{"title":"Innovative device for surgical wound irrigation: a preclinical testing and pilot clinical validation study.","authors":"Yoshinori Hayashi, Takehiro Noda, Yasutaka Samizo, Kenji Fujimoto, Eiji Uemoto, Kotaro Yamashita, Takuro Saito, Koji Tanaka, Kazuyoshi Yamamoto, Tomoki Makino, Tsuyoshi Takahashi, Yukinori Kurokawa, Shogo Kobayashi, Hidetoshi Eguchi, Yuichiro Doki, Kiyokazu Nakajima","doi":"10.1080/13645706.2024.2317196","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Surgical site infection (SSI) poses a substantial postoperative challenge, affecting patient recovery and healthcare costs. While surgical wound irrigation is pivotal in SSI reduction, consensus on the optimal method remains elusive. We developed a novel device for surgical wound irrigation and conducted preclinical and clinical evaluations to evaluate its efficacy and safety.</p><p><strong>Methods: </strong>Two preclinical experiments using swine were performed. In the washability test, two contaminated wound model were established, and the cleansing rate between the device and the conventional method were compared. In the contamination test, the irrigation procedure with a fluorescent solution assessed the surrounding contamination of drapes. Subsequently, a clinical trial involving patients undergoing abdominal surgery was conducted.</p><p><strong>Results: </strong>The washability test demonstrated significantly higher cleansing rates with the device method (86.4% and 82.5%) compared to the conventional method (65.2% and 65.1%) in two contamination models. The contamination test revealed a smaller contaminated region with the device method than the conventional method. In the clinical trial involving 17 abdominal surgery cases, no superficial SSIs or adverse events related to device use were observed.</p><p><strong>Conclusions: </strong>Our newly developed device exhibits potential for achieving more effective and safe SSI control compared to conventional wound irrigation.</p>","PeriodicalId":18537,"journal":{"name":"Minimally Invasive Therapy & Allied Technologies","volume":" ","pages":"200-206"},"PeriodicalIF":1.7000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Minimally Invasive Therapy & Allied Technologies","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/13645706.2024.2317196","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/18 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"SURGERY","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Surgical site infection (SSI) poses a substantial postoperative challenge, affecting patient recovery and healthcare costs. While surgical wound irrigation is pivotal in SSI reduction, consensus on the optimal method remains elusive. We developed a novel device for surgical wound irrigation and conducted preclinical and clinical evaluations to evaluate its efficacy and safety.
Methods: Two preclinical experiments using swine were performed. In the washability test, two contaminated wound model were established, and the cleansing rate between the device and the conventional method were compared. In the contamination test, the irrigation procedure with a fluorescent solution assessed the surrounding contamination of drapes. Subsequently, a clinical trial involving patients undergoing abdominal surgery was conducted.
Results: The washability test demonstrated significantly higher cleansing rates with the device method (86.4% and 82.5%) compared to the conventional method (65.2% and 65.1%) in two contamination models. The contamination test revealed a smaller contaminated region with the device method than the conventional method. In the clinical trial involving 17 abdominal surgery cases, no superficial SSIs or adverse events related to device use were observed.
Conclusions: Our newly developed device exhibits potential for achieving more effective and safe SSI control compared to conventional wound irrigation.
期刊介绍:
Minimally Invasive Therapy and Allied Technologies (MITAT) is an international forum for endoscopic surgeons, interventional radiologists and industrial instrument manufacturers. It is the official journal of the Society for Medical Innovation and Technology (SMIT) whose membership includes representatives from a broad spectrum of medical specialities, instrument manufacturing and research. The journal brings the latest developments and innovations in minimally invasive therapy to its readers. What makes Minimally Invasive Therapy and Allied Technologies unique is that we publish one or two special issues each year, which are devoted to a specific theme. Key topics covered by the journal include: interventional radiology, endoscopic surgery, imaging technology, manipulators and robotics for surgery and education and training for MIS.