Real-Time Ultrasound Tip Location Reduces Malposition and Radiation Exposure during Umbilical Venous Catheter Placement in Neonates: A Retrospective, Observational Study.
Vito D'Andrea, Giorgia Prontera, Francesco Cota, Alessandro Perri, Rosellina Russo, Giovanni Barone, Giovanni Vento
{"title":"Real-Time Ultrasound Tip Location Reduces Malposition and Radiation Exposure during Umbilical Venous Catheter Placement in Neonates: A Retrospective, Observational Study.","authors":"Vito D'Andrea, Giorgia Prontera, Francesco Cota, Alessandro Perri, Rosellina Russo, Giovanni Barone, Giovanni Vento","doi":"10.1159/000538905","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The umbilical venous catheter is a vital access device in neonatal intensive care units for preterm and critically ill infants. Correct positioning is crucial, as malpositioning can lead to severe complications. According to international guidelines, the position of the umbilical venous catheter tip must be assessed in real time; traditionally, the catheter is visualized with a thoracoabdominal X-ray, but one of the most effective and safest methods is therefore real-time ultrasound.</p><p><strong>Methods: </strong>This study compares real-time ultrasound and traditional X-ray methods for assessing umbilical venous catheter tip location in 461 cases. The rate of tip malposition was analyzed retrospectively. The secondary aim was to assess indwelling time of umbilical venous catheters and reasons of removal.</p><p><strong>Results: </strong>Real-time ultrasound tip location, found to be more reliable and efficient, demonstrated a significantly lower incidence of primary malpositioning compared to X-ray assessments (9.6 vs. 75.9%). The study also highlighted the association of real-time ultrasound with reduced catheter manipulation, fewer radiographs, and higher indwelling times of umbilical venous catheter. The multiple logistic regression showed a high probability of the central safe position of the umbilical venous catheter tip using real-time ultrasound tip location (odds ratio 29.5, 95% confidence interval: 17.4-49.4).</p><p><strong>Conclusion: </strong>The findings support the adoption of real-time ultrasound in clinical settings to enhance umbilical venous catheter placement accuracy and minimize associated risks. A minimal training investment is needed to attain the proficiency to visualize the umbilical venous catheters, offering a substantial advantage in terms of both cost-effectiveness for the procedure and enhanced patient safety.</p>","PeriodicalId":94152,"journal":{"name":"Neonatology","volume":" ","pages":"1-6"},"PeriodicalIF":0.0000,"publicationDate":"2024-06-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Neonatology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1159/000538905","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: The umbilical venous catheter is a vital access device in neonatal intensive care units for preterm and critically ill infants. Correct positioning is crucial, as malpositioning can lead to severe complications. According to international guidelines, the position of the umbilical venous catheter tip must be assessed in real time; traditionally, the catheter is visualized with a thoracoabdominal X-ray, but one of the most effective and safest methods is therefore real-time ultrasound.
Methods: This study compares real-time ultrasound and traditional X-ray methods for assessing umbilical venous catheter tip location in 461 cases. The rate of tip malposition was analyzed retrospectively. The secondary aim was to assess indwelling time of umbilical venous catheters and reasons of removal.
Results: Real-time ultrasound tip location, found to be more reliable and efficient, demonstrated a significantly lower incidence of primary malpositioning compared to X-ray assessments (9.6 vs. 75.9%). The study also highlighted the association of real-time ultrasound with reduced catheter manipulation, fewer radiographs, and higher indwelling times of umbilical venous catheter. The multiple logistic regression showed a high probability of the central safe position of the umbilical venous catheter tip using real-time ultrasound tip location (odds ratio 29.5, 95% confidence interval: 17.4-49.4).
Conclusion: The findings support the adoption of real-time ultrasound in clinical settings to enhance umbilical venous catheter placement accuracy and minimize associated risks. A minimal training investment is needed to attain the proficiency to visualize the umbilical venous catheters, offering a substantial advantage in terms of both cost-effectiveness for the procedure and enhanced patient safety.
导言:脐静脉导管是新生儿重症监护室中早产儿和重症婴儿的重要通路设备。正确定位至关重要,因为定位不当会导致严重的并发症。根据国际指南,必须实时评估脐静脉导管尖端的位置;传统上,导管通过胸腹部 X 光片观察,但最有效、最安全的方法之一是实时超声:本研究比较了 461 个病例中评估脐静脉导管尖端位置的实时超声和传统 X 光方法。回顾性分析了导管尖端错位率。次要目的是评估脐静脉导管的留置时间和移除原因:与 X 光评估相比,实时超声尖端定位更可靠、更高效,原发性错位的发生率明显较低(9.6% 对 75.9%)。该研究还强调了实时超声与导管操作减少、X 光检查次数减少和脐静脉导管留置时间延长的关系。多元逻辑回归结果显示,使用实时超声检查导管尖端位置,脐静脉导管尖端中心安全位置的概率很高(几率比29.5,95%置信区间:17.4-49.4):研究结果支持在临床环境中采用实时超声来提高脐静脉导管置管的准确性,并将相关风险降至最低。只需投入极少的培训费用即可熟练掌握脐静脉导管的可视化操作,在手术的成本效益和提高患者安全性方面都具有很大的优势。