Alex J. Nusbickel, Stephen Allan Petty, Steven J. Ross, Alex Parker, Juan Vilaro, Mustafa M. Ahmed
{"title":"Left ventricular assist device implantation outcomes in patients with subcutaneous implantable cardioverter-defibrillators: A case series","authors":"Alex J. Nusbickel, Stephen Allan Petty, Steven J. Ross, Alex Parker, Juan Vilaro, Mustafa M. Ahmed","doi":"10.1016/j.ahjo.2024.100426","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Left ventricular assist devices (LVADs) may induce electromagnetic interference (EMI) affecting implanted cardiac devices, including more novel subcutaneous implantable cardiac defibrillators (S-ICDs).</p></div><div><h3>Methods</h3><p>In this case series, the authors retrospectively reviewed courses of 6 patients with S-ICDs who underwent LVAD implantation at a single center.</p></div><div><h3>Results</h3><p>Of the 6 patients reviewed, 4 experienced inappropriate ICD shocks, of which 3 resulted from EMI. Five of the 6 patients ultimately had S-ICD therapies disabled.</p></div><div><h3>Conclusions</h3><p>Due to EMI resulting in inappropriate shocks and improved tolerability of malignant arrhythmias, deactivation or removal of S-ICDs should be considered in patients undergoing LVAD implantation.</p></div>","PeriodicalId":72158,"journal":{"name":"American heart journal plus : cardiology research and practice","volume":"45 ","pages":"Article 100426"},"PeriodicalIF":1.3000,"publicationDate":"2024-07-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2666602224000697/pdfft?md5=5ad351a98a56de6a9c4bc880919e2c18&pid=1-s2.0-S2666602224000697-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American heart journal plus : cardiology research and practice","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666602224000697","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Left ventricular assist devices (LVADs) may induce electromagnetic interference (EMI) affecting implanted cardiac devices, including more novel subcutaneous implantable cardiac defibrillators (S-ICDs).
Methods
In this case series, the authors retrospectively reviewed courses of 6 patients with S-ICDs who underwent LVAD implantation at a single center.
Results
Of the 6 patients reviewed, 4 experienced inappropriate ICD shocks, of which 3 resulted from EMI. Five of the 6 patients ultimately had S-ICD therapies disabled.
Conclusions
Due to EMI resulting in inappropriate shocks and improved tolerability of malignant arrhythmias, deactivation or removal of S-ICDs should be considered in patients undergoing LVAD implantation.