建图像噪声。我们模拟了一台CT扫描仪的空间分辨率和噪声,该扫描仪使用一个0.2 mm像素的平板探测器和一个氧化硫化钆闪烁体,以及一个450 kVp的双聚焦x射线管。我们的噪声模型与实验数据和模拟测量结果的一致性在10%以内。研究表明,fpd中的串扰可以使分辨率降低30%以上,将噪声降低约2倍,并在噪声中引入相关性,因此,在评估CT图像质量时不能忽略串扰。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Spatiotemporal Variance Image Reconstruction for Thermographic Inspections Fault Detection in an Electro-Hydrostatic Actuator Using Polyscale Complexity Measures and Bayesian Classification Baseline-Free Damage Imaging for Structural Health Monitoring of Composite Lap Joint Using Ultrasonic-Guided Waves Spiking Neural Networks for Energy-Efficient Acoustic Emission-Based Monitoring IMU Optimal Rotation Rates
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1