{"title":"3-Year Outcomes of Balloon-Expandable Valves","authors":"","doi":"10.1016/j.jcin.2024.06.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>A prior Society of Thoracic Surgeons/American College of Cardiology TVT (Transcatheter Valve Therapy) Registry–based analysis reported similar 1-year clinical outcomes with small (20-mm) vs large (≥23-mm) balloon-expandable valves (BEV).</p></div><div><h3>Objectives</h3><p>The aim of this study was to describe mid-term 3-year clinical outcomes for small vs large BEV and the relationship between discharge echocardiographic mean gradient (MG) and different definitions of prothesis-patient mismatch (PPM) with clinical outcomes.</p></div><div><h3>Methods</h3><p>Using the TVT Registry with Centers for Medicare and Medicaid Services linkage, a propensity-matched analysis of patients receiving 20- vs ≥23-mm BEVs was performed. Spline curves and Kaplan-Meier plots with adjusted HRs determined the relationship between MG and 3-year mortality.</p></div><div><h3>Results</h3><p>In total, 316,091 patients were analyzed; after propensity matching, 8,100 pairs of each group were compared. The 20-mm BEV was associated with higher MGs compared with ≥23-mm BEVs (16.2 ± 7.2 mm Hg vs 11.8 ± 5.7 mm Hg; <em>P</em> < 0.0001). At 3 years, there was no difference in mortality between 20- and ≥23-mm BEVs (31.5% vs 32.5%, respectively; HR: 0.97; 95% CI: 0.90-1.05). Compared with an MG of 10 to 30 mm Hg, an MG <10 mm Hg (HR: 1.25; 95% CI:1.22-1.27) was associated with increased 3-year mortality. Measured severe PPM and predicted no PPM were associated with increased 3-year mortality (33.5% vs 32.9% vs 32.1%; <em>P</em> < 0.0001) and (33.5% vs 31.1% vs 30%; <em>P</em> < 0.0001), respectively. Low MG and severe measured PPM were associated with lower left ventricular ejection fraction (LVEF).</p></div><div><h3>Conclusions</h3><p>Patients with small-prosthesis BEVs (20 mm) had identical 3-year survival as those with larger (≥23-mm) BEV valves. Severe measured PPM and low MG (<10 mm Hg), but not predicted severe PPM, were associated with lower LVEF and increased mortality, suggesting that LVEF is the culprit for worse outcomes.</p></div>","PeriodicalId":14688,"journal":{"name":"JACC. Cardiovascular interventions","volume":null,"pages":null},"PeriodicalIF":11.7000,"publicationDate":"2024-09-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1936879824008574/pdfft?md5=208ccfa7d91dfbfede1b2c638328c8f0&pid=1-s2.0-S1936879824008574-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JACC. Cardiovascular interventions","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1936879824008574","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
A prior Society of Thoracic Surgeons/American College of Cardiology TVT (Transcatheter Valve Therapy) Registry–based analysis reported similar 1-year clinical outcomes with small (20-mm) vs large (≥23-mm) balloon-expandable valves (BEV).
Objectives
The aim of this study was to describe mid-term 3-year clinical outcomes for small vs large BEV and the relationship between discharge echocardiographic mean gradient (MG) and different definitions of prothesis-patient mismatch (PPM) with clinical outcomes.
Methods
Using the TVT Registry with Centers for Medicare and Medicaid Services linkage, a propensity-matched analysis of patients receiving 20- vs ≥23-mm BEVs was performed. Spline curves and Kaplan-Meier plots with adjusted HRs determined the relationship between MG and 3-year mortality.
Results
In total, 316,091 patients were analyzed; after propensity matching, 8,100 pairs of each group were compared. The 20-mm BEV was associated with higher MGs compared with ≥23-mm BEVs (16.2 ± 7.2 mm Hg vs 11.8 ± 5.7 mm Hg; P < 0.0001). At 3 years, there was no difference in mortality between 20- and ≥23-mm BEVs (31.5% vs 32.5%, respectively; HR: 0.97; 95% CI: 0.90-1.05). Compared with an MG of 10 to 30 mm Hg, an MG <10 mm Hg (HR: 1.25; 95% CI:1.22-1.27) was associated with increased 3-year mortality. Measured severe PPM and predicted no PPM were associated with increased 3-year mortality (33.5% vs 32.9% vs 32.1%; P < 0.0001) and (33.5% vs 31.1% vs 30%; P < 0.0001), respectively. Low MG and severe measured PPM were associated with lower left ventricular ejection fraction (LVEF).
Conclusions
Patients with small-prosthesis BEVs (20 mm) had identical 3-year survival as those with larger (≥23-mm) BEV valves. Severe measured PPM and low MG (<10 mm Hg), but not predicted severe PPM, were associated with lower LVEF and increased mortality, suggesting that LVEF is the culprit for worse outcomes.
期刊介绍:
JACC: Cardiovascular Interventions is a specialist journal launched by the Journal of the American College of Cardiology (JACC). It covers the entire field of interventional cardiovascular medicine, including cardiac, peripheral, and cerebrovascular interventions. The journal publishes studies that will impact the practice of interventional cardiovascular medicine, including clinical trials, experimental studies, and in-depth discussions by respected experts. To enhance visual understanding, the journal is published both in print and electronically, utilizing the latest technologies.