{"title":"[Design and application of alaryngeal mask and monitoring device facilitating withdrawal of endotracheal tube].","authors":"Qionglei Ding, Xiaobao Lei, Jiaxiong Deng, Xiang Wang, Tiao Li, Guicheng Li","doi":"10.3760/cma.j.cn121430-20240129-00090","DOIUrl":null,"url":null,"abstract":"<p><p>Percutaneous dilatational tracheostomy (PDT) is a surgical method for quickly establishing an artificial airway, which has been favored by clinicians because of its simple operation, small trauma and bedside operation. However, for patients with tracheal intubation in intensive care unit (ICU), the tip and balloon of the existing endotracheal tube will not only hinder percutaneous puncture, but also hinder insertion of guidewire and tracheotomy tube, and consequently affect the process of PDT. On the contrary, blind withdrawal of the existing endotracheal tube may cause the tracheal tube tipleave the glottis, leading to an emergency airway situation that endangers the patient's life. Therefore, the medical staff from intensive care medicine department of the First People's Hospital of Chenzhou designed a laryngeal mask and its monitoring device, which is convenient for withdrawal of endotracheal tube, and obtained the national utility model patent of China (patent number: ZL 2020 2 2795887.1). The device is composed of a laryngeal mask and a monitoring device. The laryngeal mask mainly includes a laryngeal mask body, a vent tube, a guidance tube and other components. The laryngeal mask body is mainly used to seal the throat and provide the air supply channel for the patient together with the ventilation tube. The main function of the guidance tube is to accommodate the tracheal tube and facilitate the withdrawal of the inserted tracheal tube. During percutaneous dilatation tracheotomy, this device can monitor the withdrawal of tracheal catheter in real time, and immediately ensure the airway patency of patients without re-intubation when the cuff of tracheal catheter exits the glottis. The utility model has the advantages of real-time monitoring, simple operation, safety and convenience, and is worthy of transformation and promotion.</p>","PeriodicalId":24079,"journal":{"name":"Zhonghua wei zhong bing ji jiu yi xue","volume":"36 6","pages":"649-651"},"PeriodicalIF":0.0000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zhonghua wei zhong bing ji jiu yi xue","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3760/cma.j.cn121430-20240129-00090","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Percutaneous dilatational tracheostomy (PDT) is a surgical method for quickly establishing an artificial airway, which has been favored by clinicians because of its simple operation, small trauma and bedside operation. However, for patients with tracheal intubation in intensive care unit (ICU), the tip and balloon of the existing endotracheal tube will not only hinder percutaneous puncture, but also hinder insertion of guidewire and tracheotomy tube, and consequently affect the process of PDT. On the contrary, blind withdrawal of the existing endotracheal tube may cause the tracheal tube tipleave the glottis, leading to an emergency airway situation that endangers the patient's life. Therefore, the medical staff from intensive care medicine department of the First People's Hospital of Chenzhou designed a laryngeal mask and its monitoring device, which is convenient for withdrawal of endotracheal tube, and obtained the national utility model patent of China (patent number: ZL 2020 2 2795887.1). The device is composed of a laryngeal mask and a monitoring device. The laryngeal mask mainly includes a laryngeal mask body, a vent tube, a guidance tube and other components. The laryngeal mask body is mainly used to seal the throat and provide the air supply channel for the patient together with the ventilation tube. The main function of the guidance tube is to accommodate the tracheal tube and facilitate the withdrawal of the inserted tracheal tube. During percutaneous dilatation tracheotomy, this device can monitor the withdrawal of tracheal catheter in real time, and immediately ensure the airway patency of patients without re-intubation when the cuff of tracheal catheter exits the glottis. The utility model has the advantages of real-time monitoring, simple operation, safety and convenience, and is worthy of transformation and promotion.