{"title":"Comparison of four D-dimer assays in the context of venous thromboembolism in the emergency department","authors":"Fabio Del Ben, Elisabetta Fontanini, Gabriella Azzarini, Marina Arini, Christian Marini, Giovanni Poli, Paola Pradella, Serena Parusso, Liliana Santarossa, Federica Targa, Lorena Zardo, Roberta Giacomello, Benedetto Morelli","doi":"10.1111/ijlh.14302","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>This observational study conducted across seven emergency care units compares the efficacy of four D-dimer detection methods, namely HemosIL D-dimer HS (HS), HemosIL D-dimer HS-500 (HS-500), VIDAS D-dimer (VIDAS), and HemosIL AcuStar D-dimer (ACUSTAR). The primary focus is on patients with a clinical suspicion of deep venous thrombosis (DVT) or pulmonary embolism (PE).</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A total of 149 samples were collected from patients with suspected DVT or PE. The confirmation of DVT/PE was based on calf ultrasound or computed tomography-Angiography. Direct comparisons were made between the different detection methods, considering both their analytical performance and clinical utility. Additionally, the impact of an age-adjusted cut-off on the diagnostic accuracy of each method was assessed.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The results revealed comparable negative predictive value, sensitivity, and specificity across the methods, with a notable exception of increased specificity for HS compared with HS-500 (50.8% vs. 41.5%, <i>p</i> = 0.03). Further analysis incorporating an age-adjusted cut-off demonstrated a significant improvement in specificity for HS. When using the age-adjusted cut-off, HS exhibited a substantial increase in specificity compared with HS-500 (63.1% vs. 49.2%, <i>p</i> = 0.004) and demonstrated significantly higher specificity compared with VIDAS (63.1% vs. 53.8%, <i>p</i> = 0.04).</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The study emphasizes the nonuniversal effect of an age-adjusted cut-off and discusses the potential necessity for different cut-off values, particularly in the case of HS-500. These findings contribute to the understanding of D-dimer detection methods in the context of DVT and PE, providing insights into their relative performances and the potential optimization through age-adjusted cut-offs.</p>\n </section>\n </div>","PeriodicalId":14120,"journal":{"name":"International Journal of Laboratory Hematology","volume":"46 5","pages":"936-945"},"PeriodicalIF":2.2000,"publicationDate":"2024-05-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Laboratory Hematology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/ijlh.14302","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
This observational study conducted across seven emergency care units compares the efficacy of four D-dimer detection methods, namely HemosIL D-dimer HS (HS), HemosIL D-dimer HS-500 (HS-500), VIDAS D-dimer (VIDAS), and HemosIL AcuStar D-dimer (ACUSTAR). The primary focus is on patients with a clinical suspicion of deep venous thrombosis (DVT) or pulmonary embolism (PE).
Methods
A total of 149 samples were collected from patients with suspected DVT or PE. The confirmation of DVT/PE was based on calf ultrasound or computed tomography-Angiography. Direct comparisons were made between the different detection methods, considering both their analytical performance and clinical utility. Additionally, the impact of an age-adjusted cut-off on the diagnostic accuracy of each method was assessed.
Results
The results revealed comparable negative predictive value, sensitivity, and specificity across the methods, with a notable exception of increased specificity for HS compared with HS-500 (50.8% vs. 41.5%, p = 0.03). Further analysis incorporating an age-adjusted cut-off demonstrated a significant improvement in specificity for HS. When using the age-adjusted cut-off, HS exhibited a substantial increase in specificity compared with HS-500 (63.1% vs. 49.2%, p = 0.004) and demonstrated significantly higher specificity compared with VIDAS (63.1% vs. 53.8%, p = 0.04).
Conclusion
The study emphasizes the nonuniversal effect of an age-adjusted cut-off and discusses the potential necessity for different cut-off values, particularly in the case of HS-500. These findings contribute to the understanding of D-dimer detection methods in the context of DVT and PE, providing insights into their relative performances and the potential optimization through age-adjusted cut-offs.
期刊介绍:
The International Journal of Laboratory Hematology provides a forum for the communication of new developments, research topics and the practice of laboratory haematology.
The journal publishes invited reviews, full length original articles, and correspondence.
The International Journal of Laboratory Hematology is the official journal of the International Society for Laboratory Hematology, which addresses the following sub-disciplines: cellular analysis, flow cytometry, haemostasis and thrombosis, molecular diagnostics, haematology informatics, haemoglobinopathies, point of care testing, standards and guidelines.
The journal was launched in 2006 as the successor to Clinical and Laboratory Hematology, which was first published in 1979. An active and positive editorial policy ensures that work of a high scientific standard is reported, in order to bridge the gap between practical and academic aspects of laboratory haematology.