{"title":"Semi-automated vertex placement for lattice radiotherapy and dosimetric verification using 3D polymer gel dosimetry","authors":"Tenzin Kunkyab, Anthony Magliari, Andrew Jirasek, Benjamin Mou, Derek Hyde","doi":"10.1002/acm2.14489","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To evaluate the feasibility of an open-source, semi-automated, and reproducible vertex placement tool to improve the efficiency of lattice radiotherapy (LRT) planning. We used polymer gel dosimetry with a Cone Beam CT (CBCT) readout to commission this LRT technique.</p>\n </section>\n \n <section>\n \n <h3> Material and methods</h3>\n \n <p>We generated a volumetric modulated arc therapy (VMAT)-based LRT plan on a 2 L NIPAM polymer gel dosimeter using our Eclipse Acuros version 15.6 AcurosXB beam model, and also recalculated the plan with a pre-clinical Acuros v18.0 dose calculation algorithm with the enhanced leaf modelling (ELM). With the assistance of the MAAS-SFRThelper software, a lattice vertex diameter of 1.5 cm and center-to-center spacing of 3 cm were used to place the spheres in a hexagonal, closed packed structure. The verification plan included four gantry arcs with 15°, 345°, 75°, 105° collimator angles. The spheres were prescribed 20 Gy to 50% of their combined volume. The 6 MV Flattening Filter Free beam energy was used to deliver the verification plan. The dosimetric accuracy of the LRT delivery was evaluated with 1D dose profiles, 2D isodose maps, and a 3D global gamma analysis.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Qualitative comparisons between the 1D dose profiles of the Eclipse plan and measured gel showed good consistency at the prescription dose mark. The average diameter measured 13.3 ± 0.2 mm (gel for v15.6), 12.6 mm (v15.6 plan), 13.1 ± 0.2 mm (gel for v18.0), and 12.3 mm (v18.0 plan). 3D gamma analysis showed that all gamma pass percent were > 95% except at 1% and 2% at the 1 mm distance to agreement criteria.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>This study presents a novel application of gel dosimetry in verifying the dosimetric accuracy of LRT, achieving excellent 3D gamma results. The treatment planning was facilitated by publicly available software that automatically placed the vertices for consistency and efficiency.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":null,"pages":null},"PeriodicalIF":2.0000,"publicationDate":"2024-08-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11540016/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.14489","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
To evaluate the feasibility of an open-source, semi-automated, and reproducible vertex placement tool to improve the efficiency of lattice radiotherapy (LRT) planning. We used polymer gel dosimetry with a Cone Beam CT (CBCT) readout to commission this LRT technique.
Material and methods
We generated a volumetric modulated arc therapy (VMAT)-based LRT plan on a 2 L NIPAM polymer gel dosimeter using our Eclipse Acuros version 15.6 AcurosXB beam model, and also recalculated the plan with a pre-clinical Acuros v18.0 dose calculation algorithm with the enhanced leaf modelling (ELM). With the assistance of the MAAS-SFRThelper software, a lattice vertex diameter of 1.5 cm and center-to-center spacing of 3 cm were used to place the spheres in a hexagonal, closed packed structure. The verification plan included four gantry arcs with 15°, 345°, 75°, 105° collimator angles. The spheres were prescribed 20 Gy to 50% of their combined volume. The 6 MV Flattening Filter Free beam energy was used to deliver the verification plan. The dosimetric accuracy of the LRT delivery was evaluated with 1D dose profiles, 2D isodose maps, and a 3D global gamma analysis.
Results
Qualitative comparisons between the 1D dose profiles of the Eclipse plan and measured gel showed good consistency at the prescription dose mark. The average diameter measured 13.3 ± 0.2 mm (gel for v15.6), 12.6 mm (v15.6 plan), 13.1 ± 0.2 mm (gel for v18.0), and 12.3 mm (v18.0 plan). 3D gamma analysis showed that all gamma pass percent were > 95% except at 1% and 2% at the 1 mm distance to agreement criteria.
Conclusion
This study presents a novel application of gel dosimetry in verifying the dosimetric accuracy of LRT, achieving excellent 3D gamma results. The treatment planning was facilitated by publicly available software that automatically placed the vertices for consistency and efficiency.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic