Is fatigue mechanism implicated in intraoral fracture of narrow dental implants? A thorough retrieval analysis of two failed implant fixtures retrieved from a single patient
Youssef S. Al Jabbari , Raymond Fournelle , Arif Sufyan , Spiros Zinelis
{"title":"Is fatigue mechanism implicated in intraoral fracture of narrow dental implants? A thorough retrieval analysis of two failed implant fixtures retrieved from a single patient","authors":"Youssef S. Al Jabbari , Raymond Fournelle , Arif Sufyan , Spiros Zinelis","doi":"10.1016/j.sdentj.2024.02.006","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>This study aimed to perform a thorough failure analysis of two fractured narrow dental implants after medium-term in vivo use.</p></div><div><h3>Materials and methods</h3><p>The top parts of two fractured Narrow Dental Implant (NDI) fixtures were retrieved from two different locations at two different times from the same patient. The NDI-specimen-1 was 12-months in service while the NDI-specimen-2 was 17-months in service. In both cases, the top parts of the fractured NDI fixtures that were attached to prosthetic components were retrieved and subjected to thorough, non-destructive and destructive testing.</p></div><div><h3>Results</h3><p>Light Microscopy <strong>(</strong>LM) and Scanning Electron Microscopy (SEM) revealed that both the retrieved fractured NDIs failed because of fatigue, characterized by beach and ratchet marks. Macroscopic examination revealed that fatigue cracks initiated at the internal thread surfaces of the implants and propagated around them until final fracture. Both samples fractured near the end of the retaining screw and followed the root of the internal thread. Optical and SEM analyses revealed a uniform distribution of irregularly shaped grains (diameter = 2 to 5 μm). X -ray Energy Dispersive Spectroscopy (EDS) analysis showed that the NDI-specimen-1 was made using Ti-14%Zr with a Vickers Hardens (HV) of 288 ± 5.</p></div><div><h3>Conclusion</h3><p>Since the fracture occurred by a fatigue; thus, an increase in fatigue resistance will be beneficial for the longevity of NDI.</p></div>","PeriodicalId":47246,"journal":{"name":"Saudi Dental Journal","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1013905224000580/pdfft?md5=2e7a751f955d5a7b36eb513d57c5ae80&pid=1-s2.0-S1013905224000580-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Saudi Dental Journal","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1013905224000580","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
This study aimed to perform a thorough failure analysis of two fractured narrow dental implants after medium-term in vivo use.
Materials and methods
The top parts of two fractured Narrow Dental Implant (NDI) fixtures were retrieved from two different locations at two different times from the same patient. The NDI-specimen-1 was 12-months in service while the NDI-specimen-2 was 17-months in service. In both cases, the top parts of the fractured NDI fixtures that were attached to prosthetic components were retrieved and subjected to thorough, non-destructive and destructive testing.
Results
Light Microscopy (LM) and Scanning Electron Microscopy (SEM) revealed that both the retrieved fractured NDIs failed because of fatigue, characterized by beach and ratchet marks. Macroscopic examination revealed that fatigue cracks initiated at the internal thread surfaces of the implants and propagated around them until final fracture. Both samples fractured near the end of the retaining screw and followed the root of the internal thread. Optical and SEM analyses revealed a uniform distribution of irregularly shaped grains (diameter = 2 to 5 μm). X -ray Energy Dispersive Spectroscopy (EDS) analysis showed that the NDI-specimen-1 was made using Ti-14%Zr with a Vickers Hardens (HV) of 288 ± 5.
Conclusion
Since the fracture occurred by a fatigue; thus, an increase in fatigue resistance will be beneficial for the longevity of NDI.
期刊介绍:
Saudi Dental Journal is an English language, peer-reviewed scholarly publication in the area of dentistry. Saudi Dental Journal publishes original research and reviews on, but not limited to: • dental disease • clinical trials • dental equipment • new and experimental techniques • epidemiology and oral health • restorative dentistry • periodontology • endodontology • prosthodontics • paediatric dentistry • orthodontics and dental education Saudi Dental Journal is the official publication of the Saudi Dental Society and is published by King Saud University in collaboration with Elsevier and is edited by an international group of eminent researchers.