Validation of the oscillometric home blood pressure monitor Braun BUA4000 with wide-range cuff in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.

IF 1.2 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE Blood Pressure Monitoring Pub Date : 2024-04-01 Epub Date: 2023-11-15 DOI:10.1097/MBP.0000000000000684
Vasileios Ntousopoulos, Angeliki Ntineri, Aikaterini Theodosiadi, Panagiota Stathopoulou, Ariadni Menti, Konstantinos G Kyriakoulis, Anastasios Kollias, George S Stergiou
{"title":"Validation of the oscillometric home blood pressure monitor Braun BUA4000 with wide-range cuff in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.","authors":"Vasileios Ntousopoulos, Angeliki Ntineri, Aikaterini Theodosiadi, Panagiota Stathopoulou, Ariadni Menti, Konstantinos G Kyriakoulis, Anastasios Kollias, George S Stergiou","doi":"10.1097/MBP.0000000000000684","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the blood pressure (BP) measurement accuracy of the Braun BUA4000 automated oscillometric upper-arm cuff device for self-home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01.</p><p><strong>Methods: </strong>Participants were recruited to fulfill the age, sex, BP and limb circumference distribution criteria of the AAMI/ESH/ISO Universal Standard in a general population using the same arm sequential BP measurement method. A single wide-range cuff of the test device was used for arm circumference range 22-42 cm.</p><p><strong>Results: </strong>A total of 98 individuals were recruited and 85 were analyzed [mean age 60.3 ± 16.1 (SD) years, 44 men, arm circumference 31.5 ± 5.1 cm, range 22-41.5 cm]. For validation criterion 1, the mean difference ± SD between the test device and reference BP readings ( N  = 255) was 0.9 ± 6.4/-0.3 ± 6.4 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual ( N  = 85) was 5.15/5.81 mmHg (systolic/diastolic; threshold ≤6.88/6.95 mmHg).</p><p><strong>Conclusion: </strong>The Braun BUA4000 automated oscillometric BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard in a general population and can be recommended for self-monitoring of BP by patients at home.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"99-102"},"PeriodicalIF":1.2000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000684","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/11/15 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To evaluate the blood pressure (BP) measurement accuracy of the Braun BUA4000 automated oscillometric upper-arm cuff device for self-home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01.

Methods: Participants were recruited to fulfill the age, sex, BP and limb circumference distribution criteria of the AAMI/ESH/ISO Universal Standard in a general population using the same arm sequential BP measurement method. A single wide-range cuff of the test device was used for arm circumference range 22-42 cm.

Results: A total of 98 individuals were recruited and 85 were analyzed [mean age 60.3 ± 16.1 (SD) years, 44 men, arm circumference 31.5 ± 5.1 cm, range 22-41.5 cm]. For validation criterion 1, the mean difference ± SD between the test device and reference BP readings ( N  = 255) was 0.9 ± 6.4/-0.3 ± 6.4 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual ( N  = 85) was 5.15/5.81 mmHg (systolic/diastolic; threshold ≤6.88/6.95 mmHg).

Conclusion: The Braun BUA4000 automated oscillometric BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard in a general population and can be recommended for self-monitoring of BP by patients at home.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
根据医疗器械进步协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)通用标准,在普通人群中验证带有宽范围袖带的家庭血压计Braun BUA4000。
目的:根据美国医疗器械发展协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)通用标准(ISO 81060-2:2018)及其1.2020-01修订版,评价Braun BUA4000自家用上臂袖带自动测血压仪在普通人群中的血压测量精度。方法:招募符合AAMI/ESH/ISO通用标准的年龄、性别、血压和肢体围度分布标准的一般人群,采用相同的臂序贯血压测量方法。臂围范围22-42 cm,采用单套宽量程袖带进行测试。结果:共纳入98例,分析85例[平均年龄60.3±16.1 (SD)岁,男性44例,臂围31.5±5.1 cm,范围22 ~ 41.5 cm]。对于验证标准1,测试装置与参考血压读数(N = 255)之间的平均差值±SD为0.9±6.4/-0.3±6.4 mmHg(收缩压/舒张压;阈值≤5±8mmhg)。对于标准2,每个个体(N = 85)的测试装置和参考血压之间的平均血压差的标准差为5.15/5.81 mmHg(收缩压/舒张压;阈值≤6.88/6.95 mmHg)。结论:博朗BUA4000全自动血压监测仪在普通人群中满足AAMI/ESH/ISO通用标准的要求,可推荐患者在家自我监测血压。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
期刊最新文献
Validation of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the ISO 81060-2:2018/AMD 1:2020 protocol. Validation of the JAMR upper-arm pressure monitor B23 according to the AAMI/ESH/ISO Universal Standard (ISO 81060- 2:2018/AMD 1:2020). Education differences in blood pressure trajectories by sex through midlife: Findings from the National Health and Nutrition Examination Survey, 1999-2018. Unattended versus conventional blood pressure measurements in hospitalized hypertensive patients. Validation of TMB-2285-BT wrist blood pressure monitor in general population according to the ISO 81060-2:2018 + Amd.1:2020 Protocol.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1