{"title":"Performance Comparison of DOI-Encoding PET Detectors Based on 1.1-mm Pitch BGO Arrays With Different Reflectors","authors":"Junwei Du;Shixian Du","doi":"10.1109/TRPMS.2024.3361891","DOIUrl":null,"url":null,"abstract":"Bismuth germanate (BGO)-based positron emission tomography (PET) detectors are potential candidates for low-dose imaging PET scanners, owing to the high stopping power and low background radiation of BGO. In this article, we compared the performance of two dual-ended readout PET detectors based on \n<inline-formula> <tex-math>$15\\times15$ </tex-math></inline-formula>\n BGO arrays. Both arrays had the same 1.1 mm pitch but utilized different reflectors—barium sulfate (BaSO4) and enhanced specular reflector film (ESR)—for high-resolution PET applications. The detectors were constructed with Hamamatsu 13361–2050-08 silicon photomultiplier arrays. Each BGO element had dimensions of \n<inline-formula> <tex-math>$1.02\\times 1.02\\times20$ </tex-math></inline-formula>\n mm3. The lateral surfaces of the BGO elements were unpolished (saw-cut), while the two ends were polished. Flood histograms showed that the detector based on the BGO array with BaSO4 reflector had much better crystal identification and depth-of-interaction (DOI) resolution. Specifically, the energy, DOI, and timing resolutions for the detector using the BGO array with BaSO4 reflector were 19.8 ± 1.5%, 4.13 ± 0.48 mm, and 2.80 ± 0.23 ns, respectively. In contrast, the values obtained using the BGO array with ESR reflector were 20.9 ± 2.1%, 7.69 ± 1.92 mm, and 2.93 ± 0.20 ns, respectively.","PeriodicalId":46807,"journal":{"name":"IEEE Transactions on Radiation and Plasma Medical Sciences","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-02-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Radiation and Plasma Medical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10422720/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Bismuth germanate (BGO)-based positron emission tomography (PET) detectors are potential candidates for low-dose imaging PET scanners, owing to the high stopping power and low background radiation of BGO. In this article, we compared the performance of two dual-ended readout PET detectors based on
$15\times15$
BGO arrays. Both arrays had the same 1.1 mm pitch but utilized different reflectors—barium sulfate (BaSO4) and enhanced specular reflector film (ESR)—for high-resolution PET applications. The detectors were constructed with Hamamatsu 13361–2050-08 silicon photomultiplier arrays. Each BGO element had dimensions of
$1.02\times 1.02\times20$
mm3. The lateral surfaces of the BGO elements were unpolished (saw-cut), while the two ends were polished. Flood histograms showed that the detector based on the BGO array with BaSO4 reflector had much better crystal identification and depth-of-interaction (DOI) resolution. Specifically, the energy, DOI, and timing resolutions for the detector using the BGO array with BaSO4 reflector were 19.8 ± 1.5%, 4.13 ± 0.48 mm, and 2.80 ± 0.23 ns, respectively. In contrast, the values obtained using the BGO array with ESR reflector were 20.9 ± 2.1%, 7.69 ± 1.92 mm, and 2.93 ± 0.20 ns, respectively.