{"title":"Analytical performance evaluation of hemoglobin A1c on an ARKRAY HA-8160 analyzer with newly-developed mobile phase buffer","authors":"Yuan Yu , Xiaoyun Zhang , Kai Lin","doi":"10.1016/j.plabm.2024.e00414","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Most glycated hemoglobin A1c (HbA1c) analytical reagents used were obtained from the analyzer's manufacturer. However, clinical laboratories need more choices for HbA1c analytical reagents to overcome the limitations of dedicated reagents for special analyzers. We developed new mobile phase buffers as HbA1c diagnostic reagents and evaluated their analytical performance for the HbA1c assay.</p></div><div><h3>Methods</h3><p>Different mobile phase buffers used as HbA1c diagnostic reagents were prepared using different concentrations of sodium salts. According to the Clinical and Laboratory Standards Institute (CLSI) recommendation guidelines, the analytical performances of the newly developed mobile phase buffers were evaluated on an ARKRAY HA-8160 Analyzer. Both quality controls and clinical blood samples were used in these experiments. To assess the quality of the newly developed mobile phase buffers, precision, accuracy, linearity, carryover, interference, bias, correlation with commercial reagents, and stability were analyzed.</p></div><div><h3>Results</h3><p>The <em>CV</em>s of intra-assay precision and interassay precision of quality control and clinical.</p><p>There were fewer than 1.00 % blood sample assays using the newly developed mobile phase buffer. The <em>RDs</em> of accuracy were less than 1.00 %. Linearity: R<sup>2</sup> = 0.9998 in the concentration range of 4.40%–17.30 %. Carryover: 0.00 %. Reagent comparison revealed that the Pearson regression equation was Y = 0.9884x+0.05692 (R<sup>2</sup> = 0.9977), and the Bland-Altman mean difference was −0.02650 % (CI: −0.2121 %–0.1591 %) between the two analytical reagents. Stability was also acceptable within 12 months. This mobile phase buffer showed good anti-interference ability.</p></div><div><h3>Conclusion</h3><p>The newly developed mobile phase buffers demonstrated good analytical performance and were suitable for clinical HbA1c assays on an ARKRAY HA-8160 Analyzer.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S235255172400060X/pdfft?md5=f33377b107c40be9e47def3bbf412eb8&pid=1-s2.0-S235255172400060X-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S235255172400060X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Most glycated hemoglobin A1c (HbA1c) analytical reagents used were obtained from the analyzer's manufacturer. However, clinical laboratories need more choices for HbA1c analytical reagents to overcome the limitations of dedicated reagents for special analyzers. We developed new mobile phase buffers as HbA1c diagnostic reagents and evaluated their analytical performance for the HbA1c assay.
Methods
Different mobile phase buffers used as HbA1c diagnostic reagents were prepared using different concentrations of sodium salts. According to the Clinical and Laboratory Standards Institute (CLSI) recommendation guidelines, the analytical performances of the newly developed mobile phase buffers were evaluated on an ARKRAY HA-8160 Analyzer. Both quality controls and clinical blood samples were used in these experiments. To assess the quality of the newly developed mobile phase buffers, precision, accuracy, linearity, carryover, interference, bias, correlation with commercial reagents, and stability were analyzed.
Results
The CVs of intra-assay precision and interassay precision of quality control and clinical.
There were fewer than 1.00 % blood sample assays using the newly developed mobile phase buffer. The RDs of accuracy were less than 1.00 %. Linearity: R2 = 0.9998 in the concentration range of 4.40%–17.30 %. Carryover: 0.00 %. Reagent comparison revealed that the Pearson regression equation was Y = 0.9884x+0.05692 (R2 = 0.9977), and the Bland-Altman mean difference was −0.02650 % (CI: −0.2121 %–0.1591 %) between the two analytical reagents. Stability was also acceptable within 12 months. This mobile phase buffer showed good anti-interference ability.
Conclusion
The newly developed mobile phase buffers demonstrated good analytical performance and were suitable for clinical HbA1c assays on an ARKRAY HA-8160 Analyzer.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.