首页 > 最新文献

Journal of Applied Clinical Medical Physics最新文献

英文 中文
Commissioning of a near real-time in vivo film dosimetry system. 近实时体内膜剂量测定系统的调试。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70525
Yunjie Yang, Seng Boh Lim, Indra Das, Jeonghoon Park, Grace Tang, Dongxu Wang, Megan Hyun, Xiuxiu He, Maria Chan
<p><strong>Background: </strong>Accurate in vivo dosimetry is crucial for dose monitoring of cardiac implantable electronic devices (CIED) and for dose verification for special procedures such as total body irradiation (TBI) and total skin electron therapy (TSET).</p><p><strong>Purpose: </strong>A new near real-time in vivo dosimetry system using radiochromic films (RCF) is investigated for clinical use in megavoltage external beam radiotherapy.</p><p><strong>Methods: </strong>The Pnt-Dos™ in vivo dosimetry system comprises of a new type of RCF and a dedicated software module. Each Pnt-Dos device is a small piece of RCF individually packed with a unique QR code for identification and record keeping. Different from the traditional film dosimetry workflow, where a film developing time of at least 16 hours is recommended, a near real-time dose readout can be achieved with the Pnt-Dos system using a novel calibration procedure. This involves an automated scanning process at user-specified time intervals, utilizing auto-region of interest (ROI) detection and triple-channel calibration to capture the time-resolved post-irradiation growth. Two standard Epson scanner models (V600/13000XL) were used to cross-validate readouts and accommodate users who may prefer to utilize existing 13000XL scanners rather than acquire an additional V600 for in vivo dosimetry. The dosimetric accuracy was evaluated over a range of 15-400 cGy. Angular dependence was studied in 45° increments over 360°, normalized to the response at 0°, at 250 cGy using a cylindrical phantom. Energy dependence was evaluated for four photon energies (6 MV, 6 MV FFF, 10 MV FFF, 15 MV) and five electron energies (6 MeV, 9 MeV, 12 MeV, 16 MeV, and 20 MeV). Long-term reproducibility/stability were assessed with nine devices with different doses under identical conditions, alongside daily scans of quality control (QC) devices over three months.</p><p><strong>Results: </strong>The system provides accurate dose measurements across high- and low-dose ranges. All readings were within specification: accuracy was < ± 5 cGy for doses ≤ 80 cGy doses (max discrepancy 6.0 cGy), and < ± 5% for doses > 80 cGy on average (max discrepancy 5.1%). Angular dependence showed a maximum variation of 2.6% ± 2.1% when the beam passed through the posterior oblique side of the device. Daily QC/reproducibility tests confirmed system constancy of 0.1% average day-to-day variation. Energy dependence analysis revealed deviations of up to 4.9% ± 2.3% for all photon and electron energies compared to 6 MV photons, indicating the need for energy correction during commissioning. Film readings were compared with ion chamber measurements at 10 × 10 cm<sup>2</sup>, d<sub>max</sub>, 100 cm SAD (photons) or 100 cm SSD (electrons). Both scanners provided comparable readouts, within 1.3 cGy for doses ≤ 80 cGy and 0.6% for doses > 80 cGy. Based on these findings, user guidelines were established to ensure optimal performance and accuracy.
背景:准确的体内剂量测定对于心脏植入式电子装置(CIED)的剂量监测以及全身照射(TBI)和全皮肤电子治疗(TSET)等特殊手术的剂量验证至关重要。目的:研究一种基于放射致色膜(RCF)的近实时体内剂量测定系统,用于临床巨压外射束放射治疗。方法:Pnt-Dos™体内剂量测定系统由新型RCF和专用软件模块组成。每个nt- dos设备都是一小块RCF,单独包装有一个独特的QR码,用于识别和记录保存。与传统的胶片剂量测定工作流程不同,传统的胶片显影时间至少为16小时,使用新颖的校准程序,Pnt-Dos系统可以实现近乎实时的剂量读数。这包括在用户指定的时间间隔内自动扫描过程,利用自动感兴趣区域(ROI)检测和三通道校准来捕获时间分辨的辐照后生长。两个标准的爱普生扫描仪型号(V600/13000XL)用于交叉验证读数,并适应可能更喜欢使用现有的13000XL扫描仪而不是购买额外的V600进行体内剂量测定的用户。剂量学准确度在15-400 cGy范围内进行了评估。角度依赖性研究在360°以上45°增量,归一化为响应在0°,在250 cGy使用圆柱形幻影。评估了4种光子能量(6 MV、6 MV FFF、10 MV FFF、15 MV)和5种电子能量(6 MeV、9 MeV、12 MeV、16 MeV和20 MeV)的能量依赖性。在相同条件下,使用9种不同剂量的设备评估长期可重复性/稳定性,并在三个月内每天扫描质量控制(QC)设备。结果:该系统在高剂量和低剂量范围内提供准确的剂量测量。所有读数均在规定范围内:平均精度为80 gy(最大误差5.1%)。当光束穿过装置的后斜侧时,角度依赖性最大变化为2.6%±2.1%。每日QC/重复性测试确认系统的稳定性为0.1%的平均每日变化。能量依赖分析显示,与6 MV光子相比,所有光子和电子能量的偏差高达4.9%±2.3%,表明在调试期间需要进行能量校正。将薄膜读数与离子室在10 × 10 cm2, dmax, 100 cm SAD(光子)或100 cm SSD(电子)下的测量结果进行比较。两种扫描仪都提供了相当的读数,剂量≤80 cGy的读数在1.3 cGy范围内,剂量≤80 cGy的读数为0.6%。基于这些发现,建立了用户指南,以确保最佳的性能和准确性。结论:新的基于薄膜的体内剂量测定系统提供了一个自动化的工作流程,能够实现一致的、时间无关的、近实时的读数,并具有用户友好的设计,简化了处理和分析,从而简化了体内剂量测定。它还提供了患者剂量测定的可追溯记录。
{"title":"Commissioning of a near real-time in vivo film dosimetry system.","authors":"Yunjie Yang, Seng Boh Lim, Indra Das, Jeonghoon Park, Grace Tang, Dongxu Wang, Megan Hyun, Xiuxiu He, Maria Chan","doi":"10.1002/acm2.70525","DOIUrl":"10.1002/acm2.70525","url":null,"abstract":"&lt;p&gt;&lt;strong&gt;Background: &lt;/strong&gt;Accurate in vivo dosimetry is crucial for dose monitoring of cardiac implantable electronic devices (CIED) and for dose verification for special procedures such as total body irradiation (TBI) and total skin electron therapy (TSET).&lt;/p&gt;&lt;p&gt;&lt;strong&gt;Purpose: &lt;/strong&gt;A new near real-time in vivo dosimetry system using radiochromic films (RCF) is investigated for clinical use in megavoltage external beam radiotherapy.&lt;/p&gt;&lt;p&gt;&lt;strong&gt;Methods: &lt;/strong&gt;The Pnt-Dos™ in vivo dosimetry system comprises of a new type of RCF and a dedicated software module. Each Pnt-Dos device is a small piece of RCF individually packed with a unique QR code for identification and record keeping. Different from the traditional film dosimetry workflow, where a film developing time of at least 16 hours is recommended, a near real-time dose readout can be achieved with the Pnt-Dos system using a novel calibration procedure. This involves an automated scanning process at user-specified time intervals, utilizing auto-region of interest (ROI) detection and triple-channel calibration to capture the time-resolved post-irradiation growth. Two standard Epson scanner models (V600/13000XL) were used to cross-validate readouts and accommodate users who may prefer to utilize existing 13000XL scanners rather than acquire an additional V600 for in vivo dosimetry. The dosimetric accuracy was evaluated over a range of 15-400 cGy. Angular dependence was studied in 45° increments over 360°, normalized to the response at 0°, at 250 cGy using a cylindrical phantom. Energy dependence was evaluated for four photon energies (6 MV, 6 MV FFF, 10 MV FFF, 15 MV) and five electron energies (6 MeV, 9 MeV, 12 MeV, 16 MeV, and 20 MeV). Long-term reproducibility/stability were assessed with nine devices with different doses under identical conditions, alongside daily scans of quality control (QC) devices over three months.&lt;/p&gt;&lt;p&gt;&lt;strong&gt;Results: &lt;/strong&gt;The system provides accurate dose measurements across high- and low-dose ranges. All readings were within specification: accuracy was &lt; ± 5 cGy for doses ≤ 80 cGy doses (max discrepancy 6.0 cGy), and &lt; ± 5% for doses &gt; 80 cGy on average (max discrepancy 5.1%). Angular dependence showed a maximum variation of 2.6% ± 2.1% when the beam passed through the posterior oblique side of the device. Daily QC/reproducibility tests confirmed system constancy of 0.1% average day-to-day variation. Energy dependence analysis revealed deviations of up to 4.9% ± 2.3% for all photon and electron energies compared to 6 MV photons, indicating the need for energy correction during commissioning. Film readings were compared with ion chamber measurements at 10 × 10 cm&lt;sup&gt;2&lt;/sup&gt;, d&lt;sub&gt;max&lt;/sub&gt;, 100 cm SAD (photons) or 100 cm SSD (electrons). Both scanners provided comparable readouts, within 1.3 cGy for doses ≤ 80 cGy and 0.6% for doses &gt; 80 cGy. Based on these findings, user guidelines were established to ensure optimal performance and accuracy.","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70525"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12935514/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147306197","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Dosimetric effects of internal margin, auto flash, and virtual bolus in VMAT-based whole-breast radiotherapy under motion variations. 运动变化下基于vmat的全乳房放射治疗中内切、自动闪光和虚拟丸的剂量学效应。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70471
Ji Hyeon Joo, Dong Woon Kim, Wontaek Kim, Jiho Nam, Donghyun Kim, Dahl Park, Youn Joo Jung, Hyun Yul Kim, Ki Seok Choo, Kyung Jin Nam, Su Bong Nam, Jae Joon Kim, Yongkan Ki

Background and purpose: This study aimed to evaluate the dosimetric performance and robustness of volumetric modulated arc therapy (VMAT) planning techniques-planning target volume with internal margin (INT), auto flash (AF), and virtual bolus (VB)-under simulated geometric changes during whole-breast radiotherapy.

Methods: Nine patients with left-sided breast cancer were included. Three planning techniques were compared: INT with a 5-mm skin-sparing internal margin, AF with a 1-cm automatic skin flash margin, and VB with a 5-mm optimization bolus applied only during the planning. Respiratory motion was simulated by shifting the plan center (± 3 mm and ± 5 mm) and recalculating the dose distributions.

Results: Under static conditions, all plans provided adequate target coverage, with planning target volume (PTV) V95% values of 96.45% (INT), 97.35% (AF), and 98.19% (VB). Under breast swelling of up to 5 mm, AF maintained the most stable coverage (PTV V95% = 99.10%), outperforming VB (95.02%) and INT (92.31%) (p < 0.001). In simulated incomplete inspiration (5 mm), VB showed superior robustness, achieving a PTV V95% of 90.46% compared with AF (85.30%) and INT (85.08%) (p = 0.008). AF met the ideal criteria in all cases under swelling conditions, whereas the VB was more effective against suboptimal respiration.

Conclusions: In VMAT breast radiation therapy (RT), the conventional INT approach was the least robust against variations, and additional techniques are required. AF effectively compensates for breast swelling, whereas VB preserves the target coverage under insufficient breath-hold conditions. Surface-guided radiation therapy (SGRT) and visual guidance are recommended to ensure accurate treatment.

背景和目的:本研究旨在评估全乳房放射治疗过程中模拟几何变化下体积调制弧线治疗(VMAT)计划技术的剂量学性能和鲁棒性。VMAT计划技术是指带内缘(INT)、自动闪光(AF)和虚拟丸(VB)的靶体积计划。方法:选取9例左侧乳腺癌患者。比较了三种规划技术:INT与5mm皮肤保留内缘,AF与1cm自动皮肤闪光边缘,VB与5mm优化丸仅在规划期间应用。通过移动计划中心(±3 mm和±5 mm)和重新计算剂量分布来模拟呼吸运动。结果:在静态条件下,各方案均具有足够的目标覆盖率,规划目标体积(PTV) V95%分别为96.45% (INT)、97.35% (AF)和98.19% (VB)。在乳房肿胀达5mm的情况下,AF保持了最稳定的覆盖率(PTV V95% = 99.10%),优于VB(95.02%)和INT (92.31%) (p结论:在VMAT乳房放射治疗(RT)中,传统的INT入路对变化的稳定期最低,需要额外的技术。AF有效地补偿乳房肿胀,而VB在屏气不足的情况下保持目标覆盖。建议表面引导放射治疗(SGRT)和视觉引导以确保准确治疗。
{"title":"Dosimetric effects of internal margin, auto flash, and virtual bolus in VMAT-based whole-breast radiotherapy under motion variations.","authors":"Ji Hyeon Joo, Dong Woon Kim, Wontaek Kim, Jiho Nam, Donghyun Kim, Dahl Park, Youn Joo Jung, Hyun Yul Kim, Ki Seok Choo, Kyung Jin Nam, Su Bong Nam, Jae Joon Kim, Yongkan Ki","doi":"10.1002/acm2.70471","DOIUrl":"10.1002/acm2.70471","url":null,"abstract":"<p><strong>Background and purpose: </strong>This study aimed to evaluate the dosimetric performance and robustness of volumetric modulated arc therapy (VMAT) planning techniques-planning target volume with internal margin (INT), auto flash (AF), and virtual bolus (VB)-under simulated geometric changes during whole-breast radiotherapy.</p><p><strong>Methods: </strong>Nine patients with left-sided breast cancer were included. Three planning techniques were compared: INT with a 5-mm skin-sparing internal margin, AF with a 1-cm automatic skin flash margin, and VB with a 5-mm optimization bolus applied only during the planning. Respiratory motion was simulated by shifting the plan center (± 3 mm and ± 5 mm) and recalculating the dose distributions.</p><p><strong>Results: </strong>Under static conditions, all plans provided adequate target coverage, with planning target volume (PTV) V95% values of 96.45% (INT), 97.35% (AF), and 98.19% (VB). Under breast swelling of up to 5 mm, AF maintained the most stable coverage (PTV V95% = 99.10%), outperforming VB (95.02%) and INT (92.31%) (p < 0.001). In simulated incomplete inspiration (5 mm), VB showed superior robustness, achieving a PTV V95% of 90.46% compared with AF (85.30%) and INT (85.08%) (p = 0.008). AF met the ideal criteria in all cases under swelling conditions, whereas the VB was more effective against suboptimal respiration.</p><p><strong>Conclusions: </strong>In VMAT breast radiation therapy (RT), the conventional INT approach was the least robust against variations, and additional techniques are required. AF effectively compensates for breast swelling, whereas VB preserves the target coverage under insufficient breath-hold conditions. Surface-guided radiation therapy (SGRT) and visual guidance are recommended to ensure accurate treatment.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70471"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12954822/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147344235","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
To shift or not to shift: identifying and correcting patient motion after couch rotations in non-coplanar intracranial radiosurgery with stereoscopic X-ray imaging. 移位或不移位:用立体x线成像识别和纠正非共面颅内放射手术中躺椅旋转后患者的运动。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70505
Vanessa Da Silva Mendes, Sylvia Garny, Lili Huang, Stephan Schönecker, Christian Trapp, Frederik Fuchs, Christopher Kurz, Claus Belka, Guillaume Landry, Michael Reiner, Stefanie Corradini

Background: Frameless linear accelerator (linac)-based image-guided stereotactic radiosurgery (SRS) or fractionated stereotactic radiotherapy (FSRT) are a widely used treatment option for intracranial lesions. Given the high radiation doses involved, it is crucial to maintain precise patient positioning throughout treatment. This requires that geometric inaccuracies arising from patient motion or setup errors are identified and corrected. With frameless immobilization, image-guidance has a greater impact, especially in non-coplanar settings that can lead to patient motion and discrepancies between couch and radiation isocenters.

Purpose: Both patient and phantom studies were conducted to assess and quantify the magnitude of geometric uncertainties after couch rotations, aiming at evaluating the clinical need for their correction to warrant a precise treatment delivery.

Methods: Intrafraction X-ray data, performed by ExacTrac Dynamic (ETD) to monitor and correct patients' position throughout treatment delivery, were collected from 50 patients treated for brain metastases in stereotactic non-coplanar schemes and immobilized by stereotactic double-layered thermoplastic mask systems: 26 patients treated in 40 single-fraction SRS (168 stereoscopic X-ray images); 24 treated with FSRT in 128 fractions (278 stereoscopic X-ray images). Additionally, a head phantom was utilized and 350 measurements under two different couch loads were carried out to distinguish true patient motion from deviations caused by couch rotations or system-related effects. For both studies, ETD stereoscopic X-rays were acquired after each couch rotation and the first measured positioning deviation was calculated by comparing X-ray images to the treatment plan's digitally reconstructed radiographs.

Results: Clinically relevant deviations were observed, exceeding clinical tolerance (≥ 0.5 mm/0.5°) mostly in the lateral and yaw directions and requiring repositioning in nearly half of the couch rotations. These deviations measuring up to 2 mm, revealed to be emerging mainly from patient motion rather than linac setup, as the phantom study showed maximum deviations of up to 0.6 mm and 0.4° when simulating a patient treatment and an interquartile range that did not exceed 0.2 mm and 0.2°.

Conclusions: These findings demonstrate the importance of a continuous intrafraction motion monitoring and repositioning in cranial stereotactic treatments, especially in non-coplanar settings.

背景:基于无框架直线加速器(linac)的图像引导立体定向放射手术(SRS)或分步立体定向放疗(FSRT)是广泛应用于颅内病变的治疗选择。鉴于所涉及的高辐射剂量,在整个治疗过程中保持精确的患者定位至关重要。这需要识别和纠正由患者运动或设置错误引起的几何不准确性。对于无框架固定,图像引导具有更大的影响,特别是在非共面环境中,可能导致患者运动和躺椅与辐射等中心之间的差异。目的:对患者和假体进行研究,以评估和量化沙发旋转后几何不确定性的大小,旨在评估其纠正的临床需要,以保证精确的治疗交付。方法:收集50例采用立体定向非共面方案并采用立体定向双层热塑性掩膜系统固定的脑转移患者的牵拉内x线数据,采用ExacTrac Dynamic (ETD)监测和纠正患者在整个治疗过程中的位置:26例采用40个单部分SRS(168个立体x线图像)治疗;24例接受FSRT治疗,分为128个组(278张立体x线图像)。此外,在两种不同的沙发负荷下,使用头部假体进行350次测量,以区分患者的真实运动与沙发旋转或系统相关影响引起的偏差。在这两项研究中,每次躺椅旋转后都获得ETD立体x射线,并通过将x射线图像与治疗计划的数字重建x射线照片进行比较,计算第一次测量的定位偏差。结果:观察到临床相关的偏差,超过临床耐受性(≥0.5 mm/0.5°),主要发生在侧向和偏航方向,近一半的沙发旋转需要重新定位。这些偏差可达2mm,显示主要来自患者运动而非直线设置,因为模拟患者治疗时,幻影研究显示最大偏差可达0.6 mm和0.4°,四分位数范围不超过0.2 mm和0.2°。结论:这些发现表明,在颅立体定向治疗中,特别是在非共面情况下,持续的屈光内运动监测和重新定位的重要性。
{"title":"To shift or not to shift: identifying and correcting patient motion after couch rotations in non-coplanar intracranial radiosurgery with stereoscopic X-ray imaging.","authors":"Vanessa Da Silva Mendes, Sylvia Garny, Lili Huang, Stephan Schönecker, Christian Trapp, Frederik Fuchs, Christopher Kurz, Claus Belka, Guillaume Landry, Michael Reiner, Stefanie Corradini","doi":"10.1002/acm2.70505","DOIUrl":"10.1002/acm2.70505","url":null,"abstract":"<p><strong>Background: </strong>Frameless linear accelerator (linac)-based image-guided stereotactic radiosurgery (SRS) or fractionated stereotactic radiotherapy (FSRT) are a widely used treatment option for intracranial lesions. Given the high radiation doses involved, it is crucial to maintain precise patient positioning throughout treatment. This requires that geometric inaccuracies arising from patient motion or setup errors are identified and corrected. With frameless immobilization, image-guidance has a greater impact, especially in non-coplanar settings that can lead to patient motion and discrepancies between couch and radiation isocenters.</p><p><strong>Purpose: </strong>Both patient and phantom studies were conducted to assess and quantify the magnitude of geometric uncertainties after couch rotations, aiming at evaluating the clinical need for their correction to warrant a precise treatment delivery.</p><p><strong>Methods: </strong>Intrafraction X-ray data, performed by ExacTrac Dynamic (ETD) to monitor and correct patients' position throughout treatment delivery, were collected from 50 patients treated for brain metastases in stereotactic non-coplanar schemes and immobilized by stereotactic double-layered thermoplastic mask systems: 26 patients treated in 40 single-fraction SRS (168 stereoscopic X-ray images); 24 treated with FSRT in 128 fractions (278 stereoscopic X-ray images). Additionally, a head phantom was utilized and 350 measurements under two different couch loads were carried out to distinguish true patient motion from deviations caused by couch rotations or system-related effects. For both studies, ETD stereoscopic X-rays were acquired after each couch rotation and the first measured positioning deviation was calculated by comparing X-ray images to the treatment plan's digitally reconstructed radiographs.</p><p><strong>Results: </strong>Clinically relevant deviations were observed, exceeding clinical tolerance (≥ 0.5 mm/0.5°) mostly in the lateral and yaw directions and requiring repositioning in nearly half of the couch rotations. These deviations measuring up to 2 mm, revealed to be emerging mainly from patient motion rather than linac setup, as the phantom study showed maximum deviations of up to 0.6 mm and 0.4° when simulating a patient treatment and an interquartile range that did not exceed 0.2 mm and 0.2°.</p><p><strong>Conclusions: </strong>These findings demonstrate the importance of a continuous intrafraction motion monitoring and repositioning in cranial stereotactic treatments, especially in non-coplanar settings.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70505"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12967575/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147377552","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Post hoc estimation of a quantitative restriction spectrum imaging biomarker for prostate cancer detection using conventional MRI. 使用常规MRI检测前列腺癌的定量限制光谱成像生物标志物的事后估计。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70543
Deondre D Do, Christopher C Conlin, Aditya Bagrodia, Matthew Cooperberg, Michael E Hahn, Mukesh Harisinghani, Gary Hollenberg, Juan Javier-Desloges, Sophia C Kamran, Christopher J Kane, Kang-Lung Lee, Michael A Liss, Daniel J A Margolis, Paul M Murphy, Nabih Nakrour, Michael A Ohliger, Thomas Osinski, Rebecca Rakow-Penner, Mariluz Rojo Domingo, Amirali Salmasi, Ahmed S Shabaik, Yuze Song, Shaun Trecarten, Natasha Wehrli, Eric P Weinberg, Sean Woolen, Anders M Dale, Tyler M Seibert

Background: Multiparametric MRI is useful for early detection of clinically significant prostate cancer (csPCa), but its standard apparent diffusion coefficient (ADC) has limited utility as a quantitative metric for automated, patient-level detection of csPCa. Restriction spectrum imaging (RSI), an advanced diffusion technique, yields a quantitative biomarker (RSIrs) that improves csPCa detection. RSIrs is typically calculated from a dedicated multi-b-value acquisition. RSIrs estimated from conventional MRI has not been studied.

Purpose: To evaluate the accuracy and validity of RSI metrics estimated post hoc from conventional diffusion-weighted imaging (DWI) to serve as a viable surrogate for a dedicated RSI acquisition.

Materials and methods: We conducted a retrospective, multicenter study of patients with both a dedicated RSI acquisition and conventional DWI. We compared three different RSI restriction score (RSIrs) calculation methods: from the dedicated acquisition (RSIrsdedicated), from conventional DWI alone (RSIrspost-hoc), and from a combination of conventional DWI with only the high b-values from the RSI acquisition (RSIrscombo). We compared these methods for quantitative agreement and csPCa detection performance (area under the receiver operating characteristic [AUC, 95% confidence interval]) of maximum RSIrs (RSIrsmax) in the prostate compared to that of minimum ADC (ADC).

Results: Data from n = 1095 patients (16 centers) were analyzed. Post hoc RSIrsmax differed systematically from RSIrsdedicated by a median of +156 (RSIrspost-hoc) and -59 (RSIrscombo), respectively. AUCs for csPCa detection were 0.51 [0.47,0.54], 0.60 [0.57,0.64], 0.70 [0.67,0.74], and 0.77 [0.74,0.80] for ADC, RSIrspost-hoc, RSIrscombo, and RSIrsdedicated, respectively.

Conclusion: Even when estimated using conventional DWI, RSIrs is a superior quantitative biomarker to ADC for automated, patient-level detection of csPCa. A dedicated RSI acquisition gives the best performance. A compromise would be to acquire high b-values (1500 and 2500 s/mm2) to complement low b-values (<1000 s/mm2) from conventional DWI.

背景:多参数MRI可用于临床显著性前列腺癌(csPCa)的早期检测,但其标准表观扩散系数(ADC)作为csPCa自动检测的定量指标的效用有限。限制光谱成像(RSI)是一种先进的扩散技术,可以产生定量的生物标志物(RSIrs),从而提高csPCa的检测。RSIrs通常是通过专用的多b值采集来计算的。传统MRI估计的RSIrs尚未得到研究。目的:评估常规弥散加权成像(DWI)事后估计的RSI指标的准确性和有效性,以作为专用RSI采集的可行替代方法。材料和方法:我们进行了一项回顾性的多中心研究,研究对象包括专用RSI采集和常规DWI。我们比较了三种不同的RSI限制评分(RSIrs)计算方法:专用采集(RSIrsdedicated),单独使用常规DWI (rsipost -hoc),以及仅使用常规DWI与RSI采集的高b值(RSIrscombo)的组合。我们比较了这些方法的定量一致性和前列腺中最大RSIrs (RSIrsmax)与最小ADC (ADC)的csPCa检测性能(接受者工作特征下面积[AUC, 95%置信区间])。结果:分析了来自16个中心n = 1095例患者的数据。Post -hoc RSIrsmax与RSIrsdedicated的中位数差异分别为+156 (rsipost -hoc)和-59 (RSIrscombo)。ADC、rsipost -hoc、RSIrscombo和RSIrsdedicated检测csPCa的auc分别为0.51[0.47,0.54]、0.60[0.57,0.64]、0.70[0.67,0.74]和0.77[0.74,0.80]。结论:即使使用传统的DWI进行估计,RSIrs也是一种优于ADC的csPCa自动、患者水平检测的定量生物标志物。专用RSI采集提供最佳性能。一种折衷方法是获取高b值(1500和2500 s/mm2),以补充传统DWI的低b值(2)。
{"title":"Post hoc estimation of a quantitative restriction spectrum imaging biomarker for prostate cancer detection using conventional MRI.","authors":"Deondre D Do, Christopher C Conlin, Aditya Bagrodia, Matthew Cooperberg, Michael E Hahn, Mukesh Harisinghani, Gary Hollenberg, Juan Javier-Desloges, Sophia C Kamran, Christopher J Kane, Kang-Lung Lee, Michael A Liss, Daniel J A Margolis, Paul M Murphy, Nabih Nakrour, Michael A Ohliger, Thomas Osinski, Rebecca Rakow-Penner, Mariluz Rojo Domingo, Amirali Salmasi, Ahmed S Shabaik, Yuze Song, Shaun Trecarten, Natasha Wehrli, Eric P Weinberg, Sean Woolen, Anders M Dale, Tyler M Seibert","doi":"10.1002/acm2.70543","DOIUrl":"https://doi.org/10.1002/acm2.70543","url":null,"abstract":"<p><strong>Background: </strong>Multiparametric MRI is useful for early detection of clinically significant prostate cancer (csPCa), but its standard apparent diffusion coefficient (ADC) has limited utility as a quantitative metric for automated, patient-level detection of csPCa. Restriction spectrum imaging (RSI), an advanced diffusion technique, yields a quantitative biomarker (RSIrs) that improves csPCa detection. RSIrs is typically calculated from a dedicated multi-b-value acquisition. RSIrs estimated from conventional MRI has not been studied.</p><p><strong>Purpose: </strong>To evaluate the accuracy and validity of RSI metrics estimated post hoc from conventional diffusion-weighted imaging (DWI) to serve as a viable surrogate for a dedicated RSI acquisition.</p><p><strong>Materials and methods: </strong>We conducted a retrospective, multicenter study of patients with both a dedicated RSI acquisition and conventional DWI. We compared three different RSI restriction score (RSIrs) calculation methods: from the dedicated acquisition (RSIrs<sub>dedicated</sub>), from conventional DWI alone (RSIrs<sub>post-hoc</sub>), and from a combination of conventional DWI with only the high b-values from the RSI acquisition (RSIrs<sub>combo</sub>). We compared these methods for quantitative agreement and csPCa detection performance (area under the receiver operating characteristic [AUC, 95% confidence interval]) of maximum RSIrs (RSIrs<sub>max</sub>) in the prostate compared to that of minimum ADC (ADC).</p><p><strong>Results: </strong>Data from n = 1095 patients (16 centers) were analyzed. Post hoc RSIrs<sub>max</sub> differed systematically from RSIrs<sub>dedicated</sub> by a median of +156 (RSIrs<sub>post-hoc</sub>) and -59 (RSIrs<sub>combo</sub>), respectively. AUCs for csPCa detection were 0.51 [0.47,0.54], 0.60 [0.57,0.64], 0.70 [0.67,0.74], and 0.77 [0.74,0.80] for ADC, RSIrs<sub>post-hoc</sub>, RSIrs<sub>combo</sub>, and RSIrs<sub>dedicated</sub>, respectively.</p><p><strong>Conclusion: </strong>Even when estimated using conventional DWI, RSIrs is a superior quantitative biomarker to ADC for automated, patient-level detection of csPCa. A dedicated RSI acquisition gives the best performance. A compromise would be to acquire high b-values (1500 and 2500 s/mm<sup>2</sup>) to complement low b-values (<1000 s/mm<sup>2</sup>) from conventional DWI.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70543"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147486030","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
An automated method to establish patient specific asymmetric intrafraction motion monitoring tolerances for prostate radiotherapy. 建立前列腺放疗患者特异性非对称屈光度内运动监测容限的自动化方法。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70507
Bradley Beeksma, John Daniel, Erin Seymour, Andrew Dipuglia

Purpose: Intrafraction motion management is recognized as a critical component of radiotherapy. While clinical trials often recommend its use, they rarely define motion management tolerances. Consequently, institutions set tolerance levels independently, often without considering patient specific anatomy, dose distributions or direction of motion. This study introduces an automated method that can be used to derive patient specific, asymmetric intrafraction motion monitoring tolerances based on individualized treatment plans.

Methods: Treatment plans for 20 prostate cancer patients receiving a simultaneous integrated boost to dominant intraprostatic lesions (DIL) were retrospectively analyzed. Referred to as the isocentre translation method, patient movement was simulated by recalculating plans with the isocentre shifted by 2, 4 and 5 mm in six different directions. The magnitude and direction that resulted in violations of dose constraints for organ at risk (OAR) rectum, bladder and urethra was determined assuming a systematic shift for all fractions. A second automated approach, the contour translation method, used Eclipse Scripting API to estimate directional tolerances via contour translation and Boolean operations, avoiding dose recalculation. Results for this method were validated against the isocentre translation method and compared for efficiency by assessing processing time.

Results: The magnitude and direction of motion to cause OAR constraint violations were organ and patient specific. Urethral violations were most sensitive to shifts toward the DIL, whereas bladder and rectum constraints were primarily affected by anterior/superior and posterior shifts, respectively. The contour translation method was demonstrated to be equivalent to the isocentre translation method within an equivalence margin of 0.5 mm. The contour translation method significantly reduced processing time (4 min 40s vs. 47 min 52s per patient).

Conclusion: The direction and extent of motion impacting OAR constraints vary by patient and organ, supporting the need for personalized intrafraction motion monitoring tolerances. The proposed contour translation method provides a practical, efficient process that is able to facilitate individualized motion management in clinical workflows.

目的:屈光内运动管理是放疗的重要组成部分。虽然临床试验经常推荐使用它,但它们很少定义运动管理耐受性。因此,机构通常不考虑患者的具体解剖结构、剂量分布或运动方向,而独立设定耐受水平。本研究介绍了一种自动化方法,可用于根据个性化治疗计划获得患者特异性、不对称的屈光内运动监测耐受性。方法:回顾性分析20例前列腺癌患者同时接受显性前列腺内病变(DIL)综合强化治疗的治疗方案。被称为等心平移法,通过重新计算等心在六个不同方向上移动2,4和5mm的计划来模拟患者的运动。在假设所有部分都有系统的转移的情况下,确定了导致危险器官(OAR)直肠、膀胱和尿道违反剂量限制的幅度和方向。第二种自动化方法是等高线平移法,它使用Eclipse Scripting API通过等高线平移和布尔运算来估计方向公差,避免了剂量的重新计算。该方法的结果与等中心翻译方法进行了验证,并通过评估处理时间来比较效率。结果:引起OAR约束违反的运动的大小和方向是器官和患者特异性的。尿道侵犯对DIL移位最为敏感,而膀胱和直肠约束主要分别受到前/上移位和后移位的影响。在0.5 mm的等效余量范围内,证明了等高线平移法与等心平移法是等效的。轮廓平移法显著缩短了处理时间(每位患者4 min 40s vs. 47 min 52s)。结论:影响桨叶约束的运动方向和程度因患者和器官而异,支持个性化屈光内运动监测耐受性的需求。所提出的轮廓平移方法提供了一种实用、高效的过程,能够促进临床工作流程中的个性化运动管理。
{"title":"An automated method to establish patient specific asymmetric intrafraction motion monitoring tolerances for prostate radiotherapy.","authors":"Bradley Beeksma, John Daniel, Erin Seymour, Andrew Dipuglia","doi":"10.1002/acm2.70507","DOIUrl":"10.1002/acm2.70507","url":null,"abstract":"<p><strong>Purpose: </strong>Intrafraction motion management is recognized as a critical component of radiotherapy. While clinical trials often recommend its use, they rarely define motion management tolerances. Consequently, institutions set tolerance levels independently, often without considering patient specific anatomy, dose distributions or direction of motion. This study introduces an automated method that can be used to derive patient specific, asymmetric intrafraction motion monitoring tolerances based on individualized treatment plans.</p><p><strong>Methods: </strong>Treatment plans for 20 prostate cancer patients receiving a simultaneous integrated boost to dominant intraprostatic lesions (DIL) were retrospectively analyzed. Referred to as the isocentre translation method, patient movement was simulated by recalculating plans with the isocentre shifted by 2, 4 and 5 mm in six different directions. The magnitude and direction that resulted in violations of dose constraints for organ at risk (OAR) rectum, bladder and urethra was determined assuming a systematic shift for all fractions. A second automated approach, the contour translation method, used Eclipse Scripting API to estimate directional tolerances via contour translation and Boolean operations, avoiding dose recalculation. Results for this method were validated against the isocentre translation method and compared for efficiency by assessing processing time.</p><p><strong>Results: </strong>The magnitude and direction of motion to cause OAR constraint violations were organ and patient specific. Urethral violations were most sensitive to shifts toward the DIL, whereas bladder and rectum constraints were primarily affected by anterior/superior and posterior shifts, respectively. The contour translation method was demonstrated to be equivalent to the isocentre translation method within an equivalence margin of 0.5 mm. The contour translation method significantly reduced processing time (4 min 40s vs. 47 min 52s per patient).</p><p><strong>Conclusion: </strong>The direction and extent of motion impacting OAR constraints vary by patient and organ, supporting the need for personalized intrafraction motion monitoring tolerances. The proposed contour translation method provides a practical, efficient process that is able to facilitate individualized motion management in clinical workflows.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70507"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12946815/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147306191","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Quantitative evaluation of a deep learning-based noise reduction algorithm in digital radiography using noise power spectrum analysis. 基于噪声功率谱分析的基于深度学习的数字放射成像降噪算法的定量评估。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70521
Sho Maruyama, Hiroki Saitou

Background: Recently, deep learning (DL)-based noise reduction (DLNR) has been introduced in clinically used digital radiography (DR) systems, reporting superior performance over conventional algorithms. However, DLNR algorithms often operate as "black boxes" with nonlinear behavior, making it essential to understand the impact of such processing on image quality under different imaging conditions.

Purpose: This study aimed to quantitatively evaluate the image quality of a commercial DLNR algorithm for DR referred to as intelligent noise reduction (INR). Specifically, we compared its noise reduction performance with that of a conventional rule-based algorithm (conventional noise reduction, Con-NR) using frequency-domain metrics with detailed noise power spectrum (NPS) analysis.

Methods: The NPS was used to assess the spatial-frequency-dependent behavior of both INR and Con-NR across varying dose levels and different objects. In this work, we introduced a supplementary metric-the NPS improvement factor (NPSIF)-to quantify noise suppression across frequency ranges and facilitate direct comparison between methods.

Results: The DL-based algorithm achieved substantial noise reduction at low-dose settings compared with the conventional method, although its advantages were less pronounced at higher dose levels. The NPSIF effectively captured frequency-specific differences, thereby offering insights into the strengths and limitations of each technique.

Conclusions: The dose-dependent performance of the DL-based algorithm suggests sensitivity to the characteristics of the training data used to develop the DL model. The findings demonstrate distinct differences in the noise suppression behavior between DL-based and conventional methods in DR and underscore the importance of detailed frequency-domain evaluation for understanding advanced image processing. Further research is warranted to integrate noise analysis with diagnostic performance metrics to comprehensively assess clinical utility.

背景:最近,基于深度学习(DL)的降噪(DLNR)已经被引入到临床使用的数字放射成像(DR)系统中,报告了比传统算法更好的性能。然而,DLNR算法通常作为具有非线性行为的“黑盒子”运行,因此有必要了解在不同成像条件下这种处理对图像质量的影响。目的:本研究旨在定量评估一种用于DR的商用DLNR算法的图像质量,即智能降噪(INR)。具体来说,我们使用频域指标和详细的噪声功率谱(NPS)分析,将其降噪性能与传统的基于规则的算法(常规降噪,Con-NR)进行了比较。方法:采用NPS评估不同剂量水平和不同对象下INR和Con-NR的空间频率依赖行为。在这项工作中,我们引入了一个补充度量- NPS改进因子(NPSIF)-来量化跨频率范围的噪声抑制,并促进方法之间的直接比较。结果:与传统方法相比,基于dl的算法在低剂量设置下取得了显著的降噪效果,尽管其优势在高剂量水平下不太明显。NPSIF有效地捕获了特定频率的差异,从而提供了对每种技术的优点和局限性的见解。结论:基于DL的算法的剂量依赖性能表明对用于开发DL模型的训练数据特征的敏感性。研究结果表明,基于dl的DR方法和传统DR方法在噪声抑制行为上存在明显差异,并强调了详细的频域评估对于理解高级图像处理的重要性。进一步的研究需要将噪声分析与诊断性能指标相结合,以全面评估临床效用。
{"title":"Quantitative evaluation of a deep learning-based noise reduction algorithm in digital radiography using noise power spectrum analysis.","authors":"Sho Maruyama, Hiroki Saitou","doi":"10.1002/acm2.70521","DOIUrl":"10.1002/acm2.70521","url":null,"abstract":"<p><strong>Background: </strong>Recently, deep learning (DL)-based noise reduction (DLNR) has been introduced in clinically used digital radiography (DR) systems, reporting superior performance over conventional algorithms. However, DLNR algorithms often operate as \"black boxes\" with nonlinear behavior, making it essential to understand the impact of such processing on image quality under different imaging conditions.</p><p><strong>Purpose: </strong>This study aimed to quantitatively evaluate the image quality of a commercial DLNR algorithm for DR referred to as intelligent noise reduction (INR). Specifically, we compared its noise reduction performance with that of a conventional rule-based algorithm (conventional noise reduction, Con-NR) using frequency-domain metrics with detailed noise power spectrum (NPS) analysis.</p><p><strong>Methods: </strong>The NPS was used to assess the spatial-frequency-dependent behavior of both INR and Con-NR across varying dose levels and different objects. In this work, we introduced a supplementary metric-the NPS improvement factor (NPSIF)-to quantify noise suppression across frequency ranges and facilitate direct comparison between methods.</p><p><strong>Results: </strong>The DL-based algorithm achieved substantial noise reduction at low-dose settings compared with the conventional method, although its advantages were less pronounced at higher dose levels. The NPSIF effectively captured frequency-specific differences, thereby offering insights into the strengths and limitations of each technique.</p><p><strong>Conclusions: </strong>The dose-dependent performance of the DL-based algorithm suggests sensitivity to the characteristics of the training data used to develop the DL model. The findings demonstrate distinct differences in the noise suppression behavior between DL-based and conventional methods in DR and underscore the importance of detailed frequency-domain evaluation for understanding advanced image processing. Further research is warranted to integrate noise analysis with diagnostic performance metrics to comprehensively assess clinical utility.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70521"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12935518/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147306228","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
From data to treatment plan: An AI-driven path for automated breast radiotherapy planning. 从数据到治疗方案:人工智能驱动的乳房放疗计划自动化路径。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70491
P Gallego, E Ambroa, J Pérez-Alija, J C Julià, N Jornet, A Matas, C Anson, A Mera, N Tejedor, H Vivancos, A Ruiz, M Barceló, A Dominguez, V Riu, J Roda, P Carrasco, S Balocco, O Díaz
<p><strong>Background: </strong>Breast cancer is one of the most prevalent malignancies in women, with radiotherapy (RT) playing a key role in its treatment. Advances in RT techniques, such as 3D conformal radiotherapy (3D-CRT) and intensity-modulated radiotherapy (IMRT), have improved dose precision and reduced side effects. However, RT modality selection and treatment planning remain manual, time-consuming, and subject to variability.</p><p><strong>Purpose: </strong>This study presents and validates TARS-B (Treatment Automation and Radiotherapy Selection for Breast Cancer), an automated framework that combines a deep learning-based decision-making module (DMF) for selecting the optimal RT technique and a fully automated treatment planning system (ATP) for generating deliverable plans that meet clinical quality standards and are deemed acceptable for clinical use.</p><p><strong>Materials and methods: </strong>TARS-B functions in two stages. First, the DMF analyzes individual patient data to determine whether 3D-CRT or IMRT is more appropriate. Second, the ATP generates the corresponding treatment plan. For 3D-CRT, a field-in-field (FiF) method is used to enhance dose homogeneity and minimize hotspots. For IMRT, the DMF provides neural network-based dose predictions, which are used to generate constraints for organs-at-risk (OARs). Both processes are fully scripted within the treatment planning system (TPS). The framework was tested on 60 breast cancer patients: 30 originally treated with 3D-CRT and 30 with IMRT. Two analyses were conducted. First, the ATP's performance was evaluated by comparing automated plans with their manually generated clinical counterparts for both techniques. Second, the full TARS-B pipeline was assessed by applying the DMF to select the RT modality and automatically generating the plan, comparing results to the original clinical plans. Dosimetric parameters, including planning target volume (PTV) coverage, OAR constraints, and low- and intermediate-dose bath, were analyzed. Planning times were also compared.</p><p><strong>Results: </strong>No statistically significant differences ( <math> <semantics><mrow><mi>p</mi> <mo>></mo> <mn>0.005</mn></mrow> <annotation>$p > 0.005$</annotation></semantics> </math> ) were found between manual and automated plans in key dosimetric metrics, including PTV coverage (V95 <math><semantics><mo>%</mo> <annotation>$%$</annotation></semantics> </math> ), hotspots (V105 <math><semantics><mo>%</mo> <annotation>$%$</annotation></semantics> </math> ), and OAR constraints, for both 3D-CRT and IMRT. TARS-B confirmed the appropriateness of 3D-CRT in all patients originally treated with it and recommended re-planning with 3D-CRT for 15 of 30 IMRT cases. Of these, 14 re-plans met all criteria; one failed due to anatomical anomalies. Re-planning led to a reduction in low-dose bath (up to 2800 <math> <semantics><msup><mi>cm</mi> <mn>3</mn></msup> <annotation>${rm cm}^3$</annotation></semantics> </math>
背景:乳腺癌是女性最常见的恶性肿瘤之一,放疗在其治疗中起着关键作用。放疗技术的进步,如三维适形放疗(3D- crt)和调强放疗(IMRT),提高了剂量精度,减少了副作用。然而,RT模式的选择和治疗计划仍然是手工的,耗时的,并受到可变性。目的:本研究提出并验证了ars -b(乳腺癌治疗自动化和放疗选择),这是一个自动化框架,结合了基于深度学习的决策模块(DMF),用于选择最佳的放疗技术,以及全自动治疗计划系统(ATP),用于生成符合临床质量标准的可交付计划,并被认为可用于临床使用。材料与方法:ars - b分两个阶段起作用。首先,DMF分析个体患者数据,以确定3D-CRT还是IMRT更合适。其次,ATP生成相应的治疗方案。对于3D-CRT,采用场内场(field-in-field, FiF)方法来增强剂量均匀性并最小化热点。对于IMRT, DMF提供了基于神经网络的剂量预测,用于生成危险器官(OARs)的限制。这两个过程都在治疗计划系统(TPS)中完全脚本化。该框架在60名乳腺癌患者身上进行了测试:30名最初接受3D-CRT治疗,30名接受IMRT治疗。进行了两项分析。首先,通过比较两种技术的自动计划和人工生成的临床计划来评估ATP的性能。其次,通过应用DMF选择RT方式并自动生成计划,将结果与原始临床计划进行比较,评估整个ars - b管道。剂量学参数,包括规划靶体积(PTV)覆盖率、OAR约束和低剂量和中剂量浴,进行了分析。计划时间也进行了比较。结果:在3D-CRT和IMRT的关键剂量学指标,包括PTV覆盖率(v95% $%$)、热点(v105% $%$)和OAR约束,手动计划和自动计划之间没有统计学上的显著差异(p > 0.005$)。ars - b证实了所有最初接受3D-CRT治疗的患者的适宜性,并建议30例IMRT病例中的15例重新计划3D-CRT。其中,14个重新规划符合所有标准;一次由于解剖异常而失败。重新规划导致低剂量浴(高达2800厘米3美元{rm厘米}^3美元)和中剂量浴(高达3000厘米3美元{rm厘米}^3美元)的减少。中剂量浴降低有统计学意义(p < 0.005$)。计划时间大幅减少:IMRT从157.4±116.2$ 157.4 pm 116.2$到2.0±1.3$ 2.0 pm 1.3$ min, 3D-CRT从112±70$ 112 pm 70$到5±4$ 5 pm 4$ min (p < 0.005$ p < 0.005$)。结论:ars - b有效地自动化了最合适的放疗技术的选择和高质量治疗方案的生成。该框架提高了工作流程效率,减少了计划时间,并保持了剂量学质量,突出了其在乳腺癌RT临床实施中的潜力。
{"title":"From data to treatment plan: An AI-driven path for automated breast radiotherapy planning.","authors":"P Gallego, E Ambroa, J Pérez-Alija, J C Julià, N Jornet, A Matas, C Anson, A Mera, N Tejedor, H Vivancos, A Ruiz, M Barceló, A Dominguez, V Riu, J Roda, P Carrasco, S Balocco, O Díaz","doi":"10.1002/acm2.70491","DOIUrl":"10.1002/acm2.70491","url":null,"abstract":"&lt;p&gt;&lt;strong&gt;Background: &lt;/strong&gt;Breast cancer is one of the most prevalent malignancies in women, with radiotherapy (RT) playing a key role in its treatment. Advances in RT techniques, such as 3D conformal radiotherapy (3D-CRT) and intensity-modulated radiotherapy (IMRT), have improved dose precision and reduced side effects. However, RT modality selection and treatment planning remain manual, time-consuming, and subject to variability.&lt;/p&gt;&lt;p&gt;&lt;strong&gt;Purpose: &lt;/strong&gt;This study presents and validates TARS-B (Treatment Automation and Radiotherapy Selection for Breast Cancer), an automated framework that combines a deep learning-based decision-making module (DMF) for selecting the optimal RT technique and a fully automated treatment planning system (ATP) for generating deliverable plans that meet clinical quality standards and are deemed acceptable for clinical use.&lt;/p&gt;&lt;p&gt;&lt;strong&gt;Materials and methods: &lt;/strong&gt;TARS-B functions in two stages. First, the DMF analyzes individual patient data to determine whether 3D-CRT or IMRT is more appropriate. Second, the ATP generates the corresponding treatment plan. For 3D-CRT, a field-in-field (FiF) method is used to enhance dose homogeneity and minimize hotspots. For IMRT, the DMF provides neural network-based dose predictions, which are used to generate constraints for organs-at-risk (OARs). Both processes are fully scripted within the treatment planning system (TPS). The framework was tested on 60 breast cancer patients: 30 originally treated with 3D-CRT and 30 with IMRT. Two analyses were conducted. First, the ATP's performance was evaluated by comparing automated plans with their manually generated clinical counterparts for both techniques. Second, the full TARS-B pipeline was assessed by applying the DMF to select the RT modality and automatically generating the plan, comparing results to the original clinical plans. Dosimetric parameters, including planning target volume (PTV) coverage, OAR constraints, and low- and intermediate-dose bath, were analyzed. Planning times were also compared.&lt;/p&gt;&lt;p&gt;&lt;strong&gt;Results: &lt;/strong&gt;No statistically significant differences ( &lt;math&gt; &lt;semantics&gt;&lt;mrow&gt;&lt;mi&gt;p&lt;/mi&gt; &lt;mo&gt;&gt;&lt;/mo&gt; &lt;mn&gt;0.005&lt;/mn&gt;&lt;/mrow&gt; &lt;annotation&gt;$p &gt; 0.005$&lt;/annotation&gt;&lt;/semantics&gt; &lt;/math&gt; ) were found between manual and automated plans in key dosimetric metrics, including PTV coverage (V95 &lt;math&gt;&lt;semantics&gt;&lt;mo&gt;%&lt;/mo&gt; &lt;annotation&gt;$%$&lt;/annotation&gt;&lt;/semantics&gt; &lt;/math&gt; ), hotspots (V105 &lt;math&gt;&lt;semantics&gt;&lt;mo&gt;%&lt;/mo&gt; &lt;annotation&gt;$%$&lt;/annotation&gt;&lt;/semantics&gt; &lt;/math&gt; ), and OAR constraints, for both 3D-CRT and IMRT. TARS-B confirmed the appropriateness of 3D-CRT in all patients originally treated with it and recommended re-planning with 3D-CRT for 15 of 30 IMRT cases. Of these, 14 re-plans met all criteria; one failed due to anatomical anomalies. Re-planning led to a reduction in low-dose bath (up to 2800 &lt;math&gt; &lt;semantics&gt;&lt;msup&gt;&lt;mi&gt;cm&lt;/mi&gt; &lt;mn&gt;3&lt;/mn&gt;&lt;/msup&gt; &lt;annotation&gt;${rm cm}^3$&lt;/annotation&gt;&lt;/semantics&gt; &lt;/math&gt;","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70491"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12951544/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147326015","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Automatic target volume segmentation for offline adaptive head-and-neck radiotherapy. 离线自适应头颈部放疗的自动靶体积分割。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70479
Gilles Moliner, Maxime Michaud, Antoine Guerin, Rodolfe Verstraet, Charles Debrigode, Philippe Lang, Joel Greffier

Purpose: To assess the clinical utility of Smartfuse (Therapanacea, France), a deformable image registration (DIR) algorithm for automatic propagation of target volumes in the context of offline adaptive head-and-neck radiotherapy.

Materials and methods: Ten patients underwent offline re-planning during head-and-neck radiotherapy. Target volumes (GTV and CTV) were manually delineated by radiation oncologists (ROs) on both the initial CT (CTi) and one re-planning CT (CTR). These manual contours were compared to those propagated by Smartfuse from CTi to CTR. The geometric agreement between DIR-propagated and RO-delineated contours was assessed using Dice similarity coefficient (DSC), 95th percentile Hausdorff distance (HD95), and surface Dice similarity coefficient (sDSC) with 0 and 2 mm thresholds. Dosimetric evaluation was conducted by comparing dose distributions from generated plans using automatically propagated target volumes (PTVDIR) with reference plans based on RO-delineated targets (PTVRO). Coverage of RO-delineated targets (GTVRO + CTVRO and PTVRO) was assessed using D95%, D50%, Dmax, and V95% ≥ 95%. Spatial dose differences were analyzed using dose difference (DD) metrics at 5% and 2% thresholds.

Results: Median DSC, HD95, sDSC0 mm and sDSC2 mm were 0.86, 4.0 mm, 0.29 and 0.73, respectively. For D95%, median relative differences between DIR and RO plans were -0.6% for GTVRO + CTVRO and -2.1% for PTVRO for D95%. All GTVRO + CTVRO reached V95% ≥ 95% with DIR plans, but only 61% of PTVRO did. Spatial DD analysis showed median pass rates of 99.2% (DD5%) and 74.5% (DD2%) for GTVRO + CTVRO, and 85.5% (DD5%) and 54.9% (DD2%) for PTVRO.

Conclusion: Smartfuse may facilitate efficient propagation of target volumes in this study. However, medical review of auto-propagated volumes remains essential, as dosimetric discrepancies may arise when relying solely on automatically generated PTV.

目的:评估Smartfuse (Therapanacea, France)的临床应用,Smartfuse是一种可变形图像配准(DIR)算法,用于离线自适应头颈部放疗背景下目标体积的自动传播。材料与方法:10例患者在头颈部放疗期间进行线下再规划。靶体积(GTV和CTV)由放射肿瘤学家(ROs)在初始CT (CTi)和一台重新规划CT (CTR)上手动划定。将这些手工轮廓与Smartfuse从CTi到CTR传播的轮廓进行比较。使用Dice相似系数(DSC)、第95百分位Hausdorff距离(HD95)和表面Dice相似系数(sDSC)评估dir传播和ro划定轮廓之间的几何一致性,阈值分别为0和2 mm。通过比较使用自动传播靶体积(PTVDIR)的生成方案与基于ro划定靶体积(PTVRO)的参考方案的剂量分布,进行剂量学评价。采用D95%、D50%、Dmax和V95%≥95%评估ro划定靶点(GTVRO + CTVRO和PTVRO)的覆盖率。使用剂量差(DD)指标在5%和2%阈值下分析空间剂量差异。结果:中位DSC、HD95、sDSC0 mm和sDSC2 mm分别为0.86、4.0 mm、0.29和0.73。对于D95%, GTVRO + CTVRO计划的DIR和RO计划的中位相对差异为-0.6%,PTVRO计划的中位相对差异为-2.1%。所有GTVRO + CTVRO均达到V95%≥95%,但只有61%的PTVRO达到V95%。空间DD分析显示GTVRO + CTVRO的中位通过率分别为99.2% (DD5%)和74.5% (DD2%), PTVRO的中位通过率分别为85.5% (DD5%)和54.9% (DD2%)。结论:在本研究中,Smartfuse可以促进靶细胞的高效繁殖。然而,对自动传播的体积进行医学检查仍然是必要的,因为仅依靠自动生成的PTV可能会产生剂量学差异。
{"title":"Automatic target volume segmentation for offline adaptive head-and-neck radiotherapy.","authors":"Gilles Moliner, Maxime Michaud, Antoine Guerin, Rodolfe Verstraet, Charles Debrigode, Philippe Lang, Joel Greffier","doi":"10.1002/acm2.70479","DOIUrl":"https://doi.org/10.1002/acm2.70479","url":null,"abstract":"<p><strong>Purpose: </strong>To assess the clinical utility of Smartfuse (Therapanacea, France), a deformable image registration (DIR) algorithm for automatic propagation of target volumes in the context of offline adaptive head-and-neck radiotherapy.</p><p><strong>Materials and methods: </strong>Ten patients underwent offline re-planning during head-and-neck radiotherapy. Target volumes (GTV and CTV) were manually delineated by radiation oncologists (ROs) on both the initial CT (CT<sub>i</sub>) and one re-planning CT (CT<sub>R</sub>). These manual contours were compared to those propagated by Smartfuse from CT<sub>i</sub> to CT<sub>R</sub>. The geometric agreement between DIR-propagated and RO-delineated contours was assessed using Dice similarity coefficient (DSC), 95th percentile Hausdorff distance (HD<sub>95</sub>), and surface Dice similarity coefficient (sDSC) with 0 and 2 mm thresholds. Dosimetric evaluation was conducted by comparing dose distributions from generated plans using automatically propagated target volumes (PTV<sub>DIR</sub>) with reference plans based on RO-delineated targets (PTV<sub>RO</sub>). Coverage of RO-delineated targets (GTV<sub>RO</sub> + CTV<sub>RO</sub> and PTV<sub>RO</sub>) was assessed using D95%, D50%, Dmax, and V95% ≥ 95%. Spatial dose differences were analyzed using dose difference (DD) metrics at 5% and 2% thresholds.</p><p><strong>Results: </strong>Median DSC, HD<sub>95</sub>, sDSC<sub>0 mm</sub> and sDSC<sub>2 mm</sub> were 0.86, 4.0 mm, 0.29 and 0.73, respectively. For D<sub>95%</sub>, median relative differences between DIR and RO plans were -0.6% for GTV<sub>RO</sub> + CTV<sub>RO</sub> and -2.1% for PTV<sub>RO</sub> for D<sub>95%</sub>. All GTV<sub>RO</sub> + CTV<sub>RO</sub> reached V95% ≥ 95% with DIR plans, but only 61% of PTV<sub>RO</sub> did. Spatial DD analysis showed median pass rates of 99.2% (DD<sub>5%</sub>) and 74.5% (DD<sub>2%</sub>) for GTV<sub>RO</sub> + CTV<sub>RO</sub>, and 85.5% (DD<sub>5%</sub>) and 54.9% (DD<sub>2%</sub>) for PTV<sub>RO</sub>.</p><p><strong>Conclusion: </strong>Smartfuse may facilitate efficient propagation of target volumes in this study. However, medical review of auto-propagated volumes remains essential, as dosimetric discrepancies may arise when relying solely on automatically generated PTV.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70479"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147443673","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Total body irradiation (TBI) combining volumetric modulated arc therapy (VMAT) for thorax at standard source axis distance (SAD) with intensity modulated radiation therapy (IMRT) at extended SAD. 在标准源轴距离(SAD)下,胸腔的全身照射(TBI)与扩展源轴距离(SAD)下的强度调制放射治疗(IMRT)相结合。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70529
Marc Delaperrière, Markus Hirt, Christian Felix Schulz, Frank André Siebert

Purpose: Total body irradiation (TBI) is an important part of conditioning regimens prior to hematopoietic stem cell transplantation (HSCT). At our institution, a conventional TBI technique employing bilateral fields at extended source-to-axis distance (eSAD) and a solid three-dimensional (3D) compensator has been standard practice since the 80s. However, its ability to minimize lung dose remains limited.

Aims: This study aims to modernize the existing TBI technique using the capabilities of our treatment planning system (TPS). The proposed hybrid method replaces solid compensators with virtual compensation by combining bilateral intensity-modulated radiotherapy (IMRT) fields at eSAD of 500 cm with volumetric modulated arc therapy (VMAT) fields at standard SAD 100 cm for the thoracic region.

Methods: At first, the Anisotropic Analytical Algorithm (AAA) V16 and AcurosXB (AXB) V16 dose calculation algorithms of the Eclipse TPS (Varian) were commissioned for use at eSAD. Dosimetric measurements were performed in water, solid water, and inhomogeneous phantoms. Subsequently, a treatment planning strategy was developed to minimize the influence of positioning uncertainties on lung dose. Anonymized computed tomography (CT) datasets representing diverse patient anatomies were utilized for treatment planning evaluation. Final validation was conducted through an end-to-end test using an anthropomorphic phantom. SAD VMAT and eSAD IMRT fields were verified using a two-dimensional (2D) dosimetry chamber array positioned at both caudal and cranial thoracic levels under SAD and eSAD setups. The resulting cumulative dose was then compared with the AXB-calculated dose.

Results: At eSAD, mean measured-calculated dose deviations were < 2% for AXB and < 4% for AAA, with confidence limits of 3.5% and 5%, respectively. These results affirmed the clinical viability of both algorithms, with AXB providing superior dose calculation accuracy. Planning studies showed consistent PTV coverage with lung dose reduction across diverse anatomies, and end-to-end phantom validation confirmed the workflow practicability. Verification of summed dose using the 2D array, with 95% of evaluated points meeting a gamma criterion of 5 mm and 5%, provided additional confidence in the dosimetric robustness of the treatment concept.

Conclusion: This study showed the feasibility of a novel hybrid eSAD IMRT/VMAT TBI technique providing a clinically viable approach enabling effective lung dose sparing while maintaining the robustness of large-field irradiation.

目的:全身照射(TBI)是造血干细胞移植(HSCT)前调理方案的重要组成部分。在我们的机构,自80年代以来,采用扩展源到轴距离(eSAD)和立体三维(3D)补偿器的双侧电场的传统TBI技术已成为标准做法。然而,其减少肺剂量的能力仍然有限。目的:本研究旨在利用我们的治疗计划系统(TPS)的能力使现有的创伤性脑损伤技术现代化。提出的混合方法通过将双侧500 cm的eSAD调强放疗(IMRT)场与胸椎区域100 cm的标准SAD体积调制电弧治疗(VMAT)场相结合,以虚拟补偿取代实体补偿器。方法:首先将Eclipse TPS (Varian)的各向异性分析算法(AAA) V16和acrosxb (AXB) V16剂量计算算法委托给eSAD使用。在水中、固体水中和非均质幻影中进行剂量测定。随后,制定了治疗计划策略,以尽量减少定位不确定性对肺剂量的影响。代表不同患者解剖结构的匿名计算机断层扫描(CT)数据集被用于治疗计划评估。最后的验证是通过使用拟人化的幻影进行端到端测试。在SAD和eSAD设置下,使用位于尾侧和颅胸水平的二维(2D)剂量室阵列验证SAD VMAT和eSAD IMRT场。然后将所得累积剂量与axb计算的剂量进行比较。结论:本研究显示了一种新型混合eSAD IMRT/VMAT TBI技术的可行性,提供了一种临床可行的方法,可以在保持大视场照射稳健性的同时有效地节省肺部剂量。
{"title":"Total body irradiation (TBI) combining volumetric modulated arc therapy (VMAT) for thorax at standard source axis distance (SAD) with intensity modulated radiation therapy (IMRT) at extended SAD.","authors":"Marc Delaperrière, Markus Hirt, Christian Felix Schulz, Frank André Siebert","doi":"10.1002/acm2.70529","DOIUrl":"10.1002/acm2.70529","url":null,"abstract":"<p><strong>Purpose: </strong>Total body irradiation (TBI) is an important part of conditioning regimens prior to hematopoietic stem cell transplantation (HSCT). At our institution, a conventional TBI technique employing bilateral fields at extended source-to-axis distance (eSAD) and a solid three-dimensional (3D) compensator has been standard practice since the 80s. However, its ability to minimize lung dose remains limited.</p><p><strong>Aims: </strong>This study aims to modernize the existing TBI technique using the capabilities of our treatment planning system (TPS). The proposed hybrid method replaces solid compensators with virtual compensation by combining bilateral intensity-modulated radiotherapy (IMRT) fields at eSAD of 500 cm with volumetric modulated arc therapy (VMAT) fields at standard SAD 100 cm for the thoracic region.</p><p><strong>Methods: </strong>At first, the Anisotropic Analytical Algorithm (AAA) V16 and AcurosXB (AXB) V16 dose calculation algorithms of the Eclipse TPS (Varian) were commissioned for use at eSAD. Dosimetric measurements were performed in water, solid water, and inhomogeneous phantoms. Subsequently, a treatment planning strategy was developed to minimize the influence of positioning uncertainties on lung dose. Anonymized computed tomography (CT) datasets representing diverse patient anatomies were utilized for treatment planning evaluation. Final validation was conducted through an end-to-end test using an anthropomorphic phantom. SAD VMAT and eSAD IMRT fields were verified using a two-dimensional (2D) dosimetry chamber array positioned at both caudal and cranial thoracic levels under SAD and eSAD setups. The resulting cumulative dose was then compared with the AXB-calculated dose.</p><p><strong>Results: </strong>At eSAD, mean measured-calculated dose deviations were < 2% for AXB and < 4% for AAA, with confidence limits of 3.5% and 5%, respectively. These results affirmed the clinical viability of both algorithms, with AXB providing superior dose calculation accuracy. Planning studies showed consistent PTV coverage with lung dose reduction across diverse anatomies, and end-to-end phantom validation confirmed the workflow practicability. Verification of summed dose using the 2D array, with 95% of evaluated points meeting a gamma criterion of 5 mm and 5%, provided additional confidence in the dosimetric robustness of the treatment concept.</p><p><strong>Conclusion: </strong>This study showed the feasibility of a novel hybrid eSAD IMRT/VMAT TBI technique providing a clinically viable approach enabling effective lung dose sparing while maintaining the robustness of large-field irradiation.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70529"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147480677","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
An X-ray and γ-ray combination strategy to exploit steep dose gradients for improved prostate cancer radiotherapy: A dosimetric comparison of Taichi Pro and Halcyon. 利用陡坡剂量梯度的x射线和γ射线联合策略改善前列腺癌放疗:太极Pro和Halcyon的剂量学比较。
IF 2.2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2026-03-01 DOI: 10.1002/acm2.70532
Kuo Li, Yong Yin, Ting Zhu, Weipeng Sun, Shihao Wang, Zhenjiang Li

Purpose: This study aimed to evaluate a novel technological platform, Taichi Pro-which integrates a 6 MV flattening filterfree linear accelerator with a 18 source Rotating Gamma System (RGS) to generate steep dose gradients via multisource focused γ-rays and noncoplanar arcs-for precision radiotherapy in prostate cancer. The work provides evidence to support the clinical adoption of hybridmodality radiotherapy devices.

Methods: Fifteen prostate cancer patients were enrolled. For each patient, a dual-modality Taichi Pro plan (RGS focused on the planning gross tumor volume (PGTV) + Linac covering the planning target volume (PTV)) and a Halcyon photon plan (three-arc VMAT) were designed while maintaining clinically tolerable dose to organs at risk (OAR). Comparative assessments included planning target volume (PTV and PGTV) metrics (D95%, Dmean, homogeneity index HI, conformity index CI), OAR doses (rectum V40/V60/D2cc, bladder V40, testis D2cc, etc.) and delivery efficiency to evaluate the ability to escalate target dose while sparing adjacent OARs.

Results: All plans met institutional clinical constraints. Taichi Pro significantly increased PGTV Dmean (79.48 Gy ± 1.75 Gy) compared to Halcyon (73.25 Gy ± 0.55 24 Gy, P < 0.001) and Dmax (122.74 Gy ± 8.69 Gy) compared to Halcyon (76.15 Gy ± 0.79 Gy, P < 0.001), albeit poorer homogeneity (HI: 0.50 ± 0.09 for Taichi Pro vs. 0.06 ± 0.01 for Halcyon, though within clinically acceptable limits). Taichi Pro significantly reduced rectum V60 (3.97% ± 3.25% vs. Halcyon 7.46% ± 4.78%, P = 0.016), and D2cc (61.61 Gy ± 5.01 Gy vs. Halcyon 65.29 Gy ± 4.52 Gy, P = 0.040). Taichi Pro also significantly reduced testis D2cc (2.39 ± 1.99 Gy) compared to (3.17 Gy ± 1.40 Gy, P = 0.006). Halcyon demonstrated significantly shorter beam-on time (1.81 ± 0.23 minutes vs. 5.05 ± 1.59 minutes for Taichi Pro, P < 0.001).

Conclusion: Utilizing the steep dose gradient characteristic of the RGS, the Taichi Pro dual-modality system effectively achieved target dose escalation while simultaneously improving sparing of adjacent OARs. This approach holds the potential for enhancing patient treatment outcomes and quality of life.

目的:本研究旨在评估一种新的技术平台太极pro,该平台集成了一个6 MV无滤波平坦线性加速器和一个18源旋转伽马系统(RGS),通过多源聚焦γ射线和非共面弧产生陡峭的剂量梯度,用于前列腺癌的精确放疗。这项工作为临床采用混合模式放射治疗装置提供了证据。方法:纳入15例前列腺癌患者。对于每位患者,设计了双模态太极Pro计划(RGS专注于计划总肿瘤体积(PGTV) + Linac覆盖计划靶体积(PTV))和Halcyon光子计划(三弧VMAT),同时保持对危险器官(OAR)的临床耐受剂量。比较评估包括计划靶体积(PTV和PGTV)指标(D95%、Dmean、均匀性指数HI、符合性指数CI)、OAR剂量(直肠V40/V60/D2cc、膀胱V40、睾丸D2cc等)和递送效率,以评估在不影响邻近OARs的情况下提高靶剂量的能力。结果:所有方案均满足机构临床约束。与Halcyon相比,太极拳Pro显著提高了PGTV Dmean (79.48 Gy±1.75 Gy) (73.25 Gy±0.55 24 Gy, pmax (122.74 Gy±8.69 Gy) (76.15 Gy±0.79 Gy, p60(3.97%±3.25%比Halcyon 7.46%±4.78%,P = 0.016)和D2cc (61.61 Gy±5.01 Gy比Halcyon 65.29 Gy±4.52 Gy, P = 0.040)。太极Pro也显著降低睾丸D2cc(2.39±1.99 Gy)与(3.17 Gy±1.40 Gy, P = 0.006)。Halcyon的射束时间明显缩短(1.81±0.23分钟),太极Pro(5.05±1.59分钟)。结论:利用RGS的陡剂量梯度特性,太极Pro双模态系统有效地实现了目标剂量递增,同时提高了相邻桨的保护。这种方法具有提高患者治疗结果和生活质量的潜力。
{"title":"An X-ray and γ-ray combination strategy to exploit steep dose gradients for improved prostate cancer radiotherapy: A dosimetric comparison of Taichi Pro and Halcyon.","authors":"Kuo Li, Yong Yin, Ting Zhu, Weipeng Sun, Shihao Wang, Zhenjiang Li","doi":"10.1002/acm2.70532","DOIUrl":"10.1002/acm2.70532","url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to evaluate a novel technological platform, Taichi Pro-which integrates a 6 MV flattening filterfree linear accelerator with a 18 source Rotating Gamma System (RGS) to generate steep dose gradients via multisource focused γ-rays and noncoplanar arcs-for precision radiotherapy in prostate cancer. The work provides evidence to support the clinical adoption of hybridmodality radiotherapy devices.</p><p><strong>Methods: </strong>Fifteen prostate cancer patients were enrolled. For each patient, a dual-modality Taichi Pro plan (RGS focused on the planning gross tumor volume (PGTV) + Linac covering the planning target volume (PTV)) and a Halcyon photon plan (three-arc VMAT) were designed while maintaining clinically tolerable dose to organs at risk (OAR). Comparative assessments included planning target volume (PTV and PGTV) metrics (D<sub>95%</sub>, D<sub>mean</sub>, homogeneity index HI, conformity index CI), OAR doses (rectum V<sub>40</sub>/V<sub>60</sub>/D<sub>2cc</sub>, bladder V<sub>40</sub>, testis D<sub>2cc</sub>, etc.) and delivery efficiency to evaluate the ability to escalate target dose while sparing adjacent OARs.</p><p><strong>Results: </strong>All plans met institutional clinical constraints. Taichi Pro significantly increased PGTV D<sub>mean</sub> (79.48 Gy ± 1.75 Gy) compared to Halcyon (73.25 Gy ± 0.55 24 Gy, P < 0.001) and D<sub>max</sub> (122.74 Gy ± 8.69 Gy) compared to Halcyon (76.15 Gy ± 0.79 Gy, P < 0.001), albeit poorer homogeneity (HI: 0.50 ± 0.09 for Taichi Pro vs. 0.06 ± 0.01 for Halcyon, though within clinically acceptable limits). Taichi Pro significantly reduced rectum V<sub>60</sub> (3.97% ± 3.25% vs. Halcyon 7.46% ± 4.78%, P = 0.016), and D<sub>2cc</sub> (61.61 Gy ± 5.01 Gy vs. Halcyon 65.29 Gy ± 4.52 Gy, P = 0.040). Taichi Pro also significantly reduced testis D<sub>2cc</sub> (2.39 ± 1.99 Gy) compared to (3.17 Gy ± 1.40 Gy, P = 0.006). Halcyon demonstrated significantly shorter beam-on time (1.81 ± 0.23 minutes vs. 5.05 ± 1.59 minutes for Taichi Pro, P < 0.001).</p><p><strong>Conclusion: </strong>Utilizing the steep dose gradient characteristic of the RGS, the Taichi Pro dual-modality system effectively achieved target dose escalation while simultaneously improving sparing of adjacent OARs. This approach holds the potential for enhancing patient treatment outcomes and quality of life.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 3","pages":"e70532"},"PeriodicalIF":2.2,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12949144/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"147317135","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Journal of Applied Clinical Medical Physics
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1