Reduction with zinc — Impact on the determination of nitrite and nitrate ions using microfluidic paper-based analytical devices

IF 4.1 Q1 CHEMISTRY, ANALYTICAL Talanta Open Pub Date : 2024-08-10 DOI:10.1016/j.talo.2024.100347
Mika I. Umeda , Kaewta Danchana , Takatoshi Fujii , Eiichi Hino , Yusuke Date , Kaoru Aoki , Takashi Kaneta
{"title":"Reduction with zinc — Impact on the determination of nitrite and nitrate ions using microfluidic paper-based analytical devices","authors":"Mika I. Umeda ,&nbsp;Kaewta Danchana ,&nbsp;Takatoshi Fujii ,&nbsp;Eiichi Hino ,&nbsp;Yusuke Date ,&nbsp;Kaoru Aoki ,&nbsp;Takashi Kaneta","doi":"10.1016/j.talo.2024.100347","DOIUrl":null,"url":null,"abstract":"<div><p>We used a microfluidic paper-based analytical device (μPAD) to investigate the influence that zinc reduction exerts on the determination of nitrite and nitrate ions in natural water samples. The μPAD consists of layered channels for the reduction of nitrate to nitrite with zinc powder and the subsequent detection of nitrite with Griess reagent. The amount of zinc, number of layers, and reaction time for the reduction were optimized to obtain an intense signal for nitrate. Initially, the sensitivity to nitrate corresponded to 55% that of nitrite, which implied an incomplete reduction. We found, however, that zinc decreased the sensitivity to nitrite in both the μPAD and spectrophotometry. The sensitivity to nitrite was decreased by 48% in spectrophotometry and 68% in the μPAD following the reaction with zinc. One of the reasons for the decreased sensitivity is attributed to the production of ammonia, as we elucidated that both nitrite and nitrate produced ammonia via the reaction with zinc. The results suggest that the total concentration of nitrite and nitrate must be corrected by constructing a calibration curve for nitrite with zinc, in addition to developing curves for nitrate with zinc and for nitrite without zinc. Using these calibration curves, the absorbance at different concentration ratios of nitrite and nitrate ions could be reproduced via calculation using the calibration curves with zinc for nitrite and nitrate. Eventually, the developed μPAD was applied to the determination of nitrite and nitrate ions in natural water samples, and the results were compared with those using a conventional spectrophotometric method. The results of the μPAD are in good agreement with those of conventional spectrophotometry, which suggests that the μPAD is reliable for the measurement of nitrite and nitrate ions in natural water samples.</p></div>","PeriodicalId":436,"journal":{"name":"Talanta Open","volume":"10 ","pages":"Article 100347"},"PeriodicalIF":4.1000,"publicationDate":"2024-08-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2666831924000614/pdfft?md5=13992163d8ee60e291dc8566daaa7495&pid=1-s2.0-S2666831924000614-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Talanta Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666831924000614","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

We used a microfluidic paper-based analytical device (μPAD) to investigate the influence that zinc reduction exerts on the determination of nitrite and nitrate ions in natural water samples. The μPAD consists of layered channels for the reduction of nitrate to nitrite with zinc powder and the subsequent detection of nitrite with Griess reagent. The amount of zinc, number of layers, and reaction time for the reduction were optimized to obtain an intense signal for nitrate. Initially, the sensitivity to nitrate corresponded to 55% that of nitrite, which implied an incomplete reduction. We found, however, that zinc decreased the sensitivity to nitrite in both the μPAD and spectrophotometry. The sensitivity to nitrite was decreased by 48% in spectrophotometry and 68% in the μPAD following the reaction with zinc. One of the reasons for the decreased sensitivity is attributed to the production of ammonia, as we elucidated that both nitrite and nitrate produced ammonia via the reaction with zinc. The results suggest that the total concentration of nitrite and nitrate must be corrected by constructing a calibration curve for nitrite with zinc, in addition to developing curves for nitrate with zinc and for nitrite without zinc. Using these calibration curves, the absorbance at different concentration ratios of nitrite and nitrate ions could be reproduced via calculation using the calibration curves with zinc for nitrite and nitrate. Eventually, the developed μPAD was applied to the determination of nitrite and nitrate ions in natural water samples, and the results were compared with those using a conventional spectrophotometric method. The results of the μPAD are in good agreement with those of conventional spectrophotometry, which suggests that the μPAD is reliable for the measurement of nitrite and nitrate ions in natural water samples.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
锌还原--对使用微流控纸质分析装置测定亚硝酸盐和硝酸盐离子的影响
我们利用微流体纸基分析装置(μPAD)研究了锌还原对天然水样中亚硝酸盐和硝酸根离子测定的影响。μPAD 由分层通道组成,用于用锌粉将硝酸盐还原成亚硝酸盐,然后用格里丝试剂检测亚硝酸盐。对锌的用量、层数和还原反应时间进行了优化,以获得强烈的硝酸盐信号。最初,对硝酸盐的灵敏度为亚硝酸盐的 55%,这意味着还原不完全。但我们发现,锌降低了μPAD和分光光度法对亚硝酸盐的灵敏度。与锌反应后,分光光度法对亚硝酸盐的敏感性降低了 48%,μPAD 降低了 68%。灵敏度降低的原因之一是氨的产生,因为我们发现亚硝酸盐和硝酸盐都会通过与锌的反应产生氨。结果表明,亚硝酸盐和硝酸盐的总浓度必须通过构建一条含锌亚硝酸盐的校准曲线来校正,此外还必须绘制含锌硝酸盐和不含锌亚硝酸盐的曲线。利用这些校准曲线,通过计算亚硝酸盐和硝酸盐的含锌校准曲线,可以再现亚硝酸盐和硝酸盐离子在不同浓度比时的吸光度。最后,将开发的 μPAD 应用于天然水样中亚硝酸盐和硝酸盐离子的测定,并将结果与传统的分光光度法进行了比较。结果表明,μPAD 与传统分光光度法的结果非常吻合,这表明μPAD 可以可靠地测定天然水样中的亚硝酸盐和硝酸盐离子。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Talanta Open
Talanta Open Chemistry-Analytical Chemistry
CiteScore
5.20
自引率
0.00%
发文量
86
审稿时长
49 days
期刊最新文献
Continuous-flow analysis of nitrogen compounds in environmental water using a copper–zinc reduction coil Development of engineered Zn-MOF/g-C3N4 based photoelectrochemical system for real-time sensors and removal of naproxen in wastewater Development of nanomaterial-supported molecularly imprinted polymer/receptor-like sensor for the detection of rosuvastatin from binary mixtures Trace-level quantification of NDMA in levosulpuride active pharmaceutical ingredient and tablet formulation Using UFLC-MS/MS Enhancing isomer specificity in mass spectrometry by combining silver ion adduction and ion mobility
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1