Validation of an automated sample preparation module directly connected to LC-MS/MS (CLAM-LC-MS/MS system) and comparison with conventional immunoassays for quantitation of tacrolimus and cyclosporin A in a clinical setting.

IF 1.2 Q4 PHARMACOLOGY & PHARMACY Journal of Pharmaceutical Health Care and Sciences Pub Date : 2024-01-08 DOI:10.1186/s40780-023-00318-6
Tsutomu Shimada, Daisuke Kawakami, Arimi Fujita, Rintaro Yamamoto, Satoshi Hara, Kiyoaki Ito, Ichiro Mizushima, Shinji Kitajima, Yasunori Iwata, Norihiko Sakai, Mitsuhiro Kawano, Takashi Wada, Yoshimichi Sai
{"title":"Validation of an automated sample preparation module directly connected to LC-MS/MS (CLAM-LC-MS/MS system) and comparison with conventional immunoassays for quantitation of tacrolimus and cyclosporin A in a clinical setting.","authors":"Tsutomu Shimada, Daisuke Kawakami, Arimi Fujita, Rintaro Yamamoto, Satoshi Hara, Kiyoaki Ito, Ichiro Mizushima, Shinji Kitajima, Yasunori Iwata, Norihiko Sakai, Mitsuhiro Kawano, Takashi Wada, Yoshimichi Sai","doi":"10.1186/s40780-023-00318-6","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Therapeutic drug monitoring (TDM) systems generally use either liquid chromatography/tandem mass spectrometry (LC-MS/MS) or immunoassay, though both methodologies have disadvantages. In this study, we aimed to evaluate whether a CLAM-LC-MS/MS system, which consists of a sample preparation module directly connected to LC-MS/MS, could be used for clinical TDM work for immunosuppressive drugs in whole blood, which requires a hemolytic process. For this purpose, we prospectively validated this system for clinical measurement of tacrolimus and cyclosporin A in patients' whole blood. The results were also compared with those of commercial immunoassays.</p><p><strong>Methods: </strong>Whole blood from patients treated with tacrolimus or cyclosporin A at the Department of Nephrology and Departments of Rheumatology, Kanazawa University Hospital, from May 2018 to July 2019 was collected with informed consent, and drug concentrations were measured by CLAM-LC-MS/MS and by chemiluminescence immunoassay (CLIA) for tacrolimus and affinity column-mediated immunoassay (ACMIA) for cyclosporin A. Correlations between the CLAM-LC-MS/MS and immunoassay results were analyzed.</p><p><strong>Results: </strong>Two hundred and twenty-four blood samples from 80 patients were used for tacrolimus measurement, and 76 samples from 21 patients were used for cyclosporin A. Intra- and inter-assay precision values of quality controls were less than 7%. There were significant correlations between CLAM-LC-MS/MS and the immunoassays for tacrolimus and cyclosporin A (Spearman rank correlation coefficients: 0.861, 0.941, P < 0.00001 in each case). The drug concentrations measured by CLAM-LC-MS/MS were about 20% lower than those obtained using the immunoassays. CLAM-LC-MS/MS maintenance requirements did not interfere with clinical operations. Compared to manual pretreatment, automated pretreatment by CLAM showed lower inter-assay precision values and greatly reduced the pretreatment time.</p><p><strong>Conclusions: </strong>The results obtained by CLAM-LC-MS/MS were highly correlated with those of commercial immunoassay methods. CLAM-LC-MS/MS offers advantages in clinical TDM practice, including simple, automatic pretreatment, low maintenance requirement, and avoidance of interference.</p>","PeriodicalId":16730,"journal":{"name":"Journal of Pharmaceutical Health Care and Sciences","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-01-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10773076/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Pharmaceutical Health Care and Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s40780-023-00318-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PHARMACOLOGY & PHARMACY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Therapeutic drug monitoring (TDM) systems generally use either liquid chromatography/tandem mass spectrometry (LC-MS/MS) or immunoassay, though both methodologies have disadvantages. In this study, we aimed to evaluate whether a CLAM-LC-MS/MS system, which consists of a sample preparation module directly connected to LC-MS/MS, could be used for clinical TDM work for immunosuppressive drugs in whole blood, which requires a hemolytic process. For this purpose, we prospectively validated this system for clinical measurement of tacrolimus and cyclosporin A in patients' whole blood. The results were also compared with those of commercial immunoassays.

Methods: Whole blood from patients treated with tacrolimus or cyclosporin A at the Department of Nephrology and Departments of Rheumatology, Kanazawa University Hospital, from May 2018 to July 2019 was collected with informed consent, and drug concentrations were measured by CLAM-LC-MS/MS and by chemiluminescence immunoassay (CLIA) for tacrolimus and affinity column-mediated immunoassay (ACMIA) for cyclosporin A. Correlations between the CLAM-LC-MS/MS and immunoassay results were analyzed.

Results: Two hundred and twenty-four blood samples from 80 patients were used for tacrolimus measurement, and 76 samples from 21 patients were used for cyclosporin A. Intra- and inter-assay precision values of quality controls were less than 7%. There were significant correlations between CLAM-LC-MS/MS and the immunoassays for tacrolimus and cyclosporin A (Spearman rank correlation coefficients: 0.861, 0.941, P < 0.00001 in each case). The drug concentrations measured by CLAM-LC-MS/MS were about 20% lower than those obtained using the immunoassays. CLAM-LC-MS/MS maintenance requirements did not interfere with clinical operations. Compared to manual pretreatment, automated pretreatment by CLAM showed lower inter-assay precision values and greatly reduced the pretreatment time.

Conclusions: The results obtained by CLAM-LC-MS/MS were highly correlated with those of commercial immunoassay methods. CLAM-LC-MS/MS offers advantages in clinical TDM practice, including simple, automatic pretreatment, low maintenance requirement, and avoidance of interference.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
验证直接连接到 LC-MS/MS 系统(CLAM-LC-MS/MS 系统)的自动样品制备模块,并将其与传统的免疫测定法进行比较,用于在临床环境中定量检测他克莫司和环孢素 A。
背景:治疗药物监测(TDM)系统通常使用液相色谱/串联质谱法(LC-MS/MS)或免疫测定法,但这两种方法都有缺点。在本研究中,我们旨在评估由直接连接到 LC-MS/MS 的样品制备模块组成的 CLAM-LC-MS/MS 系统是否可用于临床 TDM 工作,以检测需要溶血过程的全血中的免疫抑制剂。为此,我们对该系统进行了前瞻性验证,用于临床检测患者全血中的他克莫司和环孢素 A。我们还将结果与商用免疫测定法进行了比较:2018年5月至2019年7月期间,金泽大学医院肾脏内科和风湿病科收集了接受他克莫司或环孢素A治疗的患者的全血,并在知情同意的情况下,通过CLAM-LC-MS/MS和化学发光免疫测定(CLIA)测量了他克莫司的药物浓度,以及亲和柱介导免疫测定(ACMIA)测量了环孢素A的药物浓度。分析了 CLAM-LC-MS/MS 和免疫测定结果之间的相关性:对 80 名患者的 224 份血样进行了他克莫司测定,对 21 名患者的 76 份血样进行了环孢素 A 测定。CLAM-LC-MS/MS 与他克莫司和环孢素 A 的免疫测定之间存在明显的相关性(斯皮尔曼秩相关系数:0.861,0.9%):0.861, 0.941, P 结论:CLAM-LC-MS/MS 与商业免疫测定方法得出的结果高度相关。CLAM-LC-MS/MS 在临床 TDM 实践中具有简单、自动预处理、维护要求低和避免干扰等优势。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.80
自引率
0.00%
发文量
29
审稿时长
8 weeks
期刊最新文献
Compatibility of hypokalaemia caused by low-dose prednisolone plus abiraterone acetate therapy for metastatic castration-resistant prostate cancer. A qualitative study on the current status and problems of pharmacists in home healthcare from the viewpoint of care managers in medically underpopulated areas in Japan. Effects of famotidine use during pregnancy: an observational cohort study. Development and validation of the prediction score for augmented renal clearance in critically Ill Japanese adults. Investigating the hypothermic effects of fluoroquinolone antimicrobials on non-bacterial fever model mice.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1