Verification of analytical performance of alpha-fetoprotein assay on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda
Imad-Eddine El khamlichi, Issam Mokhtari, Nisma Douzi, Oussama Grari, Soufiane Beyyoudh, Dounia El Moujtahide, El houcine Sebbar, Mohammed Choukri
{"title":"Verification of analytical performance of alpha-fetoprotein assay on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda","authors":"Imad-Eddine El khamlichi, Issam Mokhtari, Nisma Douzi, Oussama Grari, Soufiane Beyyoudh, Dounia El Moujtahide, El houcine Sebbar, Mohammed Choukri","doi":"10.30574/gscbps.2024.26.1.0035","DOIUrl":null,"url":null,"abstract":"The aim of our study was the verification of the analytical performance of alpha-fetoprotein determination on the Abbott CI analyzer utilizing the immuno-chemiluminescence method. The verification process was conducted in the biochemistry laboratory of Mohammed VI University Hospital of Oujda. The working methodology adapted is based on the recommendations of the protocol of the French accreditation committee (COFRAC) accreditation technical guide (GTA) 04, by the evaluation of reproducibility and repeatability. The results obtained by this evaluation were overall satisfactory and have meet the recommended criteria set by supplier and the French society of clinical biology. This study shows that the biochemistry laboratory of Mohammed VI University Hospital of Oujda can deliver an accurate and precise results which can be used for clinical diagnosis and decision making.","PeriodicalId":12808,"journal":{"name":"GSC Biological and Pharmaceutical Sciences","volume":"254 ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-01-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"GSC Biological and Pharmaceutical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.30574/gscbps.2024.26.1.0035","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The aim of our study was the verification of the analytical performance of alpha-fetoprotein determination on the Abbott CI analyzer utilizing the immuno-chemiluminescence method. The verification process was conducted in the biochemistry laboratory of Mohammed VI University Hospital of Oujda. The working methodology adapted is based on the recommendations of the protocol of the French accreditation committee (COFRAC) accreditation technical guide (GTA) 04, by the evaluation of reproducibility and repeatability. The results obtained by this evaluation were overall satisfactory and have meet the recommended criteria set by supplier and the French society of clinical biology. This study shows that the biochemistry laboratory of Mohammed VI University Hospital of Oujda can deliver an accurate and precise results which can be used for clinical diagnosis and decision making.
我们的研究旨在验证雅培 CI 分析仪利用免疫化学发光法测定甲胎蛋白的分析性能。验证过程在乌季达穆罕默德六世大学医院的生化实验室进行。所采用的工作方法是根据法国认证委员会(COFRAC)认证技术指南(GTA)04 协议的建议,对再现性和重复性进行评估。评估结果总体上令人满意,符合供应商和法国临床生物学会推荐的标准。这项研究表明,乌季达穆罕默德六世大学医院的生化实验室能够提供准确和精确的结果,可用于临床诊断和决策。