Siti Hajar Zuber, Muhammad Fahmi Rizal Abdul Hadi, Damilola Oluwafemi Samson, Jayapramila Jayamani, Nor Ain Rabaiee, Mohd Zahri Abdul Aziz, Nurul Ab Aziz Hashikin, Chee Keat Ying, Mohd Fahmi Mohd Yusof, Rokiah Hashim
{"title":"Dosimetric Analysis of <i>Rhizophora</i>-based Phantom Material in Radiation Therapy Applications Using Monte Carlo GATE Simulation.","authors":"Siti Hajar Zuber, Muhammad Fahmi Rizal Abdul Hadi, Damilola Oluwafemi Samson, Jayapramila Jayamani, Nor Ain Rabaiee, Mohd Zahri Abdul Aziz, Nurul Ab Aziz Hashikin, Chee Keat Ying, Mohd Fahmi Mohd Yusof, Rokiah Hashim","doi":"10.4103/jmp.jmp_75_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aims to determine the percentage depth dose (PDD) of a phantom material made from soy-lignin bonded <i>Rhizophora</i> spp. particleboard coated with a gloss finish by using Monte Carlo Geant4 Application for Tomographic Emission (GATE) simulation.</p><p><strong>Materials and methods: </strong>The particleboard was fabricated using a hot pressing technique at target density of 1.0 g·cm<sup>-3</sup> and the elemental fraction was recorded for the simulation. The PDD was simulated in the GATE simulation using the linear accelerator Elekta Synergy model for the water phantom and <i>Rhizophora</i> phantom, and the results were compared with the experimental PDD performed by several studies. Beam flatness and beam symmetry were also measured in this study.</p><p><strong>Results: </strong>The simulated PDD for <i>Rhizophora</i> and water was in agreement with the experimental PDD of water with overall discrepancies of 0% to 8.7% at depth ranging from 1.0 to 15.0 cm. In the GATE simulation, all the points passed the clinical 3%/3 mm criterion in comparison with water, with the final percentage of 2.34% for <i>Rhizophora</i> phantom and 2.49% for the water phantom simulated in GATE. Both the symmetries are all within the range of an acceptable value of 2.0% according to the recommendation, with the beam symmetry of the water phantom and <i>Rhizophora</i> phantom at 0.58% and 0.28%, respectively.</p><p><strong>Conclusions: </strong>The findings of this study provide the necessary foundation to confidently use the phantom for radiotherapy purposes, especially in treatment planning.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"48 4","pages":"358-364"},"PeriodicalIF":0.7000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10783191/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_75_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/5 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: This study aims to determine the percentage depth dose (PDD) of a phantom material made from soy-lignin bonded Rhizophora spp. particleboard coated with a gloss finish by using Monte Carlo Geant4 Application for Tomographic Emission (GATE) simulation.
Materials and methods: The particleboard was fabricated using a hot pressing technique at target density of 1.0 g·cm-3 and the elemental fraction was recorded for the simulation. The PDD was simulated in the GATE simulation using the linear accelerator Elekta Synergy model for the water phantom and Rhizophora phantom, and the results were compared with the experimental PDD performed by several studies. Beam flatness and beam symmetry were also measured in this study.
Results: The simulated PDD for Rhizophora and water was in agreement with the experimental PDD of water with overall discrepancies of 0% to 8.7% at depth ranging from 1.0 to 15.0 cm. In the GATE simulation, all the points passed the clinical 3%/3 mm criterion in comparison with water, with the final percentage of 2.34% for Rhizophora phantom and 2.49% for the water phantom simulated in GATE. Both the symmetries are all within the range of an acceptable value of 2.0% according to the recommendation, with the beam symmetry of the water phantom and Rhizophora phantom at 0.58% and 0.28%, respectively.
Conclusions: The findings of this study provide the necessary foundation to confidently use the phantom for radiotherapy purposes, especially in treatment planning.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.