Effects of sterilization, conditioning, and thermal aging on the retention of zirconia hybrid abutments: A laboratory study

IF 3.7 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE Clinical Implant Dentistry and Related Research Pub Date : 2024-04-29 DOI:10.1111/cid.13332
Nalah Alghanaim, Matthias Kern, Christine Yazigi, M. Sad Chaar
{"title":"Effects of sterilization, conditioning, and thermal aging on the retention of zirconia hybrid abutments: A laboratory study","authors":"Nalah Alghanaim,&nbsp;Matthias Kern,&nbsp;Christine Yazigi,&nbsp;M. Sad Chaar","doi":"10.1111/cid.13332","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objective</h3>\n \n <p>To investigate the effects of sterilization, conditioning method, and thermal aging on the retentive strength of two-piece zirconia abutments.</p>\n </section>\n \n <section>\n \n <h3> Materials and Methods</h3>\n \n <p>In total, 128 stock zirconia abutments were divided into four groups (<i>n</i> = 32) according to the conditioning parameters: (A) air-abrasion using 50 μm alumina particles/1.0 bar, (B) 50 μm/2.0 bar, (C) 100 μm/1.0 bar, and (D) 100 μm/2.0 bar. All abutments were bonded onto titanium bases using DTK adhesive resin and stored in water bath (37°C) for 72 h. Each group was subdivided into two subgroups (<i>n</i> = 16), group 1 was disinfected, whereas group 2 followed disinfection and autoclave sterilization. Half of the specimens of each subgroup (<i>n</i>&amp;amp;#x02009;=&amp;amp;#x02009;8) was directly subjected to the axial retention test (groups N), while the other half was first subjected to 150&amp;amp;#x02009;days of thermocycling followed by retention test (groups T). Statistical analysis was performed with three-way ANOVA, additional statistical analysis was performed by using separate one-way ANOVAs followed by the Tukey's post-hoc test for post hoc pairwise comparisons among groups.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The highest median retention strength was recorded for group B2N (1390 N), whereas the lowest strength was recorded for group C1T (688 N). No significant interaction (<i>p</i> ≥ 0.05) was detected between the different variables; conditioning method, sterilization, and the thermal cycling regarding the effect on the resulting retention. However, the sterilization always showed a positive effect. Thermocycling presented an adversely significant effect only in the absence of sterilization (<i>p</i> &lt; 0.05), with the exception of subgroups A. For the sterilized groups, thermocycling had no statistically significant effect on the retention.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>Steam autoclaving increased the retention of hybrid zirconia abutments. DTK adhesive resin for two-piece zirconia abutments performed well after sterilization and thermocycling.</p>\n </section>\n </div>","PeriodicalId":50679,"journal":{"name":"Clinical Implant Dentistry and Related Research","volume":"26 4","pages":"679-687"},"PeriodicalIF":3.7000,"publicationDate":"2024-04-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/cid.13332","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Implant Dentistry and Related Research","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/cid.13332","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

To investigate the effects of sterilization, conditioning method, and thermal aging on the retentive strength of two-piece zirconia abutments.

Materials and Methods

In total, 128 stock zirconia abutments were divided into four groups (n = 32) according to the conditioning parameters: (A) air-abrasion using 50 μm alumina particles/1.0 bar, (B) 50 μm/2.0 bar, (C) 100 μm/1.0 bar, and (D) 100 μm/2.0 bar. All abutments were bonded onto titanium bases using DTK adhesive resin and stored in water bath (37°C) for 72 h. Each group was subdivided into two subgroups (n = 16), group 1 was disinfected, whereas group 2 followed disinfection and autoclave sterilization. Half of the specimens of each subgroup (n&amp;#x02009;=&amp;#x02009;8) was directly subjected to the axial retention test (groups N), while the other half was first subjected to 150&amp;#x02009;days of thermocycling followed by retention test (groups T). Statistical analysis was performed with three-way ANOVA, additional statistical analysis was performed by using separate one-way ANOVAs followed by the Tukey's post-hoc test for post hoc pairwise comparisons among groups.

Results

The highest median retention strength was recorded for group B2N (1390 N), whereas the lowest strength was recorded for group C1T (688 N). No significant interaction (p ≥ 0.05) was detected between the different variables; conditioning method, sterilization, and the thermal cycling regarding the effect on the resulting retention. However, the sterilization always showed a positive effect. Thermocycling presented an adversely significant effect only in the absence of sterilization (p < 0.05), with the exception of subgroups A. For the sterilized groups, thermocycling had no statistically significant effect on the retention.

Conclusion

Steam autoclaving increased the retention of hybrid zirconia abutments. DTK adhesive resin for two-piece zirconia abutments performed well after sterilization and thermocycling.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
消毒、调理和热老化对氧化锆混合基台固位的影响:实验室研究
材料和方法将 128 个氧化锆基台按照调理参数分为四组(n = 32):(A) 使用 50 μm 氧化铝颗粒/1.0 bar 空磨,(B) 50 μm/2.0 bar 空磨,(C) 100 μm/1.0 bar 空磨,(D) 100 μm/2.0 bar 空磨。所有基台都使用 DTK 粘接树脂粘接在钛基台上,并在水浴(37°C)中保存 72 小时。每组又分为两个亚组(n = 16),第 1 组进行消毒,第 2 组进行消毒和高压灭菌。每个子组的一半样本(n&amp;#x02009;=&amp;#x02009;8)直接进行轴向保留试验(N 组),另一半样本先进行 150&amp;#x02009;d 的热循环,然后进行保留试验(T 组)。统计分析采用三因素方差分析,其他统计分析采用单独的单因素方差分析,然后采用 Tukey's 事后检验进行组间配对比较。结果 B2N 组的中位保留强度最高(1390 N),而 C1T 组的中位保留强度最低(688 N)。在不同变量(调节方法、灭菌和热循环)对所产生的保持力的影响方面,没有发现明显的交互作用(p ≥ 0.05)。然而,灭菌总是显示出积极的影响。结论蒸汽高压灭菌提高了混合氧化锆基台的固位力。用于两件式氧化锆基台的 DTK 粘接树脂在灭菌和热循环后表现良好。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
6.00
自引率
13.90%
发文量
103
审稿时长
4-8 weeks
期刊介绍: The goal of Clinical Implant Dentistry and Related Research is to advance the scientific and technical aspects relating to dental implants and related scientific subjects. Dissemination of new and evolving information related to dental implants and the related science is the primary goal of our journal. The range of topics covered by the journals will include but be not limited to: New scientific developments relating to bone Implant surfaces and their relationship to the surrounding tissues Computer aided implant designs Computer aided prosthetic designs Immediate implant loading Immediate implant placement Materials relating to bone induction and conduction New surgical methods relating to implant placement New materials and methods relating to implant restorations Methods for determining implant stability A primary focus of the journal is publication of evidenced based articles evaluating to new dental implants, techniques and multicenter studies evaluating these treatments. In addition basic science research relating to wound healing and osseointegration will be an important focus for the journal.
期刊最新文献
Issue Information Featured Cover A transcrestal sinus floor elevation strategy based on a haptic robot system: An in vitro study Influence of repeated implant‐abutment manipulation on the prevalence of peri‐implant diseases in complete arch restorations. A retrospective analysis Biocompatibility and dimensional stability through the use of 3D‐printed scaffolds made by polycaprolactone and bioglass‐7: An in vitro and in vivo study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1