Validation of a novel professional automated auscultatory upper-arm cuff blood pressure monitor in a general population according to the AAMI/ESH/ISO Universal Standard: KOROT V2 Doctor (InBody BPBIO280KV).

IF 1.2 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE Blood Pressure Monitoring Pub Date : 2024-08-01 Epub Date: 2024-05-17 DOI:10.1097/MBP.0000000000000707
Aikaterini Theodosiadi, Vasileios Ntousopoulos, Angeliki Ntineri, Panagiota Stathopoulou, Ariadni Menti, Konstantinos G Kyriakoulis, Anastasios Kollias, George S Stergiou
{"title":"Validation of a novel professional automated auscultatory upper-arm cuff blood pressure monitor in a general population according to the AAMI/ESH/ISO Universal Standard: KOROT V2 Doctor (InBody BPBIO280KV).","authors":"Aikaterini Theodosiadi, Vasileios Ntousopoulos, Angeliki Ntineri, Panagiota Stathopoulou, Ariadni Menti, Konstantinos G Kyriakoulis, Anastasios Kollias, George S Stergiou","doi":"10.1097/MBP.0000000000000707","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>A novel automated auscultatory upper arm cuff blood pressure (BP) monitor KOROT V2 Doctor (InBody BPBIO280KV) was developed for professional use. An electronic stethoscope embedded in the device cuff records the Korotkoff sounds, which are graphically displayed during deflation allowing visual evaluation by the healthcare professional. The device provides automated measurements of BP and this study evaluated its accuracy.</p><p><strong>Methods: </strong>The requirements of the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01 were applied. Participants were recruited to fulfill the age, sex, BP, arm circumference, and cuff distribution criteria of the Universal Standard and its Amendment in a general population using the same arm sequential measurement method. Three cuffs of the test device were tested for arm circumference 23-28, 28-35, and 33-42 cm.</p><p><strong>Results: </strong>Data from 85 individuals were analyzed [mean age: 56.4 ± 16.0 (SD) years, 50 men, arm circumference 23-42 cm]. For validation Criterion 1, the mean difference ±SD between the test device and reference BP readings (N = 255) was -1.3 ± 6.0/1.5 ± 5.0 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For Criterion 2, the SD of the averaged BP differences per individual (N = 85) was 4.61/3.48 mmHg (systolic/diastolic; threshold ≤6.82/6.78 mmHg).</p><p><strong>Conclusion: </strong>The KOROT V2 Doctor (InBody BPBIO280KV) device for professional use, which provides automated auscultatory measurements with visual display of the Korotkoff sounds, comfortably fulfills all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in a general population and can be recommended for clinical use.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000707","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/17 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: A novel automated auscultatory upper arm cuff blood pressure (BP) monitor KOROT V2 Doctor (InBody BPBIO280KV) was developed for professional use. An electronic stethoscope embedded in the device cuff records the Korotkoff sounds, which are graphically displayed during deflation allowing visual evaluation by the healthcare professional. The device provides automated measurements of BP and this study evaluated its accuracy.

Methods: The requirements of the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01 were applied. Participants were recruited to fulfill the age, sex, BP, arm circumference, and cuff distribution criteria of the Universal Standard and its Amendment in a general population using the same arm sequential measurement method. Three cuffs of the test device were tested for arm circumference 23-28, 28-35, and 33-42 cm.

Results: Data from 85 individuals were analyzed [mean age: 56.4 ± 16.0 (SD) years, 50 men, arm circumference 23-42 cm]. For validation Criterion 1, the mean difference ±SD between the test device and reference BP readings (N = 255) was -1.3 ± 6.0/1.5 ± 5.0 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For Criterion 2, the SD of the averaged BP differences per individual (N = 85) was 4.61/3.48 mmHg (systolic/diastolic; threshold ≤6.82/6.78 mmHg).

Conclusion: The KOROT V2 Doctor (InBody BPBIO280KV) device for professional use, which provides automated auscultatory measurements with visual display of the Korotkoff sounds, comfortably fulfills all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in a general population and can be recommended for clinical use.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
根据 AAMI/ESH/ISO 通用标准在普通人群中验证新型专业自动听诊式上臂袖带血压计:KOROT V2 Doctor (InBody BPBIO280KV)。
目的:为专业人员开发了一种新型自动听诊式上臂袖带血压(BP)监测仪 KOROT V2 Doctor(InBody BPBIO280KV)。设备袖带中嵌入的电子听诊器可记录 Korotkoff 音,放气时以图形显示,以便医护人员进行可视化评估。该设备可自动测量血压,本研究对其准确性进行了评估:研究采用了医疗仪器促进协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)通用标准(ISO 81060-2:2018)及其 1.2020-01 修正案的要求。采用相同的手臂顺序测量方法,在普通人群中招募符合通用标准及其修订版规定的年龄、性别、血压、臂围和袖带分布标准的参与者。测试设备的三个袖带分别测试了臂围 23-28、28-35 和 33-42 厘米:分析了 85 人的数据[平均年龄:56.4 ± 16.0 (SD) 岁,男性 50 人,臂围 23-42 厘米]。对于验证标准 1,测试设备和参考血压读数(N = 255)之间的平均差(±SD)为-1.3 ± 6.0/1.5 ± 5.0 mmHg(收缩压/舒张压;阈值≤5 ± 8 mmHg)。对于标准 2,每个人(N = 85)的平均血压差异的 SD 为 4.61/3.48 mmHg(收缩压/舒张压;阈值≤6.82/6.78 mmHg):KOROT V2 Doctor(InBody BPBIO280KV)专业设备可提供自动听诊测量,并可视化显示 Korotkoff 音,在普通人群中可轻松满足 AAMI/ESH/ISO 通用标准(ISO 81060-2:2018)的所有要求,建议临床使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
期刊最新文献
Does greater adherence to a healthy dietary pattern correspond to a better body composition and cardiorespiratory fitness in adults with overweight and primary hypertension? Data from the EXERDIET-HTA study. Validation of the TMB-2296-BT blood pressure monitor in adults according to the ISO 81060-2:2018 + Amd.1:2020. Inflammation-based markers, especially the uric acid/albumin ratio, are associated with non-dipper pattern in newly diagnosed treatment-naive hypertensive patients. Validating the accuracy of Omron HEM-7372T1-AZAZ (BP5460) in monitoring blood pressure according to the ISO 81060-2:2018+Amd 1:2020 protocol in the general population. Blood pressure variability at rest and during pressor challenges in patients with acute ischemic stroke.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1