Aryan Niknam Maleki, Alexander Thompson, M. Runciman, Julia Murray, G. Mylonas
{"title":"A soft hydraulic endorectal actuator for prostate radiotherapy","authors":"Aryan Niknam Maleki, Alexander Thompson, M. Runciman, Julia Murray, G. Mylonas","doi":"10.1109/ROBIO58561.2023.10355037","DOIUrl":null,"url":null,"abstract":"Despite advances in radiotherapy, motion error remains a challenge in prostate radiotherapy. Rectal obturators and endorectal balloons may reduce motion error and improve outcomes but have limitations. We aimed to create a deployable rectal obturator with precise angle control to personalise to a patient’s rectal anatomy, by using an antagonistic pair of \"muscle\" actuators to flex and extend the device. Results on deployability, angle control, and radial stiffness are presented here. The device can be compressed down to 16 x 3 x 91 mm, and be deployed to maximum dimensions of 24 x 25.5 x 77 mm. The device provides radial stiffness that may be sufficient to stabilise the rectum during radiotherapy. Angle control can be achieved with an average change of 7.5°/ml inflation in the extensor actuator.","PeriodicalId":505134,"journal":{"name":"2023 IEEE International Conference on Robotics and Biomimetics (ROBIO)","volume":"49 3","pages":"1-6"},"PeriodicalIF":0.0000,"publicationDate":"2023-12-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2023 IEEE International Conference on Robotics and Biomimetics (ROBIO)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ROBIO58561.2023.10355037","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Despite advances in radiotherapy, motion error remains a challenge in prostate radiotherapy. Rectal obturators and endorectal balloons may reduce motion error and improve outcomes but have limitations. We aimed to create a deployable rectal obturator with precise angle control to personalise to a patient’s rectal anatomy, by using an antagonistic pair of "muscle" actuators to flex and extend the device. Results on deployability, angle control, and radial stiffness are presented here. The device can be compressed down to 16 x 3 x 91 mm, and be deployed to maximum dimensions of 24 x 25.5 x 77 mm. The device provides radial stiffness that may be sufficient to stabilise the rectum during radiotherapy. Angle control can be achieved with an average change of 7.5°/ml inflation in the extensor actuator.
尽管放疗技术不断进步,但运动误差仍是前列腺放疗的一大难题。直肠闭塞器和肛门直肠内球囊可减少运动误差并改善治疗效果,但也有局限性。我们的目标是通过使用一对拮抗的 "肌肉 "致动器来弯曲和伸展装置,创造出一种具有精确角度控制功能的可展开直肠闭锁器,以便根据患者的直肠解剖结构进行个性化设计。本文介绍了该装置的可展开性、角度控制和径向刚度结果。该装置可压缩至 16 x 3 x 91 毫米,展开后的最大尺寸为 24 x 25.5 x 77 毫米。该装置提供的径向硬度足以在放疗期间稳定直肠。角度控制可通过伸展致动器 7.5°/ml 的平均充气变化来实现。