Evaluation of conditions to minimize mercury losses from leaching test eluates

IF 4.1 Q1 CHEMISTRY, ANALYTICAL Talanta Open Pub Date : 2024-05-13 DOI:10.1016/j.talo.2024.100326
L. Kelley, R.C. Delapp, A.C. Garrabrants, D.S. Kosson
{"title":"Evaluation of conditions to minimize mercury losses from leaching test eluates","authors":"L. Kelley,&nbsp;R.C. Delapp,&nbsp;A.C. Garrabrants,&nbsp;D.S. Kosson","doi":"10.1016/j.talo.2024.100326","DOIUrl":null,"url":null,"abstract":"<div><p>A series of experiments are described that provide recommendations of modifications to leaching test specifications to minimize losses of mercury from eluates during leaching tests and management of analytical samples. EPA methods of the Leaching Environmental Assessment Framework (LEAF) were used as a reference point for what steps are most likely to see losses, including filtration, storage of unpreserved mercury contaminated material, and materials of construction for vessels. The potential sorption of mercury to borosilicate glass and PTFE lab materials was tested to determine what materials are suitable to replace HDPE and polypropylene. The experimental findings indicate that bottles constructed of PTFE and Type I (borosilicate) glass are recommended as extraction and containment vessels for mercury-containing liquids. Type I syringe filtration of aqueous mercury samples at pH 2 through 9 through 0.45 µm PTFE filters housed in polypropylene is acceptable; however, sorption losses were observed to occur at pH &gt; 9. Storage of unpreserved liquids containing mercury over extended leaching test intervals did not lead to significant losses of mercury when coupled with minimized headspace. All mercury-containing eluates should include preservation with Optima HNO<sub>3</sub> to a 1 % concentration and gold (III) standard to a concentration of 200 µg/L, followed by refrigeration at &lt;6 °C prior to analysis.</p></div>","PeriodicalId":436,"journal":{"name":"Talanta Open","volume":"9 ","pages":"Article 100326"},"PeriodicalIF":4.1000,"publicationDate":"2024-05-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2666831924000407/pdfft?md5=9a0820aa87035bef9f0575f7a8ca9acb&pid=1-s2.0-S2666831924000407-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Talanta Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666831924000407","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

A series of experiments are described that provide recommendations of modifications to leaching test specifications to minimize losses of mercury from eluates during leaching tests and management of analytical samples. EPA methods of the Leaching Environmental Assessment Framework (LEAF) were used as a reference point for what steps are most likely to see losses, including filtration, storage of unpreserved mercury contaminated material, and materials of construction for vessels. The potential sorption of mercury to borosilicate glass and PTFE lab materials was tested to determine what materials are suitable to replace HDPE and polypropylene. The experimental findings indicate that bottles constructed of PTFE and Type I (borosilicate) glass are recommended as extraction and containment vessels for mercury-containing liquids. Type I syringe filtration of aqueous mercury samples at pH 2 through 9 through 0.45 µm PTFE filters housed in polypropylene is acceptable; however, sorption losses were observed to occur at pH > 9. Storage of unpreserved liquids containing mercury over extended leaching test intervals did not lead to significant losses of mercury when coupled with minimized headspace. All mercury-containing eluates should include preservation with Optima HNO3 to a 1 % concentration and gold (III) standard to a concentration of 200 µg/L, followed by refrigeration at <6 °C prior to analysis.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估最大限度减少浸出试验洗脱液中汞损失的条件
文中描述了一系列实验,提出了修改浸出测试规格的建议,以尽量减少浸出测试和分析样本管理期间洗脱液中汞的损失。浸出环境评估框架(LEAF)中的 EPA 方法被用作最有可能出现损失的步骤的参考点,包括过滤、未保存的汞污染材料的储存以及容器的结构材料。测试了汞对硼硅玻璃和聚四氟乙烯实验室材料的潜在吸附性,以确定哪些材料适合替代高密度聚乙烯和聚丙烯。实验结果表明,建议使用聚四氟乙烯和 I 型(硼硅酸盐)玻璃瓶作为含汞液体的提取和盛放容器。在 pH 值为 2 到 9 的情况下,用 I 型注射器通过 0.45 微米的聚四氟乙烯过滤器(装在聚丙烯中)过滤汞水样是可以接受的;但在 pH 值为 9 时,会出现吸附损失。所有含汞洗脱液都应该用 Optima HNO3 保存到 1 % 的浓度,用金 (III) 标准保存到 200 µg/L 的浓度,然后在分析前在 <6 °C 下冷藏。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Talanta Open
Talanta Open Chemistry-Analytical Chemistry
CiteScore
5.20
自引率
0.00%
发文量
86
审稿时长
49 days
期刊最新文献
MCM-41 based dispersive micro-solid phase extraction of selected cephalosporin antibiotic residues from water samples prior to liquid chromatographic quantification Bone fragility in Type 2 Diabetes Mellitus. Influence of sex and cardiovascular disease in a pilot study using metabolomics Molecularly imprinted biomimetic plasmonic sensor decorated with gold nanoparticles for selective and sensitive detection of bisphenol A Post-burn and post-blast rapid detection of trace and bulk energetics by 3D-printed cone spray ionization mass spectrometry Continuous-flow analysis of nitrogen compounds in environmental water using a copper–zinc reduction coil
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1