Motion-induced dose perturbations in photon radiotherapy and proton therapy measured by deformable liver-shaped 3D dosimeters in an anthropomorphic phantom

Simon Vindbæk , Stefanie Ehrbar , Esben Worm , Ludvig Muren , Stephanie Tanadini-Lang , Jørgen Petersen , Peter Balling , Per Poulsen
{"title":"Motion-induced dose perturbations in photon radiotherapy and proton therapy measured by deformable liver-shaped 3D dosimeters in an anthropomorphic phantom","authors":"Simon Vindbæk ,&nbsp;Stefanie Ehrbar ,&nbsp;Esben Worm ,&nbsp;Ludvig Muren ,&nbsp;Stephanie Tanadini-Lang ,&nbsp;Jørgen Petersen ,&nbsp;Peter Balling ,&nbsp;Per Poulsen","doi":"10.1016/j.phro.2024.100609","DOIUrl":null,"url":null,"abstract":"<div><h3>Background and purpose</h3><p>The impact of intrafractional motion and deformations on clinical radiotherapy delivery has so far only been investigated by simulations as well as point and planar dose measurements. The aim of this study was to combine anthropomorphic 3D dosimetry with a deformable abdominal phantom to measure the influence of intra-fractional motion and gating in photon radiotherapy and evaluate the applicability in proton therapy.</p></div><div><h3>Material and methods</h3><p>An abdominal phantom was modified to hold a deformable anthropomorphic 3D dosimeter shaped as a human liver. A liver-specific photon radiotherapy and a proton pencil beam scanning therapy plan were delivered to the phantom without motion as well as with 12 mm sinusoidal motion while using either no respiratory gating or respiratory gating.</p></div><div><h3>Results</h3><p>Using the stationary irradiation as reference the local 3 %/2 mm 3D gamma index pass rate of the motion experiments in the planning target volume (PTV) was above 97 % (photon) and 78 % (proton) with gating whereas it was below 74 % (photon) and 45 % (proton) without gating.</p></div><div><h3>Conclusions</h3><p>For the first time a high-resolution deformable anthropomorphic 3D dosimeter embedded in a deformable abdominal phantom was applied for experimental validation of both photon and proton treatments of targets exhibiting respiratory motion. It was experimentally shown that gating improves dose coverage and the geometrical accuracy for both photon radiotherapy and proton therapy.</p></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"31 ","pages":"Article 100609"},"PeriodicalIF":3.4000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2405631624000794/pdfft?md5=4d59bb653313cf7c63ec5bcea269a7c2&pid=1-s2.0-S2405631624000794-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631624000794","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background and purpose

The impact of intrafractional motion and deformations on clinical radiotherapy delivery has so far only been investigated by simulations as well as point and planar dose measurements. The aim of this study was to combine anthropomorphic 3D dosimetry with a deformable abdominal phantom to measure the influence of intra-fractional motion and gating in photon radiotherapy and evaluate the applicability in proton therapy.

Material and methods

An abdominal phantom was modified to hold a deformable anthropomorphic 3D dosimeter shaped as a human liver. A liver-specific photon radiotherapy and a proton pencil beam scanning therapy plan were delivered to the phantom without motion as well as with 12 mm sinusoidal motion while using either no respiratory gating or respiratory gating.

Results

Using the stationary irradiation as reference the local 3 %/2 mm 3D gamma index pass rate of the motion experiments in the planning target volume (PTV) was above 97 % (photon) and 78 % (proton) with gating whereas it was below 74 % (photon) and 45 % (proton) without gating.

Conclusions

For the first time a high-resolution deformable anthropomorphic 3D dosimeter embedded in a deformable abdominal phantom was applied for experimental validation of both photon and proton treatments of targets exhibiting respiratory motion. It was experimentally shown that gating improves dose coverage and the geometrical accuracy for both photon radiotherapy and proton therapy.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在拟人模型中使用可变形肝形三维剂量计测量光子放射治疗和质子治疗中运动引起的剂量扰动
背景和目的迄今为止,人们仅通过模拟以及点剂量和平面剂量测量来研究点内运动和变形对临床放射治疗的影响。本研究旨在将拟人三维剂量测量与可变形腹部模型相结合,测量光子放疗中的点内运动和门控的影响,并评估其在质子治疗中的适用性。结果以静止辐照为参考,规划靶体积(PTV)内运动实验的局部 3 %/2 mm 3D 伽玛指数通过率在有门控的情况下高于 97 %(光子)和 78 %(质子),而在无门控的情况下低于 74 %(光子)和 45 %(质子)。结论首次将嵌入可变形腹部模型中的高分辨率可变形拟人三维剂量计用于对表现出呼吸运动的目标进行光子和质子治疗的实验验证。实验表明,门控提高了光子放疗和质子治疗的剂量覆盖率和几何精度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Physics and Imaging in Radiation Oncology
Physics and Imaging in Radiation Oncology Physics and Astronomy-Radiation
CiteScore
5.30
自引率
18.90%
发文量
93
审稿时长
6 weeks
期刊最新文献
Results of 2023 survey on the use of synthetic computed tomography for magnetic resonance Imaging-only radiotherapy: Current status and future steps Head and neck automatic multi-organ segmentation on Dual-Energy Computed Tomography Automatic segmentation for magnetic resonance imaging guided individual elective lymph node irradiation in head and neck cancer patients Development of a novel 3D-printed dynamic anthropomorphic thorax phantom for evaluation of four-dimensional computed tomography Technical feasibility of delivering a simultaneous integrated boost in partial breast irradiation
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1