Evaluating the impact of scan body angulation and geometric attachments on the accuracy of complete-arch digital implant impressions: A comparison of two intraoral scanners.

IF 3.4 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry Pub Date : 2025-02-01 Epub Date: 2023-12-08 DOI:10.1111/jopr.13807
Ra'fat I Farah, Bandar Alresheedi, Saad Alazmi, Sanaa N Al-Haj Ali
{"title":"Evaluating the impact of scan body angulation and geometric attachments on the accuracy of complete-arch digital implant impressions: A comparison of two intraoral scanners.","authors":"Ra'fat I Farah, Bandar Alresheedi, Saad Alazmi, Sanaa N Al-Haj Ali","doi":"10.1111/jopr.13807","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To investigate the effects of scan body (SB) angulation and geometric attachments (GAs) on the deviations of complete-arch digital implant impressions obtained using intraoral scanners (IOSs), in relation to the gold-standard desktop scanner.</p><p><strong>Material and methods: </strong>Two IOSs (iTero and Omnicam), two SB angulations (0 degrees and 30 degrees), and GAs for the SBs were investigated. SBs were attached to an edentulous maxillary cast with the following implant analogs: parallel 0-degree at positions #13, #23, and #26, and 30-degree posteriorly tilted at position #16. The cast was digitized using a reference desktop scanner, followed by ten consecutive digital scans using each IOS (five scans with GAs and five without GAs, n = 20). Meshes obtained from the IOS scans were superimposed on the master reference mesh. Linear distance and 3D surface deviations from the reference mesh were calculated. A three-way multivariate analysis of variance (MANOVA) was employed to assess the effects of angulation, IOS type, and GAs on the combined dependent variables (α = 0.05).</p><p><strong>Results: </strong>No significant three-way interaction was observed between IOS type, SB angulation, and GAs for combined deviations from the reference scan (p = 0.56). Simple main effect analysis revealed that iTero exhibited significantly lower mean 3D surface and linear deviations than Omnicam (p < 0.05). Additionally, the use of GAs significantly reduced deviations (p < 0.001), with mean deviation reductions for both scanners ranging from 26-33 micrometers (µm) for 3D deviations and 15-21 µm for linear distance deviations. No differences were found between angled and parallel SBs regarding 3D surface and linear distance deviations (p ≥ 0.05).</p><p><strong>Conclusions: </strong>ITero demonstrated significantly smaller deviations, and the use of GAs led to significantly reduced distance and 3D surface deviations. SB angulations did not impact scan deviations.</p>","PeriodicalId":49152,"journal":{"name":"Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry","volume":" ","pages":"174-181"},"PeriodicalIF":3.4000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jopr.13807","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/8 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To investigate the effects of scan body (SB) angulation and geometric attachments (GAs) on the deviations of complete-arch digital implant impressions obtained using intraoral scanners (IOSs), in relation to the gold-standard desktop scanner.

Material and methods: Two IOSs (iTero and Omnicam), two SB angulations (0 degrees and 30 degrees), and GAs for the SBs were investigated. SBs were attached to an edentulous maxillary cast with the following implant analogs: parallel 0-degree at positions #13, #23, and #26, and 30-degree posteriorly tilted at position #16. The cast was digitized using a reference desktop scanner, followed by ten consecutive digital scans using each IOS (five scans with GAs and five without GAs, n = 20). Meshes obtained from the IOS scans were superimposed on the master reference mesh. Linear distance and 3D surface deviations from the reference mesh were calculated. A three-way multivariate analysis of variance (MANOVA) was employed to assess the effects of angulation, IOS type, and GAs on the combined dependent variables (α = 0.05).

Results: No significant three-way interaction was observed between IOS type, SB angulation, and GAs for combined deviations from the reference scan (p = 0.56). Simple main effect analysis revealed that iTero exhibited significantly lower mean 3D surface and linear deviations than Omnicam (p < 0.05). Additionally, the use of GAs significantly reduced deviations (p < 0.001), with mean deviation reductions for both scanners ranging from 26-33 micrometers (µm) for 3D deviations and 15-21 µm for linear distance deviations. No differences were found between angled and parallel SBs regarding 3D surface and linear distance deviations (p ≥ 0.05).

Conclusions: ITero demonstrated significantly smaller deviations, and the use of GAs led to significantly reduced distance and 3D surface deviations. SB angulations did not impact scan deviations.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估扫描体角度和几何附着物对全弓数字种植印模精度的影响:两种口内扫描仪的比较。
目的:探讨扫描体(SB)角度和几何附着物对口腔内扫描仪(ios)获得的全弓数字种植印模偏差的影响,并与金标准桌面扫描仪进行比较。材料和方法:研究了两种口腔内扫描仪(iTero和Omnicam),两种扫描体角度(0度和30度)和SBs的几何附着体(GAs)。SBs附着在无牙上颌铸造体上,植入物类似物如下:在#13、#23和#26位置平行0度,在#16位置后倾斜30度。使用参考桌面扫描仪对模型进行数字化,然后使用每个IOS进行10次连续数字扫描(5次带GAs扫描,5次不带GAs扫描,n = 20)。从IOS扫描中获得的网格被叠加在主参考网格上。计算了与参考网格的直线距离和三维曲面偏差。采用三因素方差分析(MANOVA)评估角度、IOS类型和GAs对组合因变量的影响(α = 0.05)。结果:IOS类型、SB角度和GAs与参考扫描的综合偏差之间没有显著的三向相互作用(p = 0.56)。简单的主效应分析显示,iTero的平均三维表面和线性偏差明显低于Omnicam (p)。结论:iTero的平均三维表面和线性偏差明显小于Omnicam,使用GAs可以显著减少距离和三维表面偏差。SB角度不影响扫描偏差。这篇文章受版权保护。版权所有。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
7.90
自引率
15.00%
发文量
171
审稿时长
6-12 weeks
期刊介绍: The Journal of Prosthodontics promotes the advanced study and practice of prosthodontics, implant, esthetic, and reconstructive dentistry. It is the official journal of the American College of Prosthodontists, the American Dental Association-recognized voice of the Specialty of Prosthodontics. The journal publishes evidence-based original scientific articles presenting information that is relevant and useful to prosthodontists. Additionally, it publishes reports of innovative techniques, new instructional methodologies, and instructive clinical reports with an interdisciplinary flair. The journal is particularly focused on promoting the study and use of cutting-edge technology and positioning prosthodontists as the early-adopters of new technology in the dental community.
期刊最新文献
Nanocomposites soft and hard denture liners containing ZrO2 nanoparticles: An in vitro evaluation of surface roughness and Candida albicans adhesion. Effect of coffee thermal cycling on the surface properties and stainability of additively manufactured denture base resins in different layer thicknesses. Effects of screw channel angulation on the fracture resistance of one-piece zirconia crown with titanium base. Evaluating the impact of scan body angulation and geometric attachments on the accuracy of complete-arch digital implant impressions: A comparison of two intraoral scanners. Effect of different preparation designs of minimally invasive occlusal onlays on the accuracy of different intraoral scanners: An in vitro study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1