{"title":"Mechanical Effects of Medical Device Attachment to Human Tympanic Membrane.","authors":"Arash Ebrahimian, Hossein Mohammadi, Nima Maftoon","doi":"10.1007/s10162-024-00942-5","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Several treatment methods for hearing disorders rely on attaching medical devices to the tympanic membrane. This study aims to systematically analyze the effects of the material and geometrical properties and location of the medical devices attached to the tympanic membrane on middle-ear vibrations.</p><p><strong>Methods: </strong>A finite-element model of the human middle ear was employed to simulate the effects of attachment of medical devices. Various types of material and geometrical properties, locations, and modeling scenarios were investigated for the medical device.</p><p><strong>Results: </strong>The attachment of the device magnifies the effects of anti-resonances of the middle ear. Additionally, the variations of the material properties of the device significantly alter the middle-ear resonance frequency while changes in the umbo and stapes footplate motions are negligible at frequencies above 5 kHz. Furthermore, modeling the device as a point mass cannot accurately represent the implanted middle-ear behavior. The variations of the diameter and height of the medical device have negligible effects on the middle-ear vibrations at frequencies below 200 Hz but can have considerable impacts at higher frequencies. The effects of changing the device height were negligible at frequencies above 2 kHz. We also discuss the effects of medical device attachment on the vibration patterns of the tympanic membrane as well as the impacts of the variations of the location of the device on the stapes footplate responses.</p><p><strong>Conclusion: </strong>The findings of our study aid the development and optimization of new therapeutic devices, attached to the tympanic membrane, to have the least adverse effects on middle-ear vibrations.</p>","PeriodicalId":56283,"journal":{"name":"Jaro-Journal of the Association for Research in Otolaryngology","volume":" ","pages":"285-302"},"PeriodicalIF":2.4000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11150348/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jaro-Journal of the Association for Research in Otolaryngology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10162-024-00942-5","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/4/1 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"NEUROSCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Several treatment methods for hearing disorders rely on attaching medical devices to the tympanic membrane. This study aims to systematically analyze the effects of the material and geometrical properties and location of the medical devices attached to the tympanic membrane on middle-ear vibrations.
Methods: A finite-element model of the human middle ear was employed to simulate the effects of attachment of medical devices. Various types of material and geometrical properties, locations, and modeling scenarios were investigated for the medical device.
Results: The attachment of the device magnifies the effects of anti-resonances of the middle ear. Additionally, the variations of the material properties of the device significantly alter the middle-ear resonance frequency while changes in the umbo and stapes footplate motions are negligible at frequencies above 5 kHz. Furthermore, modeling the device as a point mass cannot accurately represent the implanted middle-ear behavior. The variations of the diameter and height of the medical device have negligible effects on the middle-ear vibrations at frequencies below 200 Hz but can have considerable impacts at higher frequencies. The effects of changing the device height were negligible at frequencies above 2 kHz. We also discuss the effects of medical device attachment on the vibration patterns of the tympanic membrane as well as the impacts of the variations of the location of the device on the stapes footplate responses.
Conclusion: The findings of our study aid the development and optimization of new therapeutic devices, attached to the tympanic membrane, to have the least adverse effects on middle-ear vibrations.
期刊介绍:
JARO is a peer-reviewed journal that publishes research findings from disciplines related to otolaryngology and communications sciences, including hearing, balance, speech and voice. JARO welcomes submissions describing experimental research that investigates the mechanisms underlying problems of basic and/or clinical significance.
Authors are encouraged to familiarize themselves with the kinds of papers carried by JARO by looking at past issues. Clinical case studies and pharmaceutical screens are not likely to be considered unless they reveal underlying mechanisms. Methods papers are not encouraged unless they include significant new findings as well. Reviews will be published at the discretion of the editorial board; consult the editor-in-chief before submitting.