Silvia Mas-Peiro, Alberto Alperi, Marisa Avvedimento, Ander Regueiro, Pablo Avanzas, Domenico Angellotti, Giovanni Esposito, Pablo Vidal-Cales, Siamak Mohammadi, Julio Farjat-Pasos, Cesar Moris, Josep Rodés-Cabau
{"title":"Optimal oversizing in transcatheter aortic valve replacement with the self-expanding Evolut valve system.","authors":"Silvia Mas-Peiro, Alberto Alperi, Marisa Avvedimento, Ander Regueiro, Pablo Avanzas, Domenico Angellotti, Giovanni Esposito, Pablo Vidal-Cales, Siamak Mohammadi, Julio Farjat-Pasos, Cesar Moris, Josep Rodés-Cabau","doi":"10.25270/jic/24.00011","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Valve oversizing has been associated with reduced paravalvular leaks (PVL) and valve migration risk. However, no optimal cut-off oversizing value has been defined for the Evolut system (Medtronic). The aim of this study was to assess the relationship between the degree of oversizing and moderate-to-severe PVL and determine the optimal oversizing cut-off value.</p><p><strong>Methods: </strong>We conducted a multicenter study that included 740 consecutive patients with multidetector computed tomography (CT) data. Valve size was selected according to manufacturer recommendations, with oversizing ranging from 10% to 30%. The primary endpoint was moderate-to-severe PVL.</p><p><strong>Results: </strong>The median age was 84 years (79-87 years), with 58.4% women, and a median EuroSCORE II of 4.1% (2.4-7.3%). Moderate-to-severe PVL was observed in 7.0% of the patients. An inverse relationship was found between oversizing and both PVL (11.3%, 8.6%, 5.4%, and 2.7% for quartiles Q1 to Q4; P = .007) and the need for post-dilation (P = .016). The multivariable analysis showed an association between oversizing and PVL (OR: 0.915 for each 1%-increase, P = .002). The optimal oversizing cut-off value to predict PVL was 20%, and PVL was significantly higher in patients with oversizing less than 20% (10.5% vs.4.2%, P less than .001). There were no differences in major clinical events according to the degree of oversizing, and a higher oversizing did not translate into an increased risk of permanent pacemaker (18.4% vs18.3%, P = .976).</p><p><strong>Conclusions: </strong>In TAVR with the Evolut valve, a higher oversizing was associated with lower rates of moderate-to-severe PVL and a lower need for post-dilation, with no negative impact on procedural and early clinical outcomes. A 20% oversizing threshold could be suggested to reduce PVLs. Further prospective studies are warranted to validate optimal oversizing for this valve system.</p>","PeriodicalId":49261,"journal":{"name":"Journal of Invasive Cardiology","volume":" ","pages":""},"PeriodicalIF":1.6000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Invasive Cardiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.25270/jic/24.00011","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: Valve oversizing has been associated with reduced paravalvular leaks (PVL) and valve migration risk. However, no optimal cut-off oversizing value has been defined for the Evolut system (Medtronic). The aim of this study was to assess the relationship between the degree of oversizing and moderate-to-severe PVL and determine the optimal oversizing cut-off value.
Methods: We conducted a multicenter study that included 740 consecutive patients with multidetector computed tomography (CT) data. Valve size was selected according to manufacturer recommendations, with oversizing ranging from 10% to 30%. The primary endpoint was moderate-to-severe PVL.
Results: The median age was 84 years (79-87 years), with 58.4% women, and a median EuroSCORE II of 4.1% (2.4-7.3%). Moderate-to-severe PVL was observed in 7.0% of the patients. An inverse relationship was found between oversizing and both PVL (11.3%, 8.6%, 5.4%, and 2.7% for quartiles Q1 to Q4; P = .007) and the need for post-dilation (P = .016). The multivariable analysis showed an association between oversizing and PVL (OR: 0.915 for each 1%-increase, P = .002). The optimal oversizing cut-off value to predict PVL was 20%, and PVL was significantly higher in patients with oversizing less than 20% (10.5% vs.4.2%, P less than .001). There were no differences in major clinical events according to the degree of oversizing, and a higher oversizing did not translate into an increased risk of permanent pacemaker (18.4% vs18.3%, P = .976).
Conclusions: In TAVR with the Evolut valve, a higher oversizing was associated with lower rates of moderate-to-severe PVL and a lower need for post-dilation, with no negative impact on procedural and early clinical outcomes. A 20% oversizing threshold could be suggested to reduce PVLs. Further prospective studies are warranted to validate optimal oversizing for this valve system.
期刊介绍:
The Journal of Invasive Cardiology will consider for publication suitable articles on topics pertaining to the invasive treatment of patients with cardiovascular disease.