Dose optimization in newborn abdominal radiography: Assessing the added value of additional filtration on radiation dose and image quality using an anthropomorphic phantom
Annie-Lyne Petit , Rabih Alwan , Julien Behr , Paul Calame , Marion Lenoir , Hubert Ducou le Pointe , Éric Delabrousse
{"title":"Dose optimization in newborn abdominal radiography: Assessing the added value of additional filtration on radiation dose and image quality using an anthropomorphic phantom","authors":"Annie-Lyne Petit , Rabih Alwan , Julien Behr , Paul Calame , Marion Lenoir , Hubert Ducou le Pointe , Éric Delabrousse","doi":"10.1016/j.redii.2024.100045","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Abdominal radiographs remain useful in newborns. Given the high radiation sensitivity of this population, it is necessary to optimize acquisition techniques to minimize radiation exposure.</p></div><div><h3>Objective</h3><p>Evaluate the effects of three additional filtrations on radiation dose and image quality in abdominal X-rays of newborns using an anthropomorphic phantom.</p></div><div><h3>Material and method</h3><p>Abdominal radiographs of an anthropomorphic newborn phantom were performed using acquisition parameters ranging from 55 to 70 kV and from 0.4 to 2.5 mAs, without and with three different additional filtrations: 0.1 mm copper (Cu) + 1 mm aluminum (Al), 0.2 mm copper + 1 mm aluminum, and 2 mm aluminum. For each X-ray the dose area product (DAP) was measured, the signal-to-noise ratio (SNR) was calculated, and image quality (IQ) was evaluated by two blinded radiologists using the absolute visual grading analysis (VGA) method.</p></div><div><h3>Results</h3><p>Adding an additional filtration resulted in a significant reduction in DAP, with a decrease of 42% using 2 mm Al filtration, 65% with 0.1 mm Cu + 1 mm Al filtration, and 78% with 0.2 mm Cu + 1 mm Al filtration (<em>p</em> < 0.01). The addition of 2 mm aluminum filtration does not significantly decrease the SNR (<em>p</em> = 0.31), CNR (<em>p</em> = 0.52) or the IQ (<em>p</em> = 0.12 and 0.401 for reader 1 and 2, respectively). However, adding copper-containing filtration leads to a significant decrease in, SNR, CNR and IQ.</p></div><div><h3>Conclusion</h3><p>Adding a 2 mm Al additional filtration for abdominal radiographs in newborns can significantly reduce the radiation dose without causing a significant decrease in image quality.</p></div>","PeriodicalId":74676,"journal":{"name":"Research in diagnostic and interventional imaging","volume":"10 ","pages":"Article 100045"},"PeriodicalIF":0.0000,"publicationDate":"2024-04-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2772652524000061/pdfft?md5=74dae0f6a82cde59ddb22e12c46456d8&pid=1-s2.0-S2772652524000061-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Research in diagnostic and interventional imaging","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2772652524000061","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Abdominal radiographs remain useful in newborns. Given the high radiation sensitivity of this population, it is necessary to optimize acquisition techniques to minimize radiation exposure.
Objective
Evaluate the effects of three additional filtrations on radiation dose and image quality in abdominal X-rays of newborns using an anthropomorphic phantom.
Material and method
Abdominal radiographs of an anthropomorphic newborn phantom were performed using acquisition parameters ranging from 55 to 70 kV and from 0.4 to 2.5 mAs, without and with three different additional filtrations: 0.1 mm copper (Cu) + 1 mm aluminum (Al), 0.2 mm copper + 1 mm aluminum, and 2 mm aluminum. For each X-ray the dose area product (DAP) was measured, the signal-to-noise ratio (SNR) was calculated, and image quality (IQ) was evaluated by two blinded radiologists using the absolute visual grading analysis (VGA) method.
Results
Adding an additional filtration resulted in a significant reduction in DAP, with a decrease of 42% using 2 mm Al filtration, 65% with 0.1 mm Cu + 1 mm Al filtration, and 78% with 0.2 mm Cu + 1 mm Al filtration (p < 0.01). The addition of 2 mm aluminum filtration does not significantly decrease the SNR (p = 0.31), CNR (p = 0.52) or the IQ (p = 0.12 and 0.401 for reader 1 and 2, respectively). However, adding copper-containing filtration leads to a significant decrease in, SNR, CNR and IQ.
Conclusion
Adding a 2 mm Al additional filtration for abdominal radiographs in newborns can significantly reduce the radiation dose without causing a significant decrease in image quality.