Repeatability of Noninvasive Break-Up Time Measurements using Keratograph Oculus 3.

Q4 Medicine Ceska a Slovenska Oftalmologie Pub Date : 2024-01-01 DOI:10.31348/2024/26
Lenka Musilová, Alena Vodičková, František Pluháček
{"title":"Repeatability of Noninvasive Break-Up Time Measurements using Keratograph Oculus 3.","authors":"Lenka Musilová, Alena Vodičková, František Pluháček","doi":"10.31348/2024/26","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>The primary aim of this study is to evaluate the repeatability of noninvasive break-up time (NIBUT) measurement by keratograph when it is determined from one, two or three partial measurements, and to recommend a suitable methodology for practice. Another goal is to verify that repeated measurements do not affect the measured value.</p><p><strong>Material and methods: </strong>Thirty-eight healthy volunteers (30 women and 8 men) aged between 19 and 50 years old were included in the study, in which only one eye of each volunteer was measured. The study was designed as a prospective one. Each subject adapted to the local conditions of the laboratory for 15 minutes and subsequently underwent two series of NIBUT measurements (test, retest) on an OCULUS 3 Keratograph. The minimum time interval between the two series was 10 minutes, in which each series contained three partial measurements approximately 3 three measurements in the given series. Repeatability was assessed by a Bland-Altman analysis and expressed as a repeatability coefficient. In every case, only the time of the first break-up of the tear film was monitored.</p><p><strong>Results: </strong>The statistical analysis did not show statistically significant differences both between partial measurements of NIBUT in the individual series (p = 0.92, p = 0.81) and when comparing all six measurements (p = 0.95). The mean values of the partial measurements ranged from 13.6 s to 14.4 s. The repeatability coefficients were found to be 15.0 s, 12.1 s and 10.0 s for methodologies A, B and C, respectively. A supplementary analysis for 12 eyes with low NIBUT (&lt; 10 s) showed statistically significantly better repeatability in this group, with coefficients of 7.0 s (methodology A), 6.0 s (B) and 4.6 s (C).</p><p><strong>Conclusion: </strong>Determination of NIBUT from three consecutive measurements (with a sufficient interval of ideally a few minutes) significantly improves repeatability. Such repeated NIBUT measurements do not have a significant effect on the measured value. The mentioned methodology for measuring NIBUT on a keratograph can be recommended for use in practice.</p>","PeriodicalId":39839,"journal":{"name":"Ceska a Slovenska Oftalmologie","volume":"80 Ahead of print","pages":"266-270"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ceska a Slovenska Oftalmologie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31348/2024/26","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Aim: The primary aim of this study is to evaluate the repeatability of noninvasive break-up time (NIBUT) measurement by keratograph when it is determined from one, two or three partial measurements, and to recommend a suitable methodology for practice. Another goal is to verify that repeated measurements do not affect the measured value.

Material and methods: Thirty-eight healthy volunteers (30 women and 8 men) aged between 19 and 50 years old were included in the study, in which only one eye of each volunteer was measured. The study was designed as a prospective one. Each subject adapted to the local conditions of the laboratory for 15 minutes and subsequently underwent two series of NIBUT measurements (test, retest) on an OCULUS 3 Keratograph. The minimum time interval between the two series was 10 minutes, in which each series contained three partial measurements approximately 3 three measurements in the given series. Repeatability was assessed by a Bland-Altman analysis and expressed as a repeatability coefficient. In every case, only the time of the first break-up of the tear film was monitored.

Results: The statistical analysis did not show statistically significant differences both between partial measurements of NIBUT in the individual series (p = 0.92, p = 0.81) and when comparing all six measurements (p = 0.95). The mean values of the partial measurements ranged from 13.6 s to 14.4 s. The repeatability coefficients were found to be 15.0 s, 12.1 s and 10.0 s for methodologies A, B and C, respectively. A supplementary analysis for 12 eyes with low NIBUT (< 10 s) showed statistically significantly better repeatability in this group, with coefficients of 7.0 s (methodology A), 6.0 s (B) and 4.6 s (C).

Conclusion: Determination of NIBUT from three consecutive measurements (with a sufficient interval of ideally a few minutes) significantly improves repeatability. Such repeated NIBUT measurements do not have a significant effect on the measured value. The mentioned methodology for measuring NIBUT on a keratograph can be recommended for use in practice.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
使用 Keratograph Oculus 3 进行无创破裂时间测量的重复性。
目的:本研究的主要目的是评估角膜塑形镜测量无创破裂时间(NIBUT)的重复性,即通过一次、两次或三次局部测量确定的重复性,并为实践推荐合适的方法。另一个目标是验证重复测量不会影响测量值:研究对象包括 38 名健康志愿者(30 名女性和 8 名男性),年龄在 19 岁至 50 岁之间,其中只测量了每名志愿者的一只眼睛。该研究为前瞻性研究。每位受试者在实验室适应当地条件 15 分钟,然后在 OCULUS 3 角膜塑形镜上进行两组 NIBUT 测量(测试、复测)。两个系列之间的最小时间间隔为 10 分钟,其中每个系列包含三次局部测量,每次测量约为给定系列中的三次测量。重复性通过布兰-阿尔特曼分析进行评估,并以重复性系数表示。在每种情况下,只监测泪膜第一次破裂的时间:统计分析结果表明,在单个系列的 NIBUT 部分测量值之间(p = 0.92,p = 0.81)以及在比较所有六次测量值时(p = 0.95),均无显著统计学差异。方法 A、B 和 C 的重复性系数分别为 15.0 秒、12.1 秒和 10.0 秒。对 12 只 NIBUT 值低(< 10 秒)的眼睛进行的补充分析表明,这组眼睛的重复性在统计学上明显更好,系数分别为 7.0 秒(方法 A)、6.0 秒(B)和 4.6 秒(C):结论:通过三次连续测量(间隔时间最好为几分钟)确定 NIBUT 可显著提高重复性。这种重复的 NIBUT 测量对测量值没有明显影响。上述在角膜塑形镜上测量 NIBUT 的方法可推荐在实践中使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Ceska a Slovenska Oftalmologie
Ceska a Slovenska Oftalmologie Medicine-Ophthalmology
CiteScore
0.80
自引率
0.00%
发文量
0
期刊介绍: - Redakce přijímá pouze práce vyhovující po odborné stránce, které jsou na odpovídající profesionální a formální úrovni. - Uveřejněná práce se stává majetkem časopisu, přetisknout její část nebo obrázek lze jen s citací původu. - Rukopis zasílejte v originále a dobře čitelné kopii (je nutná také kopie tabulek, legend, podle možnosti i obrázků). - Listy číslujte v pravém horním rohu a spojujte svorkou, nesešívejte. Přijímáme práce psané na jedné straně kvalitního bílého nelesklého papíru formátu A4 (neprůklepový) na psacím stroji nebo počítači s obvyklými typy.
期刊最新文献
ULTRASOUND EXAMINATION OF THE ORBIT IN PATIENTS WITH THYROIDASSOCIATED ORBITOPATHY - EXAMINATION GUIDE AND RECOMMENDATIONS FOR EVERYDAY PRACTICE. A REVIEW. Diagnostic Importance of OCT Pachymetry in Keratoconus. Improvement of Visual Field Defects after Neuroembolization Treatment of Intracranial Aneurysms. Case Reports. Pars Plana Vitrectomy in the Treatment of Rhegmatogenous Retinal Detachment. Refractive Errors Among Members of the Armed Forces of the Czech Republic.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1