Claudio Fania, Antonella Giletto, Alice Niero, Paolo Palatini
{"title":"Accuracy of the KOROT P3 Accurate automated auscultatory blood pressure measuring device for professional use in people with extra-large arms.","authors":"Claudio Fania, Antonella Giletto, Alice Niero, Paolo Palatini","doi":"10.1097/MBP.0000000000000696","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To determine the accuracy of the KOROT P3 Accurate (previously InBody BPBIO480KV) monitor, an automated auscultatory blood pressure (BP) measuring device developed for professional use, in people with extra-large arms according to the ISO81060-2 2018 protocol.</p><p><strong>Methods: </strong>The KOROT P3 Accurate was tested in 37 subjects with upper-arm circumference ranging from >42 to 53 cm using a mercury sphygmomanometer coupled to a 20 × 40 cm tronco-conical cuff as the reference standard.</p><p><strong>Results: </strong>The mean BP difference between the device and the observers' reference measurements was 1.2 ± 2.0 mmHg for systolic BP and 1.0 ± 2.0 mmHg for diastolic BP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also criterion 2 was satisfied being the standard deviations ± 1.7 mmHg for systolic BP and ± 1.6 for diastolic BP, well below the maximum values required by the protocol (±6.84/6.87 mmHg). Scatterplots of device-reference systolic and diastolic BP differences showed similar accuracy across the range of participants' BP, arm circumference and upper-arm slant angle.</p><p><strong>Conclusions: </strong>These data show that the KOROT P3 Accurate monitor satisfied the ISO 81060-2:2018 standard requirements in a special population of people with extra-large arms ranging from >42 to 53 cm.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000696","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/23 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To determine the accuracy of the KOROT P3 Accurate (previously InBody BPBIO480KV) monitor, an automated auscultatory blood pressure (BP) measuring device developed for professional use, in people with extra-large arms according to the ISO81060-2 2018 protocol.
Methods: The KOROT P3 Accurate was tested in 37 subjects with upper-arm circumference ranging from >42 to 53 cm using a mercury sphygmomanometer coupled to a 20 × 40 cm tronco-conical cuff as the reference standard.
Results: The mean BP difference between the device and the observers' reference measurements was 1.2 ± 2.0 mmHg for systolic BP and 1.0 ± 2.0 mmHg for diastolic BP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also criterion 2 was satisfied being the standard deviations ± 1.7 mmHg for systolic BP and ± 1.6 for diastolic BP, well below the maximum values required by the protocol (±6.84/6.87 mmHg). Scatterplots of device-reference systolic and diastolic BP differences showed similar accuracy across the range of participants' BP, arm circumference and upper-arm slant angle.
Conclusions: These data show that the KOROT P3 Accurate monitor satisfied the ISO 81060-2:2018 standard requirements in a special population of people with extra-large arms ranging from >42 to 53 cm.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.