Endovascular navigation in patients: vessel-based registration of electromagnetic tracking to preoperative images

Erik Nypan, Geir Arne Tangen, Reidar Brekken, Petter Aadahl, F. Manstad-Hulaas
{"title":"Endovascular navigation in patients: vessel-based registration of electromagnetic tracking to preoperative images","authors":"Erik Nypan, Geir Arne Tangen, Reidar Brekken, Petter Aadahl, F. Manstad-Hulaas","doi":"10.3389/fradi.2024.1320535","DOIUrl":null,"url":null,"abstract":"Electromagnetic tracking of instruments combined with preoperative images can supplement fluoroscopy for guiding endovascular aortic repair (EVAR). The aim of this study was to evaluate the in-vivo accuracy of a vessel-based registration algorithm for matching electromagnetically tracked positions of an endovascular instrument to preoperative computed tomography angiography. Five patients undergoing elective EVAR were included, and a clinically available semi-automatic 3D–3D registration algorithm, based on similarity measures computed over the entire image, was used for reference. Accuracy was reported as target registration error (TRE) evaluated in manually selected anatomic landmarks on bony structures, placed close to the volume-of-interest. The median TRE was 8.2 mm (range: 7.1 mm to 16.1 mm) for the vessel-based registration algorithm, compared to 2.2 mm (range: 1.8 mm to 3.7 mm) for the reference algorithm. This illustrates that registration based on intraoperative electromagnetic tracking is feasible, but the accuracy must be improved before clinical use.","PeriodicalId":73101,"journal":{"name":"Frontiers in radiology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-01-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in radiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fradi.2024.1320535","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Electromagnetic tracking of instruments combined with preoperative images can supplement fluoroscopy for guiding endovascular aortic repair (EVAR). The aim of this study was to evaluate the in-vivo accuracy of a vessel-based registration algorithm for matching electromagnetically tracked positions of an endovascular instrument to preoperative computed tomography angiography. Five patients undergoing elective EVAR were included, and a clinically available semi-automatic 3D–3D registration algorithm, based on similarity measures computed over the entire image, was used for reference. Accuracy was reported as target registration error (TRE) evaluated in manually selected anatomic landmarks on bony structures, placed close to the volume-of-interest. The median TRE was 8.2 mm (range: 7.1 mm to 16.1 mm) for the vessel-based registration algorithm, compared to 2.2 mm (range: 1.8 mm to 3.7 mm) for the reference algorithm. This illustrates that registration based on intraoperative electromagnetic tracking is feasible, but the accuracy must be improved before clinical use.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
患者的血管内导航:基于血管的电磁跟踪与术前图像配准
电磁追踪器械结合术前图像可作为透视检查的补充,用于指导血管内主动脉修复术(EVAR)。本研究旨在评估一种基于血管的套准算法的体内准确性,该算法用于将电磁跟踪的血管内器械位置与术前计算机断层扫描血管造影相匹配。其中包括五名接受择期 EVAR 的患者,并使用临床上可用的半自动 3D-3D 配准算法作为参考,该算法基于对整个图像计算的相似性度量。精确度以目标配准误差(TRE)来报告,目标配准误差是在靠近感兴趣容积的骨性结构上手动选择的解剖地标进行评估的。基于血管的配准算法的目标配准误差中值为 8.2 毫米(范围:7.1 毫米至 16.1 毫米),而参考算法的目标配准误差中值为 2.2 毫米(范围:1.8 毫米至 3.7 毫米)。这说明基于术中电磁追踪的配准是可行的,但在临床使用前必须提高准确性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.20
自引率
0.00%
发文量
0
期刊最新文献
Wideband radiofrequency pulse sequence for evaluation of myocardial scar in patients with cardiac implantable devices. Value of interventional radiology and their contributions to modern medical systems Feasibility study to unveil the potential: considerations of constrained spherical deconvolution tractography with unsedated neonatal diffusion brain MRI data. Automated intracranial vessel segmentation of 4D flow MRI data in patients with atherosclerotic stenosis using a convolutional neural network Standardized evaluation of the extent of resection in glioblastoma with automated early post-operative segmentation
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1