Dean Hobbis, Michael D. Armstrong, Samir H. Patel, Riley C. Tegtmeier, Brady S. Laughlin, Shadi Chitsazzadeh, Edward L. Clouser, Jennifer L. Smetanick, Justin Pettit, Justin D. Gagneur, Joshua B. Stoker, Yi Rong, Courtney R. Buckey
{"title":"Comprehensive clinical implementation, workflow, and FMEA of bespoke silicone bolus cast from 3D printed molds using open-source resources","authors":"Dean Hobbis, Michael D. Armstrong, Samir H. Patel, Riley C. Tegtmeier, Brady S. Laughlin, Shadi Chitsazzadeh, Edward L. Clouser, Jennifer L. Smetanick, Justin Pettit, Justin D. Gagneur, Joshua B. Stoker, Yi Rong, Courtney R. Buckey","doi":"10.1002/acm2.14498","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Bolus materials have been used for decades in radiotherapy. Most frequently, these materials are utilized to bring dose closer to the skin surface to cover superficial targets optimally. While cavity filling, such as nasal cavities, is desirable, traditional commercial bolus is lacking, requiring other solutions. Recently, investigators have worked on utilizing 3D printing technology, including commercially available solutions, which can overcome some challenges with traditional bolus.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To utilize failure modes and effects analysis (FMEA) to successfully implement a comprehensive 3D printed bolus solution to replace commercial bolus in our clinic using a series of open-source (or free) software products.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>3D printed molds for bespoke bolus were created by exporting the DICOM structures of the bolus designed in the treatment planning system and manipulated to create a multipart mold for 3D printing. A silicone (Ecoflex 00–30) mixture is poured into the mold and cured to form the bolus. Molds for sheet bolus of five thicknesses were also created. A comprehensive FMEA was performed to guide workflow adjustments and QA steps.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The process map identified 39 and 30 distinct steps for the bespoke and flat sheet bolus workflows, respectively. The corresponding FMEA highlighted 119 and 86 failure modes, with 69 shared between the processes. Misunderstanding of plan intent was a potential cause for most of the highest-scoring failure modes, indicating that physics and dosimetry involvement early in the process is paramount.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>FMEA informed the design and implementation of QA steps to guarantee a safe and high-quality comprehensive implementation of silicone bolus from 3D printed molds. This approach allows for greater adaptability not afforded by traditional bolus, as well as potential dissemination to other clinics due to the open-source nature of the workflow.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":null,"pages":null},"PeriodicalIF":2.0000,"publicationDate":"2024-08-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11539970/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.14498","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Bolus materials have been used for decades in radiotherapy. Most frequently, these materials are utilized to bring dose closer to the skin surface to cover superficial targets optimally. While cavity filling, such as nasal cavities, is desirable, traditional commercial bolus is lacking, requiring other solutions. Recently, investigators have worked on utilizing 3D printing technology, including commercially available solutions, which can overcome some challenges with traditional bolus.
Purpose
To utilize failure modes and effects analysis (FMEA) to successfully implement a comprehensive 3D printed bolus solution to replace commercial bolus in our clinic using a series of open-source (or free) software products.
Methods
3D printed molds for bespoke bolus were created by exporting the DICOM structures of the bolus designed in the treatment planning system and manipulated to create a multipart mold for 3D printing. A silicone (Ecoflex 00–30) mixture is poured into the mold and cured to form the bolus. Molds for sheet bolus of five thicknesses were also created. A comprehensive FMEA was performed to guide workflow adjustments and QA steps.
Results
The process map identified 39 and 30 distinct steps for the bespoke and flat sheet bolus workflows, respectively. The corresponding FMEA highlighted 119 and 86 failure modes, with 69 shared between the processes. Misunderstanding of plan intent was a potential cause for most of the highest-scoring failure modes, indicating that physics and dosimetry involvement early in the process is paramount.
Conclusion
FMEA informed the design and implementation of QA steps to guarantee a safe and high-quality comprehensive implementation of silicone bolus from 3D printed molds. This approach allows for greater adaptability not afforded by traditional bolus, as well as potential dissemination to other clinics due to the open-source nature of the workflow.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic