Preparation and certification of benzoic acid reference material for calorimetry analysis

{"title":"Preparation and certification of benzoic acid reference material for calorimetry analysis","authors":"","doi":"10.1016/j.meaene.2024.100016","DOIUrl":null,"url":null,"abstract":"<div><p>Benzoic acid reference material is indispensable for bomb calorimeter instrument calibration and validation of gross calorific value (GCV) analysis of any substance. In this work, we demonstrated the preparation of benzoic acid reference material through a homogeneity study, round-robin analysis, and stability study. Two-factor analysis of variance (ANOVA) test for gross calorific value in the randomly selected sub-samples of benzoic acid exhibits a lower <em>F</em><sub><em>TS</em></sub> value than the <em>F</em><sub><em>crit</em></sub> value, indicating that the samples are sufficiently homogeneous. The calculated uncertainty of between-bottle (<em>u</em><sub><em>bb</em></sub>) and uncertainty of homogeneity (<em>u</em><sub><em>hom</em></sub>) for GCV of benzoic acid in the sub-samples were found as 1.82 and 4.42 ​cal/g respectively. We found that the observed homogeneity (<em>u</em><sub><em>hom (Finding)</em></sub>) value is lower than the assumed homogeneity (<em>u</em><sub><em>hom (Assume)</em></sub>) value for the prepared benzoic acid reference material. Overall observations confirm that the sub-samples are sufficiently homogeneous. Moreover, the round-robin analysis/or inter-laboratory comparison analysis was conducted to assign the gross calorific value and determine the characterization uncertainty. The seventh order of Grubbs' analysis was done using robust estimator Alogoritm A to assign the GCV of benzoic acid. Finally, the measurement uncertainty of the assigned GCV of benzoic acid was calculated with the combined uncertainties from various sources.</p></div>","PeriodicalId":100897,"journal":{"name":"Measurement: Energy","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-07-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2950345024000162/pdfft?md5=043b511c377475bdbd886ff5c0dda2c6&pid=1-s2.0-S2950345024000162-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Measurement: Energy","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2950345024000162","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Benzoic acid reference material is indispensable for bomb calorimeter instrument calibration and validation of gross calorific value (GCV) analysis of any substance. In this work, we demonstrated the preparation of benzoic acid reference material through a homogeneity study, round-robin analysis, and stability study. Two-factor analysis of variance (ANOVA) test for gross calorific value in the randomly selected sub-samples of benzoic acid exhibits a lower FTS value than the Fcrit value, indicating that the samples are sufficiently homogeneous. The calculated uncertainty of between-bottle (ubb) and uncertainty of homogeneity (uhom) for GCV of benzoic acid in the sub-samples were found as 1.82 and 4.42 ​cal/g respectively. We found that the observed homogeneity (uhom (Finding)) value is lower than the assumed homogeneity (uhom (Assume)) value for the prepared benzoic acid reference material. Overall observations confirm that the sub-samples are sufficiently homogeneous. Moreover, the round-robin analysis/or inter-laboratory comparison analysis was conducted to assign the gross calorific value and determine the characterization uncertainty. The seventh order of Grubbs' analysis was done using robust estimator Alogoritm A to assign the GCV of benzoic acid. Finally, the measurement uncertainty of the assigned GCV of benzoic acid was calculated with the combined uncertainties from various sources.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
制备和认证用于量热分析的苯甲酸标准物质
苯甲酸标准物质是炸弹量热仪校准和验证任何物质总热值(GCV)分析所不可或缺的。在这项工作中,我们通过均一性研究、循环分析和稳定性研究证明了苯甲酸标准物质的制备方法。在随机抽取的苯甲酸子样品中,总热值的双因素方差分析(ANOVA)检验显示 FTS 值低于 Fcrit 值,表明样品具有足够的均匀性。计算得出的苯甲酸子样品总热值的瓶间不确定度(ubb)和均质性不确定度(uhom)分别为 1.82 和 4.42 卡/克。我们发现观察到的均一性(uhom (Finding))值低于制备的苯甲酸标准物质的假定均一性(uhom (Assume))值。总体观察结果证实,子样品具有足够的均匀性。此外,还进行了循环分析/或实验室间比对分析,以确定总热值和表征不确定性。使用稳健估计器 Alogoritm A 进行格拉布斯七阶分析,以确定苯甲酸的总热值。最后,利用各种来源的综合不确定性,计算出分配的苯甲酸 GCV 的测量不确定性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
PMU-based voltage estimation and distributed generation effects in active distribution networks An optimization approach for enhancing energy efficiency, reducing CO2 emission, and improving lubrication reliability in roller bearings using ABC algorithm Analysis of transmission pathways of combustion-induced vibration in a diesel engine using wavelet cross-correlation analysis method Accelerated lithium-ion battery cycle lifetime testing by condition-based reference performance tests New parameters for the capacitive accelerometer to reduce its measurement error and power consumption
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1