Francesca Giovannetti, Marta Sacchetti, Marco Marenco, Ludovico Alisi, Giacomo Visioli, Alice Bruscolini, Alessandro Lambiase
{"title":"New disposable esthesiometer (KeraSenseⓇ) to improve diagnosis and management of neurotrophic keratitis","authors":"Francesca Giovannetti, Marta Sacchetti, Marco Marenco, Ludovico Alisi, Giacomo Visioli, Alice Bruscolini, Alessandro Lambiase","doi":"10.1016/j.jtos.2024.03.009","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>To validate the use, repeatability, and reproducibility of a new, cost-effective, disposable, sterile device (KeraSenseⓇ, Dompè farmaceutici SpA, Milan Italy) compared to Cochet-Bonnet (CB) esthesiometer. Secondly, to identify a simple, safe, rapid, and low-cost test to diagnose neurotrophic keratitis (NK).</p></div><div><h3>Methods</h3><p>16 patients with diagnosis of NK stage I, 25 patients with diabetes mellitus (DM), and 26 healthy subjects were included in the study. Corneal sensitivity (CS) was assessed by CB and KeraSenseⓇ. Repeatability, accuracy, and reproducibility of the novel disposable aesthesiometer were assessed. Specificity, sensitivity, and cut-off value for NK diagnosis were calculated by ROC curve analysis.</p></div><div><h3>Results</h3><p>All NK patients showed a CS ≤ 40 mm, while none of the healthy patients showed a CS value < 50 mm. Significant agreement was found between CB measurements and the single use esthesiometer evaluations of CS (p < 0.001). Repeatability evaluations of the single use esthesiometer showed 100% agreement between different measurements (p < 0.001). Reproducibility evaluations showed 99.6% concordance between different operators (p < 0.001). A 55 mm value of the single use esthesiometer was adequate to exclude an NK diagnosis, while all NK patients showed a value ≤ 35 mm.</p></div><div><h3>Conclusions</h3><p>Corneal hypo/anaesthesia is considered the hallmark of NK. The use of the novel single-use esthesiometer will allow for a diagnostic improvement in NK, sparing time and guaranteeing patients’ safety. Diabetic patients despite normal corneal findings may show impairment of CS, suggesting a preclinical stage of NK, requiring a close follow-up.</p></div>","PeriodicalId":54691,"journal":{"name":"Ocular Surface","volume":"32 ","pages":"Pages 192-197"},"PeriodicalIF":5.9000,"publicationDate":"2024-03-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ocular Surface","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1542012424000375","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
To validate the use, repeatability, and reproducibility of a new, cost-effective, disposable, sterile device (KeraSenseⓇ, Dompè farmaceutici SpA, Milan Italy) compared to Cochet-Bonnet (CB) esthesiometer. Secondly, to identify a simple, safe, rapid, and low-cost test to diagnose neurotrophic keratitis (NK).
Methods
16 patients with diagnosis of NK stage I, 25 patients with diabetes mellitus (DM), and 26 healthy subjects were included in the study. Corneal sensitivity (CS) was assessed by CB and KeraSenseⓇ. Repeatability, accuracy, and reproducibility of the novel disposable aesthesiometer were assessed. Specificity, sensitivity, and cut-off value for NK diagnosis were calculated by ROC curve analysis.
Results
All NK patients showed a CS ≤ 40 mm, while none of the healthy patients showed a CS value < 50 mm. Significant agreement was found between CB measurements and the single use esthesiometer evaluations of CS (p < 0.001). Repeatability evaluations of the single use esthesiometer showed 100% agreement between different measurements (p < 0.001). Reproducibility evaluations showed 99.6% concordance between different operators (p < 0.001). A 55 mm value of the single use esthesiometer was adequate to exclude an NK diagnosis, while all NK patients showed a value ≤ 35 mm.
Conclusions
Corneal hypo/anaesthesia is considered the hallmark of NK. The use of the novel single-use esthesiometer will allow for a diagnostic improvement in NK, sparing time and guaranteeing patients’ safety. Diabetic patients despite normal corneal findings may show impairment of CS, suggesting a preclinical stage of NK, requiring a close follow-up.
期刊介绍:
The Ocular Surface, a quarterly, a peer-reviewed journal, is an authoritative resource that integrates and interprets major findings in diverse fields related to the ocular surface, including ophthalmology, optometry, genetics, molecular biology, pharmacology, immunology, infectious disease, and epidemiology. Its critical review articles cover the most current knowledge on medical and surgical management of ocular surface pathology, new understandings of ocular surface physiology, the meaning of recent discoveries on how the ocular surface responds to injury and disease, and updates on drug and device development. The journal also publishes select original research reports and articles describing cutting-edge techniques and technology in the field.